ECG Technician Responsibilities and Scope of Practice

  • ECG technicians perform diagnostic ECG recordings and may include Holter, event monitoring, and stress testing under supervision.

  • Scope includes patient preparation, electrode placement, recording acquisition, artifact recognition, and basic rhythm recognition for reporting.

  • Technicians do not interpret or diagnose rhythms; they report findings to a physician or supervising clinician.

  • Patient education about the procedure is within scope; obtaining informed consent is typically the physician's responsibility.

  • Technicians must recognize life-threatening rhythms (e.g., ventricular fibrillation, asystole) and immediately alert the care team.

  • Infection control: hand hygiene and cleaning reusable electrodes between patients are standard responsibilities.

  • Documentation of the procedure, patient response, and any artifacts is a core technician duty.

  • Technicians must know and follow facility protocols for critical value reporting and escalation.

Quick reference

Patient Preparation, Positioning, and Electrode Placement

  • Standard 12-lead placement: V1 4th intercostal space right sternal border; V2 4th ICS left sternal border; V3 midway between V2 and V4; V4 5th ICS midclavicular line; V5 5th ICS anterior axillary line; V6 5th ICS midaxillary line.

  • Limb leads: RA right arm, LA left arm, RL right leg, LL left leg; avoid placing limb electrodes on bony prominences or over joints.

  • Prepare skin by cleaning with alcohol and drying; shave excessive hair if needed to reduce artifact.

  • Position patient supine or semi-recumbent; ensure limbs are relaxed and supported to minimize muscle artifact.

  • Apply electrodes firmly; check for expiration date and adequate gel; replace if dried or loose.

  • For pediatric patients, use appropriate electrode size and consider modified lead placement per protocol.

  • Document any deviations from standard placement and reason (e.g., amputation, wound).

Quick reference

ECG Recording, Equipment Operation, and Troubleshooting

  • Standard paper speed is 25 mm/s; standard gain is 10 mm/mV; verify these settings before recording.

  • Calibration pulse should be 10 mm high and 0.2 s wide (at 25 mm/s) to confirm proper gain and speed.

  • Common artifacts: baseline wander (respiration/movement), somatic tremor (muscle), AC interference (60 Hz), and poor electrode contact.

  • Troubleshoot artifact by checking electrode adhesion, skin preparation, cable connections, and patient relaxation.

  • If AC interference persists, check for nearby electrical equipment and ensure proper grounding.

  • Record a full 12-lead ECG with at least 10 seconds of rhythm strip (usually lead II) for rhythm analysis.

  • Label each lead clearly and ensure all leads are visible without overlap; adjust gain if necessary for large voltages.

  • Perform routine equipment checks and report malfunctions; do not use damaged cables or electrodes.

Quick reference

Rhythm Interpretation and Strip Labeling

  • Normal sinus rhythm: rate 60-100 bpm, regular, P wave before every QRS, PR interval 0.12-0.20 s, QRS <0.12 s.

  • Sinus bradycardia: rate <60 bpm; sinus tachycardia: rate >100 bpm; both with normal P waves.

  • Atrial fibrillation: irregularly irregular rhythm, absent P waves, fibrillatory baseline.

  • Ventricular tachycardia: wide QRS (>0.12 s), rate >100 bpm, often regular; treat as emergency if pulseless.

  • Ventricular fibrillation: chaotic, irregular waveform with no identifiable QRS; immediate defibrillation required.

  • Asystole: flat line; confirm lead connection and gain before declaring asystole.

  • Heart rate calculation: 300 / number of large boxes between R waves; or 1500 / number of small boxes.

  • Strip labeling must include patient name, date, time, lead, and any symptoms or interventions.

Quick reference

Documentation, Reporting, and Communication of Changes

  • Document all ECG findings objectively, including rate, rhythm, and any abnormalities observed, using standard terminology.

  • Report critical findings (e.g., new ST elevation, ventricular arrhythmias) immediately to the responsible clinician per protocol.

  • Use SBAR (Situation, Background, Assessment, Recommendation) for structured communication of changes.

  • Record patient symptoms (chest pain, palpitations, dyspnea) and vital signs at time of ECG.

  • Note any interventions (e.g., oxygen, medication) and patient response in the record.

  • Ensure timely entry of ECG data into the medical record; follow HIPAA and facility confidentiality policies.

  • Verify patient identification before reporting; include unique identifiers to prevent miscommunication.