Quick reference

Nuclear Medicine Technology

  • Radiopharmaceutical decay follows A = A0·e^(−λt), with λ = ln2 / T½; activity at time t is obtained by multiplying the calibrated activity by the decay factor e^(−0.693t/T½).

  • Precalibration: activity needed at use time is divided by the decay factor from calibration time to use time; Tc-99m (T½ ≈ 6.02 h) loses about half its activity every 6 hours.

  • Mo-99/Tc-99m generator: Mo-99 (T½ ≈ 66 h) decays to Tc-99m; transient equilibrium is re-established in roughly 4 half-lives of Tc-99m (~24 h), and eluate Mo-99 breakthrough limit is 0.15 kBq Mo-99 per MBq Tc-99m (0.15 µCi/mCi) at administration.

  • Alumina column generator elution uses 0.9% NaCl; the first eluate after delivery may contain higher Mo-99 and is typically discarded or assayed before clinical use.

  • Radionuclidic purity is verified with a dose calibrator or gamma camera; radiochemical purity is verified by chromatography (paper, TLC, or column) and must meet the monograph limit before patient administration.

  • Common T½ values: F-18 ≈ 110 min, Ga-68 ≈ 68 min, Tc-99m ≈ 6.02 h, I-123 ≈ 13.2 h, Tl-201 ≈ 73 h, In-111 ≈ 67 h, I-131 ≈ 8.02 d, Lu-177 ≈ 6.7 d.

  • Dose calibrator constancy is checked daily with a long-lived source (e.g., Cs-137) and accuracy at clinical energies is verified at required intervals; geometry (volume and syringe type) affects measured activity.

  • Injection site infiltration is confirmed by imaging the injection site; residual activity in the syringe and tubing must be subtracted when calculating the administered dose.

Quick reference

Nuclear Medicine Technology

  • Gamma camera: NaI(Tl) crystal converts gamma photons to light; PMTs convert light to electrical signal; the collimator defines the field of view and spatial resolution.

  • Collimator choice: low-energy high-resolution (LEHR) for Tc-99m; medium-energy for In-111 and Ga-67; high-energy for I-131; pinhole for small organs such as thyroid.

  • Energy resolution of NaI(Tl) is roughly 6–7% at 140 keV; a 20% photopeak window centered on 140 keV is standard for Tc-99m.

  • Spatial resolution improves with closer source-to-collimator distance; sensitivity and resolution trade off with collimator hole size and septal thickness.

  • Uniformity (flood) and energy-peak calibration are performed daily or per protocol; photomultiplier tube drift appears as ring or spot artifacts on uniformity images.

  • SPECT reconstruction uses filtered backprojection or iterative methods (OSEM); Butterworth and Hanning filters trade noise against resolution; attenuation correction improves myocardial and brain SPECT.

  • PET detection relies on annihilation photons (511 keV) detected in coincidence; time-of-flight and attenuation correction improve image quality; SUV = activity concentration / (injected dose / body weight).

Quick reference

Nuclear Medicine Technology

  • Cardiac perfusion: Tc-99m sestamibi or tetrofosmin for SPECT; Tl-201 for viability; stress/rest comparison identifies reversible ischemia versus fixed scar.

  • Bone scan: Tc-99m MDP localizes to osteoblastic activity; three-phase imaging (flow, blood pool, delayed) distinguishes cellulitis from osteomyelitis.

  • V/Q scan uses Tc-99m DTPA or Technegas aerosol for ventilation and Tc-99m MAA for perfusion; mismatched defects suggest pulmonary embolism (modified PIOPED criteria).

  • Renal scan: Tc-99m MAG3 for tubular secretion and ERPF; Tc-99m DTPA for GFR; captopril challenge for renovascular hypertension; Lasix washout for obstruction.

  • Thyroid: Tc-99m pertechnetate and I-123 localize to thyroid; I-131 used for therapy; pregnancy and breastfeeding are absolute contraindications to radioiodine.

  • Hepatobiliary: Tc-99m mebrofenin/IDA agents trace hepatocyte uptake and biliary excretion; GI bleed: Tc-99m sulfur colloid or tagged RBCs.

  • Sentinel node and infection imaging: Tc-99m sulfur colloid for lymphoscintigraphy; Ga-67 and In-111 WBC for infection/inflammation; F-18 FDG PET for oncology staging.

Quick reference

Nuclear Medicine Technology

  • ALARA: occupational dose limit is 50 mSv/year (5 rem) effective dose; lens of eye 150 mSv/year; extremity/skin 500 mSv/year under ICRP; public limit 1 mSv/year.

  • Pregnant declared worker: dose to fetus limited to 0.5 mSv per month (5 mSv over gestation); a declared pregnant worker may continue if limits are met.

  • Time, distance, and shielding: dose rate follows inverse-square law; doubling distance reduces exposure to one quarter; lead half-value layer for Tc-99m is about 0.3 mm Pb.

  • Personnel monitoring uses TLD or OSL badges worn at collar level (and ring badges for high hand exposure); badges are exchanged per institutional interval and not worn outside work.

  • Spills: contain, mark, and decontaminate; survey with a thin-window Geiger-Mueller or NaI probe; record removable contamination in Bq/cm² or dpm/100 cm² per license conditions.

  • Radioactive waste is held for decay-in-storage until activity is indistinguishable from background (commonly 10 half-lives) before disposal as non-radioactive; survey and document before release.

  • Patient release after therapy follows NRC/agreement-state criteria (e.g., 1.11 GBq I-131 or dose-rate limits) with written instructions; breastfeeding must stop after I-131 therapy.

Quick reference

Nuclear Medicine Technology

  • Before any study, verify patient identity, pregnancy/breastfeeding status, allergies, renal function, and prior nuclear medicine or contrast procedures.

  • Radiopharmaceuticals are administered under an authorized user's order; the package insert and institutional protocol define dose, route, and timing.

  • Patient preparation is agent-specific: fasting for FDG PET, thyroid blockade or hydration for I-131 therapy, voiding before pelvic imaging, and hydration for renal scans.

  • For FDG PET, blood glucose should be controlled (commonly < 11.1 mmol/L or per protocol), uptake period is about 60 min, and the patient rests quietly to avoid muscle uptake.

  • Document administered activity, route, time, lot number, and any residual activity in the syringe; reconcile the prescribed and administered dose after injection.

  • Recognize and manage adverse reactions: anaphylactoid responses to radiopharmaceuticals are rare; vasovagal reactions and extravasation require prompt assessment and documentation.

  • Quality control: daily dose-calibrator constancy, generator eluate assay, radiochemical purity testing, and camera flood/energy checks must be recorded before patient imaging.

Quick reference

Nuclear Medicine Technology

  • Image acquisition modes: static, dynamic, whole-body, SPECT, and gated; matrix size and zoom must match the required resolution and field of view.

  • Count statistics: noise decreases with the square root of counts; doubling counts improves signal-to-noise by about 1.41.

  • Reconstruction: filtered backprojection is fast but amplifies noise; iterative OSEM with attenuation and scatter correction is preferred for quantitative SPECT.

  • Registration and fusion of SPECT/CT or PET/CT improve localization; CT-based attenuation correction uses a bilinear conversion of Hounsfield units to attenuation coefficients.

  • Quantitation: SUV, standardized uptake ratio, and time-activity curves require decay correction, scanner calibration, and consistent uptake time.

  • Motion artifact (patient movement, respiration) produces blurring or misregistration; gating and patient immobilization reduce it.

  • Display and archiving follow DICOM standards; window/level, color maps, and annotation must be documented so images are interpretable and reproducible.