- Bremsstrahlung radiation
- Continuous x-ray spectrum produced when projectile electrons decelerate in the anode target; the maximum photon energy equals the tube voltage in keV.
- Inverse square law
- Radiation intensity is inversely proportional to the square of the distance from the source; doubling distance reduces dose to one quarter.
- Anode heel effect
- Reduced beam intensity on the anode side of the field; thicker patient parts are placed toward the cathode side to compensate.
- Half-value layer
- Absorber thickness that reduces beam intensity by half; minimum total filtration is 2.5 mm Al equivalent above 70 kVp.
- Atlas and axis
- C1 (atlas) lacks a body and supports the skull; C2 (axis) provides the dens for head rotation.
- Right main bronchus
- Wider and more vertical than the left, making it the more likely path for aspirated material.
- Diaphragm level
- At full inspiration the domes lie near the 10th posterior or 6th anterior rib; the right hemidiaphragm is normally higher.
- AP versus PA projection
- AP enters anterior and exits posterior; PA enters posterior and exits anterior; projection is defined by the part-to-image relationship.
- PA chest positioning
- Patient stands with chest against receptor, shoulders rolled forward, chin elevated, and exposure at full inspiration.
- AP pelvis rotation
- Feet internally rotated 15 degrees to bring the femoral necks into profile; beam centered between ASIS and pubic symphysis.
- Spatial resolution
- Ability to distinguish small, closely spaced objects; described in line pairs per millimeter and limited by pixel and focal spot size.
- Deviation index
- Digital measure of receptor exposure relative to the reference; a value within plus or minus 1 indicates acceptable exposure.
- Detective quantum efficiency
- Combined measure of resolution and noise performance; higher DQE indicates better image quality per unit dose.
- Collimator alignment test
- Daily QC check that the light field aligns with the radiation field within 2 percent of the source-to-image distance on each side.
- kVp accuracy test
- Measured tube voltage must be within plus or minus 5 percent of the indicated value for acceptable performance.
- ALARA principle
- All exposures should be kept as low as reasonably achievable, considering economic and social factors.
- Occupational dose limits
- 20 mSv per year averaged over 5 years for whole body; 150 mSv for lens; 500 mSv for extremities and skin.
- Lead apron attenuation
- A 0.5 mm lead-equivalent apron attenuates about 90 percent of scattered radiation at diagnostic energies.
- Gray and sievert
- Gray measures absorbed dose (J/kg); sievert measures equivalent dose, equal to absorbed dose times the radiation weighting factor.
- Linear-quadratic model
- Cell survival equals exp(-alpha D minus beta D squared); the alpha/beta ratio indicates tissue repair capacity.
- Deterministic versus stochastic
- Deterministic effects have a threshold and increase in severity; stochastic effects have no threshold and increase in probability.
- Informed consent
- Patient agreement after explanation of procedure, benefits, risks, and alternatives, required for contrast and significant-risk procedures.
- Anaphylaxis management
- Stop injection, call for help, give intramuscular epinephrine, provide oxygen, and monitor vital signs.
- Scope of practice
- Procedures a radiographer is licensed to perform under Kuwait MOH regulations; practicing beyond it is prohibited.
- Hounsfield unit scale
- CT attenuation scale with water at 0 HU, air at -1000 HU, and bone above +1000 HU.
- MRI safety zones
- Zone I public, zone II reception, zone III restricted, zone IV magnet room; ferromagnetic objects must never enter zone IV.
- Doppler effect
- Frequency shift of reflected sound from moving blood; color Doppler shows direction and velocity, spectral Doppler quantifies flow.
- Technetium-99m half-life
- Six hours; widely used for bone and cardiac imaging because of favorable energy and short half-life.
- Therapeutic ratio
- Tumor control probability divided by normal tissue complication probability; fractionation improves it by exploiting repair differences.
- Hand hygiene
- Single most effective measure to prevent healthcare-associated infection; alcohol rub for routine use, soap and water when soiled.
- Airborne precautions
- Required for tuberculosis and measles; use an N95 respirator and a negative-pressure room.
- PDSA cycle
- Plan-do-study-act framework for continuous quality improvement; tests changes on a small scale before full implementation.
- DICOM and PACS
- DICOM is the standard for image storage and transmission; PACS stores and retrieves images across the enterprise.