JBA Written Exam study materials
JBA Written Exam study resources

JBA Written Exam cheat sheet

JBA Written Exam · Keep the key ideas within reach.

Quick reference

Biosafety and Facility Management

  • Biosafety levels map to risk groups: BSL-1 (RG1, no disease), BSL-2 (RG2, moderate hazard), BSL-3 (RG3, aerosol-transmitted serious disease), BSL-4 (RG4, life-threatening, no treatment).

  • Directional airflow is from clean to contaminated zones; BSL-3/4 suites hold negative pressure differentials (typically -10 to -50 Pa) relative to corridors.

  • Biological safety cabinets: Class I = inward flow, no product protection; Class II = personnel + product + environment; Class III = gas-tight glovebox for BSL-4.

  • HEPA filters retain ≥99.97% of particles at 0.3 µm; BSL-3 exhaust HEPA is tested annually and after any structural change (decontamination or filter change).

  • Autoclave efficacy: 121 °C at 15 psi (103 kPa) for 15-30 min, or 134 °C for 3-4 min; validated with biological indicators (Geobacillus stearothermophilus spores).

  • Chemical disinfection: 70% ethanol for surfaces, 1:10 household bleach (0.5% sodium hypochlorite) for blood spills, 5-10% bleach for high-concentration organic spill; contact time 10-30 min.

  • Waste segregation: sharps into puncture-resistant containers; biohazard waste into red bags; liquid waste decontaminated before drain discharge; autoclave tape is a process indicator, not a sterilization guarantee.

  • Primary containment = cabinet/PPE; secondary containment = facility envelope, HVAC, waste systems. Both are required for BSL-3+ operations.

Quick reference

Biohazard Risk Assessment and Containment

  • Risk assessment considers agent pathogenicity, infectious dose, transmission route, host range, availability of prophylaxis, and environmental stability (BSL classification by WHO/NIH).

  • Containment level is set by the highest-risk procedure, not the agent alone: aerosol-generating steps (centrifugation, vortexing, sonication) may require a higher BSL than routine handling.

  • Biosafety vs biosecurity: biosafety addresses accidental exposure; biosecurity addresses intentional misuse, theft, or diversion - both must be assessed.

  • Risk = likelihood × consequence; residual risk is assessed after controls are applied, and re-assessed when procedures, agents, or personnel change.

  • Aerosol-generating procedures require Class II BSC use; open-bench work is acceptable only for RG1/RG2 with no significant aerosol generation.

  • Spill response hierarchy: alert, evacuate, allow aerosol settle (≥30 min), don PPE, cover with absorbent, apply disinfectant, remove, decontaminate, report.

  • Personnel training, medical surveillance, and immunization (e.g., HBV, rabies) are part of the risk control plan for RG2/RG3 work.

  • Genetically modified organisms are assessed under the same framework; containment level may be raised if the modification increases virulence or environmental fitness.

Quick reference

Biological Resource Management and Quality Control

  • Biological Resource Centers follow OECD Best Practice Guidelines and ISO 20387 (biobanking); key elements are accession, authentication, viability, and traceability.

  • Cell line authentication: STR profiling (human) or species-specific markers; misidentification is a leading cause of irreproducible research.

  • Mycoplasma contamination is detected by PCR or culture; routine screening every 1-3 months is standard for continuous cell lines.

  • Cryopreservation: slow cooling (~-1 °C/min) to -80 °C, then liquid nitrogen (-196 °C); DMSO 5-10% is a common cryoprotectant; viability on thaw should be ≥70-80%.

  • Seed lot and working stock systems: master seed and working seed lots are characterized once and used for production; passage number is tracked to limit genetic drift.

  • Quality control of culture media includes sterility, growth promotion, and pH checks; each lot is tested before use in production.

  • Inventory management requires unique identifiers, location mapping, and audit trails; a documented chain of custody is essential for regulatory acceptance.

  • Reference standards and controls must be traceable and periodically re-qualified; expired or degraded standards invalidate assay results.

Quick reference

Bioindustry Regulations and Standards

  • ICH Q7 covers GMP for active pharmaceutical ingredients; ICH Q8-Q12 cover pharmaceutical development, quality risk management, quality systems, and lifecycle management.

  • GMP pillars: quality management, personnel, premises, equipment, documentation, production, quality control, contract manufacturing, complaints/recalls, self-inspection.

