CMDA flashcards
CMDA · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Audit (ISO 19011)
- A systematic, independent, documented process for obtaining objective evidence and evaluating it objectively against audit criteria to determine the extent to which the criteria are fulfilled.
- Third-party audit
- An audit performed by an independent external organization, typically a certification body, to verify conformity against a standard such as ISO 13485 for the purpose of issuing or maintaining certification.
- Audit criteria
- The set of requirements used as a reference and against which objective evidence is compared, such as ISO 13485, 21 CFR Part 820, EU MDR, and the organization's own procedures.
- Audit evidence
- Records, statements of fact, or other verifiable information which are relevant to the audit criteria and can be confirmed independently; evidence becomes a finding only when compared to a criterion.
- Independence in auditing
- The principle that auditors must be free of bias and conflict of interest and must not audit their own work, ensuring findings are objective and conclusions are credible.
- ISO 13485 clause structure
- Five main clauses: 4 Quality management system, 5 Management responsibility, 6 Resource management, 7 Product realization, and 8 Measurement, analysis and improvement. Clause 7 holds most manufacturing and design requirements.
- 21 CFR 820 subpart J
- The CAPA subpart requiring manufacturers to establish procedures for receiving and investigating complaints, identifying existing and potential causes of nonconformity, and verifying the effectiveness of corrective and preventive actions.
- EU MDR 2017/745
- The European Medical Device Regulation that replaced the Medical Devices Directive and became applicable on 26 May 2021, setting classification rules, clinical evidence, and conformity assessment obligations.
- MDSAP jurisdictions
- Medical Device Single Audit Program covers five regulators: US FDA, Health Canada, Australia's TGA, Brazil's ANVISA, and Japan's MHLW/PMDA, allowing one audit to satisfy all five.
- Device master record
- A compilation of records containing the procedures and specifications for a finished device, required by 21 CFR 820.181 and controlled under document control requirements.
- Risk under ISO 14971
- The combination of the probability of occurrence of harm and the severity of that harm, estimated and then evaluated against acceptability criteria defined in advance in the risk management plan.
- Risk control hierarchy
- The required order of preference for risk control: first inherent safety by design, then protective measures in the device or process, and last information for safety such as labeling and training.
- Verification versus validation
- Verification confirms design outputs meet design inputs; validation confirms the device meets user needs and its intended use under actual or simulated conditions of use.
- IQ OQ PQ
- The three process validation stages: Installation Qualification confirms equipment is installed as specified; Operational Qualification challenges the process across its operating range; Performance Qualification shows consistent output under routine conditions.
- Design History File
- A compilation of records documenting the design history of a finished device, demonstrating that the design was developed in accordance with the approved design plan and 21 CFR 820.30 requirements.
- Audit program PDCA
- Audit program management follows Plan-Do-Check-Act: set program objectives, plan and schedule audits, implement them, monitor performance indicators, and improve the program.
- Auditor competence elements
- Competence combines education, work experience, audit training, knowledge of standards and regulations, knowledge of the organization's processes and products, and required personal attributes such as integrity and objectivity.
- Auditor conflict of interest
- An auditor must decline or disclose any assignment where prior responsibility, commercial pressure, gifts, or personal relationships could compromise impartiality in evaluating objective evidence.
- Audit team leader duties
- The team leader prepares the audit plan, assigns tasks, manages opening and closing meetings, controls the audit pace, resolves disagreements, and approves the final audit report.
- Scope versus criteria
- Scope defines the extent and boundaries of the audit (processes, sites, product lines); criteria define the requirements against which conformity is judged, such as standards, regulations, and internal procedures.
- Sampling in audits
- Audits examine a portion of available records and activities; sampling may be judgmental (risk-based) or statistical, and larger samples are used for complex or historically nonconforming processes.
- Audit plan content
- A documented plan includes audit objectives, scope, criteria, dates and duration, locations, audit team roles, resources, confidentiality arrangements, and provisions for follow-up.
- Checklist function
- A checklist is a memory aid covering the requirements to be examined; it guides coverage but should not restrict the auditor from pursuing evidence that emerges during the audit.
- Objective evidence
- Verifiable data, records, or observed facts that are traceable to a source and can be independently confirmed; it is the only acceptable basis for an audit finding.
- Open-ended questioning
- Questions beginning with what, why, when, where, who, or how that encourage the auditee to describe processes in detail and to volunteer information rather than answer yes or no.
- Audit trail technique
- Following a single record, product, or batch through every process step and department to verify that each stage meets its requirement and that linkages between records are consistent.
- Opening meeting purpose
- To confirm the audit scope, criteria, schedule, methods, and sampling approach; introduce the team; explain reporting and nonconformity handling; and confirm confidentiality, safety, and access arrangements.
- Major nonconformity
- A deficiency indicating the absence of a required system element, a systemic breakdown across processes, or a condition likely to result in a product that does not meet requirements or poses safety risk.
- Correction versus corrective action
- A correction fixes or contains the detected nonconformity; corrective action addresses its root cause so the nonconformity does not recur. Only corrective action prevents recurrence.
- Verification of effectiveness
- Evidence collected after a defined period showing that the corrective action eliminated the root cause and the nonconformity has not recurred, distinct from verifying that the action was implemented.
- FDA Form 483
- The form on which FDA investigators record inspectional observations of objectionable conditions or practices; the facility responds in writing, and the inspection is later classified NAI, VAI, or OAI.
- Audit report content
- A report states audit objectives, scope, criteria, team, dates, findings with objective evidence references, conclusions on QMS effectiveness, and required follow-up actions and deadlines.