ITC Thermographer Level III flashcards
ITC Thermographer Level III · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Stefan-Boltzmann law
- Total radiant exitance of a blackbody equals the Stefan-Boltzmann constant times absolute temperature to the fourth power: M = σT⁴. Radiated power rises steeply with temperature.
- Wien displacement law
- Peak emission wavelength equals 2898 micrometre-kelvin divided by absolute temperature. A 300 K surface peaks near 9.7 micrometres, favouring longwave cameras for ambient targets.
- Kirchhoff's law
- For an opaque body, emissivity equals absorptivity at each wavelength and direction, and reflectivity equals one minus emissivity. Low-emissivity surfaces mostly reflect background radiation.
- Apparent temperature
- The temperature a camera reports after applying its emissivity and background settings. It equals true surface temperature only when those inputs match the actual scene.
- Emissivity of polished aluminum
- Roughly 0.05, so polished aluminum is a poor emitter and strongly reflects its surroundings. Accurate measurement usually requires a high-emissivity coating or tape.
- Cavity effect
- Enclosed geometries such as furnace openings and deep holes trap reflections, raising apparent emissivity toward one. They behave closer to a blackbody than a flat surface.
- ISO 18434-1 scope
- Standard covering condition monitoring and diagnostics of machines using thermography. It addresses survey methods, measurement conditions, and interpretation of thermal results.
- ISO 18436-7 categories
- Defines thermographer categories: Level I performs basic measurements, Level II interprets per established techniques, Level III develops procedures and manages the program.
- ASTM E1934
- Standard guide for examining electrical and mechanical equipment with infrared thermography. It covers survey conditions, qualitative and quantitative interpretation, and reporting.
- Written practice
- Program document defining training hours, experience, examination content, acceptance criteria, and recertification intervals for each thermographer category. It controls qualification.
- Baseline trending conditions
- Record load, ambient temperature, wind, solar load, distance, and emissivity settings for every survey so later measurements are comparable to the baseline.
- Acceptance criteria
- Pre-agreed thresholds for judging a finding, such as delta-T over similar components, absolute temperature limits, or trend deviation. They must be written before the survey.
- Reflected temperature method
- Measure apparent temperature of crumpled aluminum foil near the target with emissivity set to one. Enter that value as the background temperature for the target.
- Emissivity by reference tape
- Apply black tape or paint of known emissivity to the target, allow equilibration, and compare its reading with the bare surface to derive the surface emissivity.
- Thermal resistance analogy
- Heat flow equals temperature difference divided by thermal resistance. A rising delta-T across the same path at constant load indicates increasing resistance.
- Severity classification
- Ranking findings by combining delta-T, load, component criticality, and trend history. A small delta-T on a critical component can outrank a large one elsewhere.
- Envelope leak detection
- Moisture and air leaks are visible only when a temperature difference is created by heating, cooling, or pressure difference. Without a driving force no pattern appears.
- Report content elements
- Object and location, survey conditions, camera and settings, radiometric image, temperature values and method, severity, limitations, and recommended action with priority.
- Microbolometer detector
- Uncooled longwave detector array, typically 8 to 14 micrometres, lower cost and slower response than cooled photon detectors. Common in handheld cameras.
- IFOV formula
- Instantaneous field of view equals detector pitch divided by focal length, in milliradians. It gives the pixel footprint at distance and drives spot-size calculations.
- Spot-size pixel rule
- The target should cover at least three by three pixels for a measurement, and ten by ten pixels for critical work, to avoid dilution by background temperature.
- Camera accuracy specification
- Usually stated as plus or minus two degrees Celsius or two percent of reading, whichever is greater, within a defined ambient range. Outside that range a temperature coefficient applies.
- NETD
- Noise equivalent temperature difference: the smallest resolvable temperature difference at a stated target temperature and averaging. Lower values indicate better sensitivity.
- Spectral band selection
- Longwave suits ambient targets and outdoor work; midwave suits high-temperature targets, through-flame measurement, and some plastics and glass. Match band to target temperature.
- Focus as measurement parameter
- Defocus spreads target energy over more pixels and biases the reading toward background temperature. Correct focus is required before capturing radiometric data.
- Spot-size calculation example
- With IFOV 1.3 milliradians at 10 metres, the pixel footprint is 13 millimetres. The three-by-three rule then requires a target of about 39 millimetres.
- Load correction for electrical
- Resistive losses scale roughly with the square of current. If the survey load differs from rated load, record it because delta-T changes with load.
- Uncertainty budget
- Combine emissivity, reflected background, atmospheric transmittance, distance, and camera accuracy contributions to state an expanded measurement uncertainty with a coverage factor.
- Level III duties
- Per ISO 18436-7, Level III develops procedures, manages the program, performs advanced analysis, and qualifies Level I and II personnel. Recertification is required at defined intervals.
- Blackbody verification
- Check camera measurement against a blackbody reference at multiple temperatures across the working range, at defined intervals and after repair or firmware change.
- Solar loading artifact
- Solar heating creates false hot indications and wind cooling creates false cold ones. Survey after solar load decays and record wind speed, ideally below about five metres per second.
- Thermal lag
- Thick, high-thermal-mass objects respond slowly to internal faults. Allow time for steady-state conditions before measuring to avoid understating the fault.