  • Data integrity ALCOA+: Attributable, Legible, Contemporaneous, Original, Accurate, plus Complete, Consistent, Enduring, Available.

  • 21 CFR Part 11 governs electronic records and signatures: audit trails, access controls, and validation are required.

  • ISO 13485 applies to medical devices; ISO 9001 is generic quality management; ISO 17025 applies to testing/calibration laboratories.

  • Validation lifecycle: IQ (installation) → OQ (operation) → PQ (performance); process validation follows a three-batch approach under FDA guidance.

  • Change control and deviation management are mandatory; any change to process, equipment, or supplier requires documented impact assessment.

  • Pharmacopoeial standards (USP, EP, JP) define tests and acceptance criteria for compendial products; deviations require justification.

Quick reference

Biotechnology Fundamentals and Applications

  • Central dogma: DNA → RNA → protein; PCR amplifies DNA exponentially (2^n after n cycles); qPCR quantifies starting template via Ct value.

  • CRISPR-Cas9 uses a guide RNA to target a complementary sequence adjacent to a PAM (e.g., NGG for SpCas9) and creates a double-strand break repaired by NHEJ or HDR.

  • Monoclonal antibodies are produced in CHO or hybridoma systems; upstream (bioreactor) and downstream (chromatography, filtration) define the process.

  • Bioprocess parameters: dissolved oxygen, pH, temperature, feed rate, and viable cell density; fed-batch and perfusion are the main modes.

  • Protein purification: affinity (Protein A) → ion exchange → size exclusion; purity assessed by SDS-PAGE and HPLC.

  • Vaccine platforms: live attenuated, inactivated, subunit, toxoid, mRNA, and viral vector; each has distinct cold-chain and safety profiles.

  • Enzyme kinetics: Michaelis-Menten v = Vmax[S]/(Km+[S]); Km is the substrate concentration at half-maximal velocity.

  • Biosimilars require analytical similarity, functional assays, and clinical comparability - not identical to the originator but highly similar.

Quick reference

Quality Assurance and Good Manufacturing Practice

  • QA is the system ensuring quality; QC is the testing function that verifies it; QA approves, QC measures - the two are organizationally independent.

  • Batch record review and release by the Qualified Person (EU) or authorized release (US) precedes distribution; no batch may be released without documented review.

  • Environmental monitoring (EM) in cleanrooms: viable and non-viable particle counts; Grade A/ISO 5 for critical zones; Grade B background for aseptic processing.

  • Cleaning validation demonstrates removal of product, cleaning agents, and microbes to predetermined limits; worst-case product and swab sites are selected.

  • CAPA (Corrective and Preventive Action) closes deviations; root cause analysis uses tools such as 5-Whys and fishbone diagrams.

  • Annual Product Quality Review (APQR) evaluates trends in batches, deviations, and complaints; required by EU GMP and expected by FDA.

  • Out-of-specification (OOS) results trigger a documented investigation; retesting is permitted only under a written, approved protocol.

  • Training records, SOPs, and change control form the documentary backbone of GMP; undocumented work is treated as not performed.

Quick reference

Ethics, Compliance, and Environmental Health and Safety

  • Institutional Biosafety Committee (IBC) reviews recombinant DNA and dual-use research; Institutional Review Board (IRB) reviews human-subjects research; IACUC reviews animal work.

  • Informed consent requires disclosure, comprehension, voluntariness, and capacity; it is a process, not just a signature.

  • Dual-use research of concern (DURC) is legitimate research that could be misused; oversight includes risk mitigation plans and publication review.

  • Environmental health and safety: hazard communication (SDS access), PPE selection, chemical fume hoods, and emergency eyewash/showers within 10 seconds' travel.

  • Waste hierarchy: reduce, reuse, recycle, treat, dispose; biohazardous waste must be decontaminated before leaving the facility.

  • Occupational exposure limits (OELs) and biological exposure indices guide worker protection; exposure monitoring is documented.

  • Whistleblower and non-retaliation protections apply to reporting of safety or compliance violations; retaliation is prohibited.

  • Sustainability in biotech includes solvent recovery, single-use plastic reduction, and energy-efficient HVAC; green chemistry principles (12 principles) guide process design.

Quick reference

Biosafety and Facility Management

  • BSL-2 requires handwashing sinks, autoclave access, and a biosafety manual; BSL-3 adds controlled access, sealed penetrations, and directional airflow.

