NCCT NCET: ECG Recording, Equipment Operation, and Troubleshooting

NCCT NCET · Work through your exam, one topic at a time.

All guide sections

ECG Recording, Equipment Operation, and Troubleshooting

Domain 3 covers equipment readiness, standard calibration, acquisition, and artifact troubleshooting.

Before recording

  • Inspect cables, lead wires, electrodes, paper, and battery; confirm machine settings: 25 mm/s paper speed and 10 mm/mV gain unless ordered otherwise.
  • Enter patient data accurately; select the correct lead set; ensure filters are at facility default.

Artifact and fixes

  • Somatic tremor: muscle tension, shivering, talking-warm the room, relax the patient, reposition.
  • 60-cycle AC interference: thick, fuzzy baseline-check nearby electrical devices, cable loops, electrode contact, and grounding.
  • Wandering baseline: respiration, movement, poor adhesion, lotion-re-prep skin, secure electrodes, ask the patient to stay still.
  • Flat line in one lead: check that electrode and lead wire; all leads flat: check the patient, connections, power, and cable.
  • Interrupted tracing: loose wire, paper jam, low battery.

Worked example

A tracing shows a thick, fuzzy baseline across all leads at 60 cycles per second. The technician checks that the patient is not touching metal, moves a nearby infusion pump cord, re-preps a loose LL electrode, and repeats. The tracing clears. The technician does not change gain or paper speed to fix the appearance without an order and documentation.

Traps

Ignoring artifact; failing to verify calibration; using damaged cables; not documenting settings or deviations; repeating until the tracing looks normal.