- Bremsstrahlung radiation
- Continuous X-ray spectrum produced when electrons decelerate in the anode target; maximum photon energy equals the applied kVp, and intensity rises with kVp squared.
- Line-focus principle
- The anode is bevelled so the effective focal spot is smaller than the actual focal spot, improving spatial resolution while spreading heat over a larger target area.
- Total filtration
- Sum of inherent and added filtration in the X-ray tube assembly; it removes low-energy photons, hardens the beam, and reduces patient skin dose.
- Inverse square law
- Radiation intensity is inversely proportional to the square of the distance from a point source; doubling the distance reduces the dose rate to one quarter.
- PA chest projection
- Posteroanterior projection of the chest at about 180 cm focus-to-receptor distance; it reduces cardiac magnification and scapular overlap compared with AP.
- Frog-leg lateral hip
- Lateral projection of the hip with the patient supine, knees flexed, and hips abducted about 45 degrees; it demonstrates the femoral neck and head.
- Open-mouth odontoid view
- AP projection of the upper cervical spine taken through the open mouth to show C1 and C2, including the dens and lateral masses of the atlas.
- Two-view rule
- At least two orthogonal projections at 90 degrees are required for suspected fractures, because a single view can miss displacement or a fracture line.
- Hurter-Driffield curve
- Characteristic curve plotting optical density against log relative exposure; the straight-line portion indicates contrast, and toe and shoulder indicate under- and overexposure.
- Optical density
- Logarithm of the ratio of incident to transmitted light; a density of 1.0 transmits 10 percent of light, and 2.0 transmits 1 percent.
- Photostimulable phosphor plate
- Computed radiography receptor that stores X-ray energy in metastable electron traps; a scanning laser releases blue light read by a photomultiplier tube.
- Window width and level
- Display settings in digital radiography; window width controls contrast and window level controls brightness of the image on the monitor.
- Safelight test
- Quality control test using a coin on a pre-exposed film; a visible coin shadow after development indicates unsafe safelight or excessive darkroom fog.
- Developer temperature
- Manual processing standard of about 20 to 22 degrees Celsius; each 1 degree rise requires roughly a 10 percent reduction in development time.
- Sensitometer and densitometer
- Sensitometer exposes a step wedge on film; densitometer measures optical density of each step to produce a processor control chart for quality assurance.
- Pi-line artifact
- Black line on a radiograph caused by a crack or scratch in the intensifying screen or cassette; it indicates the screen must be replaced.
- ALARA principle
- As Low As Reasonably Achievable; all exposures must be optimized so patient and staff doses are kept below dose limits and as low as practical.
- Occupational dose limit
- Effective dose of 20 mSv per year averaged over five years, with no single year exceeding 50 mSv; the eye lens limit is 20 mSv per year.
- Thermoluminescent dosimeter
- Personal monitor using lithium fluoride; absorbed radiation is stored and released as light when heated, giving a measure of cumulative dose.
- Lead apron equivalence
- A 0.25 mm lead-equivalent apron attenuates about 90 percent of diagnostic X-rays; 0.35 mm gives greater protection and aprons are checked annually for cracks.
- Femoral neck-shaft angle
- In adults the angle between the femoral neck and shaft is about 125 degrees; a reduced angle indicates coxa vara and an increased angle coxa valga.
- Radiographic densities
- Standard scale from black to white: air, fat, soft tissue and fluid, bone and calcification, and metal; each step is progressively more attenuating.
- Carina level
- The tracheal bifurcation lies at about the T4-T5 vertebral level; it is a key landmark on the PA chest radiograph for assessing lung position.
- Pneumothorax sign
- Absent lung markings peripheral to a visible pleural line on the chest radiograph, with possible mediastinal shift in a tension pneumothorax.
- Barium sulphate
- Insoluble inert positive contrast agent for the gastrointestinal tract; contraindicated if perforation is suspected because it causes peritonitis.
- Intravenous urogram
- Examination using intravenous iodinated contrast with films at about 1, 5, 15, and 25 minutes to show the nephrogram and collecting system.
- Digital subtraction angiography
- Technique that subtracts a pre-contrast mask image from the contrast image to show vessels; patient movement causes misregistration artifact.
- Double-contrast barium enema
- Study of the colon using barium sulphate plus air insufflation to coat the mucosa and distend the lumen, improving detection of small polyps.
- Anaphylaxis first-line drug
- Adrenaline 0.5 mg intramuscularly as 1:1000 solution into the anterolateral thigh, repeated as needed, for contrast-induced anaphylaxis with hypotension or bronchospasm.
- Adult BLS compression ratio
- Chest compressions at 100 to 120 per minute with a depth of 5 to 6 cm, given in a ratio of 30 compressions to 2 rescue breaths.
- Four ethical principles
- Autonomy, beneficence, non-maleficence, and justice; these guide consent, confidentiality, and decisions in radiological practice.
- 10-day rule
- Abdominal and pelvic radiography in women of reproductive age should be performed within 10 days of the onset of menstruation to avoid irradiating an early pregnancy.