  • BSC certification is performed at installation, after repair, and annually; the cabinet must be decontaminated (formaldehyde gas or VHP) before service.

  • Air changes per hour (ACH) in BSL-3 suites are typically ≥12; BSL-4 requires higher redundancy and HEPA on both supply and exhaust.

  • Autoclave load configuration affects penetration: bags and containers must allow steam contact; a Bowie-Dick test verifies air removal in porous loads.

  • Spill kit contents: absorbent, disinfectant, gloves, forceps, biohazard bags, and a written procedure; kits are checked on a defined schedule.

  • Signage at BSL-2+ doors states agent, biosafety level, responsible person, and required PPE; universal biohazard symbol is mandatory.

  • Pest control, restricted access, and visitor logging are facility controls that support biosecurity as well as biosafety.

Quick reference

Biohazard Risk Assessment and Containment

  • A risk assessment is documented, dated, and reviewed at least annually or whenever a new agent, procedure, or personnel change occurs.

  • Infectious dose varies by route: inhalation often requires fewer organisms than ingestion; route-specific assessment guides PPE and engineering controls.

  • Sharps injuries are the leading exposure route in laboratories; safe sharps devices and no-recapping policies reduce risk.

  • Pregnant, immunocompromised, and immunized workers require individualized risk assessment; some agents (e.g., CMV, Toxoplasma) pose special fetal risk.

  • Decontamination validation uses biological indicators; chemical disinfectant efficacy depends on concentration, contact time, temperature, and organic load.

  • Biosecurity tiers: inventory control, access control, personnel reliability, and transport security; select agents require registration under national rules.

  • Post-exposure management includes first aid, reporting, medical evaluation, and prophylaxis (e.g., HIV PEP within hours).

Quick reference

Biological Resource Management and Quality Control

  • ISO 20387:2018 specifies competence, impartiality, and consistency requirements for biobanks; accreditation is via a national body.

  • Sample quality metrics: pre-analytical variables (ischemia time, fixation), storage temperature, freeze-thaw cycles, and RNA integrity number (RIN).

  • Cell bank characterization includes identity, purity, sterility, mycoplasma, and adventitious agent testing (in vivo and in vitro).

  • Biological resource centers must have a documented succession and backup plan; cryogenic storage requires LN2 level monitoring and alarm systems.

  • Passage number limits are set in the master cell bank protocol; exceeding limits requires re-qualification or a new bank.

  • Contamination control in cell culture: aseptic technique, antibiotic-free cultures for detection, and quarantine of new lines.

  • Biospecimen traceability links donor consent, collection, processing, and distribution; consent scope limits downstream use.

Quick reference

Bioindustry Regulations and Standards

  • EU GMP Annex 1 (2022 revision) sets contamination control strategy requirements for sterile products, including CCS documentation and PUPSIT considerations.

  • FDA 21 CFR 210/211 covers finished pharmaceuticals; 21 CFR 600 covers biologics; 21 CFR 820 covers medical devices (QSR).

  • ICH Q9(R1) quality risk management uses tools such as FMEA, HACCP, and risk ranking/filtering; risk-based decisions must be documented.

  • ICH Q10 establishes the pharmaceutical quality system across the product lifecycle: development, technology transfer, commercial, and discontinuation.

  • Stability testing (ICH Q1A) defines long-term, intermediate, and accelerated conditions; shelf life is supported by real-time data.

  • Registration pathways: IND/NDA/BLA (US), MAA (EU), and PMDA (Japan); each requires CMC, nonclinical, and clinical modules (CTD format).

  • Serialization and track-and-trace requirements (e.g., US DSCSA, EU FMD) apply to prescription products at the unit level.

Quick reference

Biotechnology Fundamentals and Applications

  • Restriction enzymes cut at palindromic recognition sites; sticky ends ligate more efficiently than blunt ends; ligation requires compatible buffers and ATP.

  • Plasmid vectors carry an origin of replication, selectable marker, and multiple cloning site; transformation efficiency is reported as CFU per µg DNA.

  • Sanger sequencing uses chain-terminating dideoxynucleotides; next-generation sequencing uses massively parallel clonal amplification and short reads.

  • mRNA vaccines use lipid nanoparticles (LNPs) for delivery; the 5' cap and poly(A) tail improve stability and translation.

