NMTCB PET study materials
NMTCB PET study resources

NMTCB PET cheat sheet

NMTCB PET · Keep the key ideas within reach.

Quick reference

PET Instrumentation and Physics

  • Annihilation of a positron with an electron produces two 511 keV photons emitted ~180° apart; the small angular deviation (<1°) plus positron range blurs spatial resolution.

  • Detector block = scintillation crystal array coupled to photomultiplier tubes; Anger logic computes the interaction position from PMT signal ratios.

  • Time-of-flight PET requires fast scintillators (e.g., LSO, LYSO) with coincidence timing resolution in the hundreds of picoseconds, improving SNR rather than spatial resolution.

  • Coincidence timing window is typically 4–12 ns; a narrower window rejects random coincidences but risks losing true events from detector timing jitter.

  • Random coincidence rate ≈ 2·τ·R1·R2 (τ = timing window); randoms rise with the square of activity, so they dominate at high count rates.

  • Sensitivity is the fraction of emitted annihilation pairs detected; it depends on solid angle, detector thickness, and energy window, not on injected dose.

  • NECR = T²/(T + S + 2R) quantifies noise-equivalent count rate; it peaks at moderate activity and falls as randoms and dead time grow.

  • Scatter correction methods include energy-window-based (dual/triple energy windows), model-based convolution-subtraction, and Monte Carlo simulation.

  • Attenuation correction uses transmission sources (e.g., rotating 68Ge line source) or CT-based mu-maps; CT mu-maps must be converted from CT numbers to 511 keV attenuation coefficients.

Quick reference

PET Radiopharmaceuticals

  • 18F-FDG is taken up by glucose transporters and trapped after hexokinase phosphorylation (FDG-6-phosphate); it is not further metabolized in most tissues.

  • 18F has a 109.8-minute half-life and is produced in a cyclotron via the 18O(p,n)18F reaction on enriched 18O-water.

  • 68Ga has a 67.7-minute half-life and is eluted from a 68Ge/68Ga generator; 68Ge (271 d) breaks through gradually and must be tested.

  • 68Ga-DOTATATE binds somatostatin receptors (SSTR2) and is used for neuroendocrine tumors; physiologic uptake occurs in spleen, liver, pituitary, and adrenals.

  • 18F-NaF localizes to bone remodeling sites via fluoride exchange with hydroxylapatite; it is used for bone metastases and is cleared renally.

  • 11C has a 20.4-minute half-life and requires on-site cyclotron; 15O has a 2.0-minute half-life and is used for perfusion studies.

  • Radiochemical purity is verified by chromatography (e.g., ITLC or HPLC); 18F-FDG USP requires ≥90% radiochemical purity and pH 4.5–7.5.

  • FDG uptake is suppressed by hyperglycemia and insulin; patients fast ≥4–6 hours and glucose should be <200 mg/dL before injection.

Quick reference

PET Clinical Imaging Procedures

  • Standard FDG oncology dose is 10–20 mCi (370–740 MBq) IV; uptake period is 60 minutes (range 45–75), and imaging follows bladder emptying.

  • Patient preparation: fast ≥4–6 hours, avoid strenuous exercise 24 hours before, and withhold caffeine if protocol requires; document glucose level.

  • FDG normal biodistribution: brain, myocardium (variable), liver, spleen, kidneys, ureters, bladder, and GI tract; brown fat in supraclavicular/paravertebral regions.

  • CT parameters for attenuation correction use low-dose settings (e.g., 120 kVp, 30–80 mAs); diagnostic CT may be added for lesion characterization.

  • SUV = tissue activity concentration / (injected dose / body weight); lean-body-mass or BSA normalization reduces variability in obese patients.

  • A negative FDG PET does not exclude malignancy; false negatives occur with small lesions (<1 cm), low metabolic rate tumors, and hyperglycemia.

  • Inflammatory/infectious uptake, granulomas, and postsurgical changes can mimic malignancy; correlate with history and prior imaging.

  • Myocardial FDG imaging for viability requires dietary preparation (high-fat, low-carbohydrate) plus prolonged fasting to shift metabolism to fatty acids.

Quick reference

PET Quality Control

  • Daily QC: detector uniformity (blank scan or uniform phantom), energy window peak position, and coincidence timing check before patient imaging.

  • Weekly/monthly QC: detector gain and timing drift, crystal energy resolution, and coincidence timing resolution verification per manufacturer schedule.

  • Normalization scan corrects for detector efficiency variations; it must be acquired after detector changes or significant repairs.

  • Well counter QC: daily background and constancy checks with a long-lived source (e.g., 137Cs or 68Ge); chi-square test verifies counting statistics.

  • Dose calibrator QC: daily constancy with a reference source, annual accuracy and linearity checks, and geometry/volume correction.

  • Uniformity phantom scan should show <5% variation across the FOV; ring artifacts suggest a defective detector or normalization error.

  • Spatial resolution is measured with a point source or line source; FWHM in reconstructed images is typically 4–6 mm for modern scanners.

  • PET/CT alignment QC uses a point source at multiple bed positions to verify CT and PET coordinate registration within 1–2 mm.

Quick reference

PET Safety, Radiation Protection, and Regulations

  • Occupational dose limits: 50 mSv/year (5 rem) effective dose; lens 150 mSv/year; extremities 500 mSv/year.

  • Public dose limit is 1 mSv/year (100 mrem); ALARA principles require time, distance, and shielding optimization.

  • Inverse-square law: exposure decreases with the square of distance; doubling distance reduces dose to one-quarter.

  • Time, distance, and shielding: lead aprons attenuate 511 keV poorly (~10–20% with 0.5 mm Pb), so distance is the primary protection.

  • Pregnant workers may declare pregnancy and are limited to 0.5 mSv (500 mrem) to the fetus over the gestation period.

  • Radioactive package receipt requires survey of external dose rate, contamination wipe test, and documentation of package labeling.

  • Patient release criteria: dose rate at 1 meter typically <7 mrem/h; written instructions must be provided for breastfeeding and close contact.

  • Medical event (misadministration) reporting: notify NRC within 24 hours if dose differs by ≥20% from prescribed or if wrong radiopharmaceutical is administered.

  • PET radiopharmaceuticals are regulated under 10 CFR Part 35; written directives and authorized user supervision are required for administration.