NMAA flashcards
NMAA · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Tc-99m half-life
- Approximately 6.02 hours. About half the activity remains after 6 hours and about 6% after 24 hours, so elution timing and dose preparation must account for decay.
- Mo-99 breakthrough limit
- The eluate must contain no more than 0.15 kBq of Mo-99 per MBq of Tc-99m (0.15 microcurie per millicurie) at the time of patient administration.
- Decay formula
- Activity at time t equals initial activity multiplied by e raised to the power negative 0.693 t divided by half-life; used for precalibration and post-calibration corrections.
- Radiochemical purity
- The fraction of total radioactivity present in the desired chemical form. It is measured by chromatography and must meet the product monograph before administration.
- Radionuclidic purity
- The fraction of total radioactivity attributable to the intended radionuclide. It is assessed by gamma spectroscopy or by assaying for contaminants such as Mo-99.
- Dose calibrator constancy
- A daily check with a long-lived reference source, such as Cs-137, that verifies the instrument response is stable within the established tolerance before patient doses are measured.
- NaI(Tl) crystal function
- Thallium-activated sodium iodide absorbs gamma photons and emits visible light proportional to deposited energy; photomultiplier tubes convert the light into an electrical pulse.
- Photopeak window
- An energy acceptance range centered on the gamma emission, commonly 20% for Tc-99m at 140 keV, that excludes scattered photons and reduces image noise.
- Collimator selection rule
- Use low-energy high-resolution collimators for Tc-99m, medium-energy for In-111 and Ga-67, high-energy for I-131, and a pinhole collimator for small organs such as the thyroid.
- SPECT iterative reconstruction
- OSEM reconstruction models the imaging process and applies attenuation and scatter correction, yielding lower noise and better quantitation than filtered backprojection in most clinical settings.
- PET coincidence detection
- Two 511 keV annihilation photons emitted about 180 degrees apart are recorded only when detected within a narrow timing window, which localizes the annihilation event along a line of response.
- SUV definition
- Standardized uptake value equals decay-corrected activity concentration in tissue divided by injected activity per unit body weight; consistent uptake time and calibration are required for comparability.
- Tc-99m MDP bone scan
- The diphosphonate tracer adsorbs to hydroxyapatite in proportion to osteoblastic activity and blood flow, showing metastases, fractures, and infection as focal increased uptake.
- Three-phase bone scan
- Flow, blood-pool, and delayed images separate hyperemia from osteoblastic activity; focal delayed uptake with normal flow suggests osteomyelitis rather than cellulitis.
- V/Q mismatch significance
- Perfusion defects with preserved ventilation in a segment suggest pulmonary embolism; matched defects are more typical of parenchymal lung disease such as COPD.
- Tc-99m MAG3 use
- A tubular secretion agent used for renal perfusion and function imaging, including Lasix washout for obstruction and captopril challenge for renovascular hypertension.
- Tc-99m sulfur colloid
- A particulate agent taken up by the reticuloendothelial system; used for liver/spleen imaging, bone marrow imaging, gastrointestinal bleeding, and sentinel lymph node localization.
- I-131 therapy precautions
- Patients receive written radiation safety instructions, avoid prolonged close contact with others, and must stop breastfeeding; pregnancy is an absolute contraindication.
- F-18 FDG patient preparation
- The patient fasts, blood glucose is checked against protocol, and after injection rests quietly for about 60 minutes to limit physiologic muscle and brown fat uptake.
- Extravasation documentation
- Infiltration at the injection site is confirmed by imaging the site; residual syringe activity is subtracted and the event is recorded and reported per institutional policy.
- Occupational dose limit
- The annual effective dose limit for radiation workers is 50 mSv (5 rem); the lens of the eye limit is 150 mSv and the extremity or skin limit is 500 mSv.
- Inverse-square law
- Radiation exposure rate from a point source decreases with the square of distance; doubling the distance from a source reduces the exposure rate to one quarter.
- Declared pregnant worker
- A worker who formally declares pregnancy is subject to a fetal dose limit of 0.5 mSv per month, and workload may be adjusted to stay within that limit.
- Decay-in-storage
- Radioactive waste is held until its activity is indistinguishable from background, commonly after about 10 half-lives, then surveyed and documented before disposal as non-radioactive.
- Personnel dosimeter placement
- A whole-body TLD or OSL badge is worn at collar level outside protective clothing; ring badges are worn when hand exposure is likely to approach dose limits.
- Pregnancy screening
- Pregnancy or breastfeeding status is documented before radiopharmaceutical administration, because radioiodine and many agents require postponement or alternative imaging.
- Administered dose reconciliation
- The activity remaining in syringe, tubing, and injection site is measured or estimated and subtracted from the prepared activity so the recorded dose reflects actual patient uptake.
- Thyroid uptake agents
- Tc-99m pertechnetate and I-123 are used for thyroid imaging and uptake; I-131 is reserved for therapy and for selected diagnostic or ablation follow-up studies.
- Renal scan timing agents
- Tc-99m DTPA is filtered and estimates glomerular filtration rate; Tc-99m MAG3 is secreted by tubules and estimates effective renal plasma flow with better image quality.
- Count statistics rule
- Image noise is inversely proportional to the square root of collected counts, so quadrupling acquisition counts halves relative noise and improves image quality.
- CT attenuation correction
- Hounsfield units from the CT component are converted to attenuation coefficients at the emission energy, enabling quantitative SPECT or PET and improved lesion localization.
- Generator elution principle
- Saline elutes Tc-99m from the alumina column as pertechnetate; ingrowth of Tc-99m from Mo-99 reaches transient equilibrium in about four Tc-99m half-lives.