  • Bioprocess scale-up follows geometric and process parameters (tip speed, kLa, mixing time); scale-down models must be qualified.

  • Analytical methods for biologics: ELISA, surface plasmon resonance, mass spectrometry, and capillary electrophoresis for charge variants.

  • CRISPR off-target assessment uses computational prediction plus empirical assays (GUIDE-seq, CIRCLE-seq); HDR efficiency is lower than NHEJ.

Quick reference

Quality Assurance and Good Manufacturing Practice

  • Deviation classification: critical (patient safety/regulatory), major (product quality), minor (documentation); classification drives investigation depth.

  • Change control categories: minor, major, critical; critical changes require regulatory notification or prior approval before implementation.

  • Media fill (process simulation) qualifies aseptic filling; acceptance criteria are defined in Annex 1 and by the manufacturer's validated limits.

  • Supplier qualification includes audits, certificates of analysis, and quality agreements; critical suppliers are audited on a defined cycle.

  • Complaint handling and recall classification (Class I, II, III) are defined by health hazard; recall effectiveness checks are documented.

  • Validation master plan (VMP) lists all qualified systems and their status; periodic revalidation is triggered by change or trend.

  • ALCOA+ data integrity: audit trails must be enabled and reviewed; shared logins and backdating are prohibited.

Quick reference

Ethics, Compliance, and Environmental Health and Safety

  • Belmont Report principles: respect for persons, beneficence, and justice; these underpin modern human-subjects protections.

  • Declaration of Helsinki and CIOMS guidelines govern international research ethics; local law and ethics committee approval are both required.

  • Animal research follows the 3Rs (replacement, reduction, refinement); IACUC approval and humane endpoints are mandatory.

  • Conflict-of-interest disclosure is required for investigators, reviewers, and committee members; undisclosed conflicts can invalidate research.

  • Chemical safety: GHS classification and labeling; SDS sections 1-16; incompatible chemicals are stored separately.

  • Biological spill or exposure is a reportable event; root-cause analysis and corrective action follow, with records retained per regulation.

  • Environmental permits may be required for wastewater discharge, air emissions, and hazardous waste; exceedances trigger notification.

Quick reference

Biosafety and Facility Management

  • BSC airflow: 70% recirculated, 30% exhausted in Class II Type A2; Type B2 is 100% exhaust and is used with volatile chemicals.

  • Room pressure monitoring uses magnehelic gauges or sensors with alarms; loss of negative pressure is a reportable containment breach.

  • Decontamination of a BSL-3 room uses vaporized hydrogen peroxide or formaldehyde; efficacy is verified with biological indicators.

  • Emergency shower and eyewash stations are flushed weekly and inspected annually; access must be unobstructed.

  • Waste autoclave validation includes thermocouples in the load and biological indicators; cycle records are retained.

  • Facility commissioning includes airflow visualization (smoke test) and pressure decay testing of sealed rooms.

Quick reference

Biohazard Risk Assessment and Containment

  • Risk group classification (WHO) is distinct from biosafety level: RG reflects agent hazard; BSL reflects the containment applied to the procedure.

  • Routes of exposure: inhalation, ingestion, percutaneous (sharps), and mucous membrane; controls are matched to route.

  • Aerosol-generating equipment includes centrifuges (use sealed rotors), blenders, and sonicators; these are operated in a BSC or closed system.

  • PPE selection: gloves (nitrile), gown, eye protection, and respiratory protection (N95/PAPR) based on risk assessment.

  • Waste and transport: UN packaging and triple containment for Category A infectious substances (UN2814); Category B is UN3373.

  • Biorisk management standard ISO 35001 provides a framework for laboratory biorisk management systems.

Quick reference

Biological Resource Management and Quality Control

  • Cell line identity: STR loci for human lines; species-specific PCR or karyotyping for non-human lines; identity is confirmed at bank creation and after significant passage.

  • Viability assays: trypan blue exclusion and automated counters; viability below 70% may indicate poor cryopreservation or handling.

  • Adventitious agent testing includes sterility, mycoplasma, in vitro and in vivo assays, and retrovirus testing for rodent cell lines.

  • Sample storage: -80 °C for working aliquots, LN2 vapor for long-term; avoid repeated freeze-thaw; record freeze-thaw cycles.

  • Quality control of reference materials includes homogeneity, stability, and commutability assessment.

