CPSGT flashcards
CPSGT · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Alpha rhythm
- The alpha rhythm is an 8–13 Hz EEG rhythm over occipital regions when eyes are closed; it attenuates with eye opening.
- Sleep spindle
- A sleep spindle is a burst of 11–16 Hz EEG activity lasting at least 0.5 seconds, characteristic of N2 sleep.
- K-complex
- A K-complex is a negative sharp wave followed by a positive component, lasting at least 0.5 seconds, seen in N2 sleep.
- Slow-wave activity
- Slow-wave activity consists of EEG waves 0.5–2 Hz with peak-to-peak amplitude greater than 75 µV; it defines N3 when occupying at least 20% of an epoch.
- REM sleep
- REM sleep is a stage with low-amplitude mixed-frequency EEG, rapid eye movements, and the lowest chin EMG tone of the night.
- Arousal scoring
- An arousal is an abrupt EEG frequency shift lasting at least 3 seconds; in REM it requires a concurrent chin EMG increase of at least 1 second.
- Apnea definition
- An apnea is a ≥90% reduction in airflow lasting at least 10 seconds; it is obstructive when respiratory effort persists and central when effort is absent.
- Hypopnea definition
- A hypopnea is a ≥30% reduction in airflow lasting at least 10 seconds with ≥3% oxygen desaturation or an arousal.
- Sleep efficiency
- Sleep efficiency equals total sleep time divided by time in bed, multiplied by 100, and is reported as a percentage.
- MSLT diagnostic criteria
- The MSLT consists of five naps; a mean sleep latency ≤8 minutes plus at least two SOREMPs supports a narcolepsy diagnosis.
- International 10-20 system
- The international 10–20 system places electrodes using 10% and 20% distances between skull landmarks so EEG locations are reproducible across patients.
- EEG impedance limit
- EEG electrode impedance should generally be ≤5 kΩ before recording; higher impedance increases 60 Hz interference and signal loss.
- EOG electrode placement
- Left EOG is placed 1 cm below and 1 cm lateral to the left outer canthus; right EOG is placed 1 cm above and lateral to the right outer canthus.
- Chin EMG placement
- Chin EMG uses three electrodes: two below the chin and one above, arranged to record muscle tone without bridging.
- Skin preparation
- Skin preparation cleans the site with alcohol and mild abrasive to lower impedance while avoiding excessive abrasion that causes pain or bleeding.
- Oximeter placement
- The oximeter probe is placed on a warm, well-perfused finger; nail polish is removed and the waveform is checked for adequate signal.
- Airflow sensor types
- A nasal pressure cannula detects airflow reduction and hypopneas, while a thermistor detects temperature change and is better for identifying apnea.
- Pre-study calibration
- Before lights out, the technologist performs bio-calibration, checks impedances and signals, and documents any deviations from expected setup.
- EEG sampling rate
- EEG is recorded at a sampling rate of at least 200 Hz, commonly 200–500 Hz, to capture brief features needed for staging.
- EEG filter settings
- Standard EEG filter settings are a 0.3 Hz low-frequency filter and a 35 Hz high-frequency filter; changes are documented.
- EMG filter settings
- Chin EMG typically uses a 10 Hz low-frequency filter and a 100 Hz high-frequency filter to display muscle tone clearly.
- 60 Hz artifact
- 60 Hz artifact is electrical interference that appears as thick, regular high-frequency activity; check electrode impedance, ground, and cable routing.
- Sweat artifact
- Sweat artifact is a slow rolling baseline change caused by perspiration; drying the skin and improving electrode adhesion reduce it.
- Event marker
- An event marker is an annotation placed in the recording to identify snores, movements, interventions, or technical events for scoring.
- Technologist note content
- A technologist note documents medications, patient complaints, interventions, artifacts, lights out and on times, and unusual events.
- Calibration signal
- A calibration signal is a 50 µV square wave used to verify amplifier gain and channel response before the recording begins.
- CPAP titration goal
- CPAP titration adjusts positive airway pressure to eliminate apneas and hypopneas and to maintain SpO2 at or above 90%.
- Oxygen titration
- Oxygen titration adjusts supplemental oxygen flow per order to keep SpO2 at or above 90%, with flow and response documented.
- Pre-study patient instructions
- Before a sleep study, the technologist explains the sensors and procedure and instructs the patient to avoid caffeine and alcohol and wash hair.
- Emergency response
- For chest pain, difficulty breathing, or loss of consciousness, call for help, stay with the patient, and follow the emergency protocol.
- Hand hygiene
- Hand hygiene is performed before and after patient contact and before applying or removing sensors to reduce infection transmission.
- Scope of practice
- An entry-level polysomnographic technician performs supervised tasks within training and escalates complex decisions or abnormal findings to a supervisor.