Maharashtra HSC Radiology Technician (Vocational) · Work through your exam, one topic at a time.

All guide sections

Radiation Physics and X-ray Equipment

Radiation Physics and X-ray Equipment

This domain tests how the X-ray tube generates a beam and how generator settings alter it. Questions target tube components, rectification, beam quality and quantity, and attenuation with distance.

Core concepts

  • Thermionic emission from the tungsten filament releases electrons; the focusing cup directs them to the anode focal spot. A smaller focal spot sharpens the image but limits mA.
  • Electrons convert to heat (over 99%) and X-rays (under 1%). Bremsstrahlung produces a continuous spectrum; characteristic radiation produces discrete K-shell peaks.
  • kVp controls penetration and also output; mAs controls quantity. Exposure rises with mAs and approximately with kVp squared to kVp to the fifth power.
  • Full-wave rectification uses both half-cycles; high-frequency generators give a more constant potential and higher effective kVp.
  • Air kerma obeys the inverse-square law.

Worked example

A tube produces 400 µGy at 100 cm. At 200 cm: I2 = 400 × (100/200)² = 400 × 0.25 = 100 µGy. Doubling distance spreads the beam over four times the area, so intensity falls fourfold.

Traps

  • Saying kVp changes only quality; it changes output too.
  • Confusing mA (current) with mAs (current × time).
  • Treating the heel effect as focusing; it reduces intensity on the anode side.