  • Documentation: chain of custody, informed consent, and material transfer agreements (MTAs) govern distribution.

Quick reference

Bioindustry Regulations and Standards

  • GMP inspection types: routine, for-cause, and pre-approval; inspection findings are classified as critical, major, or minor (or FDA Form 483 observations).

  • Regulatory submissions use the Common Technical Document (CTD) format: Modules 1-5 (regional, quality, nonclinical, clinical).

  • ICH Q3A/Q3B set impurity limits for drug substances and products; ICH Q6A/B define specifications and test procedures.

  • Good Distribution Practice (GDP) requires temperature-controlled transport, qualification of shippers, and deviation handling.

  • Pharmacovigilance: adverse events are reported within defined timelines (e.g., 15 days for serious unexpected suspected adverse reactions in the EU).

  • Medical device classification (Class I-III in EU; Class I-III in US) drives conformity assessment and premarket submission requirements.

Quick reference

Biotechnology Fundamentals and Applications

  • PCR cycle: denaturation (~95 °C), annealing (~55-65 °C), extension (~72 °C); primer Tm and Mg2+ concentration affect specificity.

  • qPCR quantification: standard curve (absolute) or ΔΔCt (relative); Ct is inversely proportional to starting template.

  • Protein expression systems: E. coli (fast, no glycosylation), yeast (glycosylation), CHO (human-like glycosylation), and insect cells (baculovirus).

  • Chromatography: affinity captures, ion exchange polishes by charge, size exclusion desalts and removes aggregates; resin lifetime is validated.

  • Biosimilarity assessment includes analytical, functional, nonclinical, and clinical comparability; extrapolation to other indications requires justification.

  • Gene therapy vectors: AAV (non-integrating, limited cargo), lentivirus (integrating), and adenovirus (transient, immunogenic).

Quick reference

Quality Assurance and Good Manufacturing Practice

  • Quality by Design (QbD): define quality target product profile (QTPP), critical quality attributes (CQAs), and critical process parameters (CPPs); design space is established in development.

  • Process validation stages (FDA 2011): Stage 1 process design, Stage 2 process qualification (PPQ), Stage 3 continued process verification (CPV).

  • Continued process verification uses statistical process control (SPC) charts and trend analysis to detect drift.

  • Cleaning validation acceptance limits: health-based exposure limits (HBEL/PDE) and visually clean criteria; rinse and swab sampling are both used.

  • Stability program: ongoing stability studies are required annually for at least one batch per product per year.

  • Self-inspection and quality audits are scheduled at defined intervals; findings are tracked to closure with CAPA.

Quick reference

Ethics, Compliance, and Environmental Health and Safety

  • Research misconduct: fabrication, falsification, and plagiarism (FFP); honest error and differences of opinion are not misconduct.

  • Authorship requires substantial contribution, drafting or critical revision, final approval, and accountability; ghost and gift authorship are prohibited.

  • Data privacy: GDPR and HIPAA govern personal and health data; de-identification and consent scope must be documented.

  • Biosafety and ethics oversight overlap in DURC review; publication of dual-use findings may require risk mitigation or redaction.

  • Environmental monitoring includes air, water, and soil sampling for facility emissions; results are compared against permit limits.

  • Occupational health surveillance includes pre-placement, periodic, and exit examinations for workers with hazardous exposures.

Quick reference

Biosafety and Facility Management

  • Biosafety cabinet placement: away from doors, drafts, and high-traffic areas; airflow disturbance reduces containment.

  • Class II BSC airflow alarm and sash alarm must be functional; the cabinet is not used if airflow is out of specification.

  • BSL-3 facility features: self-closing doors, sealed windows, HEPA-filtered exhaust, and a pass-through autoclave or dunk tank.

  • Decontamination of equipment before removal from containment uses a validated method (VHP, formaldehyde, or double-bagging with surface disinfection).

  • Emergency power and backup systems support HVAC, freezers, and alarms; generators are tested on a defined schedule.

  • Wastewater from BSL-4 must be heat-treated or chemically decontaminated before discharge.

Quick reference

Biohazard Risk Assessment and Containment

  • Risk assessment documentation includes agent, titer, volume, procedure, route, controls, and residual risk; it is signed and dated.

  • Biological safety cabinets protect product and personnel but not from volatile chemicals unless ducted (Type B2) or fitted with a canopy.

  • Sharps policy: no recapping, no bending, no removal from syringes; disposal immediately after use into approved containers.

  • Spill outside the BSC requires evacuation and decontamination; spill inside the BSC is cleaned with the cabinet running.

  • Biosecurity inventory: select agents are tracked, secured, and inventoried; discrepancies are reported to authorities.

  • Post-exposure prophylaxis for HIV is most effective when started within hours; reporting and source testing follow protocol.

Quick reference

Biological Resource Management and Quality Control

  • Cell bank tiers: master cell bank (MCB) and working cell bank (WCB); the WCB is prepared from the MCB and used for production.

  • Mycoplasma testing methods: culture (broth and agar) and PCR; PCR is faster but should be paired with culture for release.

  • Cryopreservation: controlled-rate freezing at ~-1 °C/min; DMSO toxicity is minimized by rapid dilution on thaw.

  • Sample quality: RIN ≥7 for RNA sequencing; DNA integrity assessed by gel or fragment analysis.

  • Biobank accreditation (ISO 20387) covers governance, quality, and technical competence; audits are periodic.

  • Material transfer agreements define permitted use, ownership, and redistribution; unauthorized sharing breaches the agreement.

Quick reference

Bioindustry Regulations and Standards

  • FDA Form 483 lists inspectional observations; a Warning Letter can follow if responses are inadequate.

  • EU GMP Annex 1 requires a contamination control strategy (CCS) covering facility, equipment, personnel, and processes.

  • ICH Q11 covers development and manufacture of drug substances, including starting material selection and control strategy.

  • Stability: ICH Q1A defines long-term (25 °C/60% RH or 30 °C/65% RH), intermediate (30 °C/65% RH), and accelerated (40 °C/75% RH) conditions.

  • Serialization: US DSCSA requires unit-level traceability; EU FMD requires unique identifiers and tamper-evident seals.

  • Regulatory inspections may be announced or unannounced; access to records and facilities is a legal requirement.

Quick reference

Biotechnology Fundamentals and Applications

  • Restriction enzyme nomenclature: organism abbreviation + strain + Roman numeral (e.g., EcoRI); recognition sites are often palindromic.

  • Transformation: competent cells are prepared by chemical (CaCl2) or electroporation; efficiency is reported as CFU/µg.

  • Sequencing: Sanger reads ~700-1000 bp; NGS short reads ~100-300 bp; long-read platforms improve assembly across repeats.

  • mRNA manufacturing: in vitro transcription, capping, polyadenylation, purification, and LNP encapsulation; dsRNA removal reduces immunogenicity.

  • Bioprocess: kLa (oxygen transfer) and mixing time are scale-dependent; scale-down models must reproduce these parameters.

  • Analytical: charge variants by capillary electrophoresis; aggregation by SEC; glycan profiles by HILIC.

Quick reference

Quality Assurance and Good Manufacturing Practice

  • Deviation investigation: contain, assess impact, identify root cause, implement CAPA, verify effectiveness, and close.

  • Change control: assess impact on regulatory filings, validation status, and product quality before approval.

  • Media fill acceptance: no contaminated units in the filled units; growth promotion of media is verified before use.

  • Supplier audits: critical suppliers audited every 1-3 years; audit reports and CAPA are tracked.

  • Recall classes: Class I (serious harm/death), Class II (temporary/reversible harm), Class III (unlikely harm).

  • Validation master plan and periodic review ensure qualified status of facilities, utilities, and equipment.

Quick reference

Ethics, Compliance, and Environmental Health and Safety

  • Belmont principles: respect for persons (autonomy), beneficence (maximize benefit, minimize harm), and justice (fair distribution).

  • Declaration of Helsinki: ethics committee review, informed consent, and risk minimization for human research.

  • 3Rs: replacement (non-animal methods), reduction (fewer animals), refinement (less pain/distress); humane endpoints are defined in protocols.

  • Conflict of interest: financial and non-financial interests are disclosed and managed; undisclosed conflicts undermine trust.

  • GHS: pictograms, signal words, hazard statements, and SDS sections 1-16; incompatible chemicals stored separately.

  • Environmental permits: air, water, and hazardous waste; exceedances require notification and corrective action.

Quick reference

Biosafety and Facility Management

  • BSC certification uses NSF/ANSI 49 or EN 12469 standards; the cabinet is decontaminated before and after service.

  • Negative pressure differentials: BSL-3 suites typically -10 to -50 Pa relative to corridors; alarms alert on loss of differential.

  • HEPA filter testing: aerosol challenge (PAO or DOP) with photometer scan; leak testing is performed annually and after change.

  • Autoclave validation: thermocouples map load temperature; biological indicators confirm lethality; records retained.

  • Spill kits are checked monthly; contents include absorbent, disinfectant, PPE, and disposal bags.

  • Biosecurity: access control, visitor logs, and pest control support both biosafety and security.

Quick reference

Biohazard Risk Assessment and Containment

  • Risk assessment is documented and reviewed annually or after any change in agent, procedure, or personnel.

  • Route of exposure determines controls: inhalation → BSC/respirator; percutaneous → sharps safety; ingestion → no mouth pipetting.

  • Aerosol-generating procedures include centrifugation, vortexing, and sonication; use sealed rotors and BSC.

  • PPE: gloves, gown, eye protection, and respiratory protection based on risk assessment; PPE is removed in a defined order.

  • Transport: Category A (UN2814) requires triple packaging and trained shippers; Category B is UN3373.

  • ISO 35001 provides a biorisk management system framework for laboratories.

Quick reference

Biological Resource Management and Quality Control

  • Identity testing: STR profiling for human lines; species PCR or karyotyping for non-human lines.

  • Viability: trypan blue or automated counting; <70% viability suggests poor cryopreservation or handling.

  • Adventitious agent testing: sterility, mycoplasma, in vitro/in vivo assays, and retrovirus testing for rodent lines.

  • Storage: -80 °C for working aliquots, LN2 vapor for long-term; avoid repeated freeze-thaw.

  • Reference materials: homogeneity, stability, and commutability are assessed before use.

  • Documentation: chain of custody, consent, and MTAs govern distribution and use.

Quick reference

Bioindustry Regulations and Standards

  • GMP inspections: routine, for-cause, and pre-approval; findings classified as critical, major, or minor (Form 483).

  • CTD modules: 1 regional, 2 quality, 3 nonclinical, 4 clinical, 5 references; used for IND/NDA/BLA and MAA.

  • ICH Q3A/Q3B impurity limits; Q6A/Q6B specifications and test procedures.

  • GDP: temperature-controlled transport, shipper qualification, and deviation handling.

  • Pharmacovigilance: serious unexpected suspected adverse reactions reported within 15 days (EU).

  • Device classification drives conformity assessment and premarket submission requirements.

Quick reference

Biotechnology Fundamentals and Applications

  • PCR: denaturation ~95 °C, annealing ~55-65 °C, extension ~72 °C; primer Tm and Mg2+ affect specificity.

  • qPCR: standard curve (absolute) or ΔΔCt (relative); Ct is inversely proportional to starting template.

  • Expression systems: E. coli (no glycosylation), yeast, CHO (human-like), insect cells (baculovirus).

  • Chromatography: affinity capture, ion exchange polish, size exclusion for aggregates/desalting.

  • Biosimilarity: analytical, functional, nonclinical, and clinical comparability; extrapolation requires justification.

  • Gene therapy vectors: AAV (non-integrating, limited cargo), lentivirus (integrating), adenovirus (transient, immunogenic).

Quick reference

Quality Assurance and Good Manufacturing Practice

  • QbD: QTPP, CQAs, CPPs, and design space are defined during development.

  • Process validation stages: process design, process qualification (PPQ), continued process verification (CPV).

  • CPV uses SPC charts and trend analysis to detect drift.

  • Cleaning validation: HBEL/PDE limits, visually clean criteria, and rinse/swab sampling.

  • Ongoing stability: at least one batch per product per year is tested.

  • Self-inspection and audits: scheduled intervals, findings tracked to closure with CAPA.

Quick reference

Ethics, Compliance, and Environmental Health and Safety

  • Research misconduct: fabrication, falsification, and plagiarism (FFP); honest error is not misconduct.

  • Authorship: substantial contribution, drafting/critical revision, final approval, and accountability; ghost/gift authorship prohibited.

  • Data privacy: GDPR and HIPAA govern personal and health data; de-identification and consent scope documented.

  • DURC review overlaps biosafety and ethics; publication may require mitigation or redaction.

  • Environmental monitoring: air, water, and soil sampling compared against permit limits.

  • Occupational health surveillance: pre-placement, periodic, and exit examinations for hazardous exposures.