CT flashcards
CT · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Hounsfield unit formula
- HU = 1000 x (mu_tissue - mu_water) / mu_water. Water is 0 HU, air is -1000 HU, and dense bone is typically +1000 HU or higher. Values depend on beam energy and scanner calibration.
- CTDIvol
- Volume CT dose index estimates average dose per slice in a phantom, accounting for pitch. It is reported by the scanner and used with scan length to calculate DLP and effective dose.
- DLP calculation
- Dose-length product equals CTDIvol multiplied by scan length in centimeters. It reflects total radiation output for a series and is used to estimate effective dose with a body-region k-factor.
- Pitch definition
- Pitch equals table feed per gantry rotation divided by total beam collimation. Pitch greater than 1 reduces dose and scan time but may increase noise; pitch less than 1 increases dose.
- Window width and level
- Window level sets the center HU of the display range; window width sets the range of HU values shown. Narrow width increases contrast; wide width displays bone and soft tissue together.
- Metformin and contrast
- For patients with eGFR below 30 or acute kidney injury risk, hold metformin for 48 hours after iodinated contrast and resume only after renal function is confirmed stable.
- Contrast extravasation response
- Stop the injection, remove the cannula, elevate the limb, and apply a cold compress unless a warm-compress or hyaluronidase protocol is specified. Document volume and monitor for compartment syndrome.
- Anaphylactoid reaction steps
- Stop contrast administration, call for help, maintain the airway, give oxygen, and administer epinephrine per protocol for severe reactions. Monitor vital signs and document timing of events.
- Pediatric dose reduction
- Use weight- or size-based kV and mAs, limit scan length, avoid multiphase scanning, and apply iterative reconstruction. These measures lower dose while preserving diagnostic image quality in children.
- Effective dose formula
- Effective dose in mSv equals DLP multiplied by a body-region k-factor, approximately 0.015 for body and 0.0021 for head. It estimates stochastic risk from a CT study.
- Helical pitch effect
- Increasing pitch speeds table coverage and lowers dose but may raise image noise. Decreasing pitch increases overlap and dose while improving z-axis sampling and reformat quality.
- Multidetector row CT
- Multiple detector rows allow thinner slices, faster coverage, and improved z-axis resolution. Isotropic voxels enable high-quality multiplanar reformats and three-dimensional reconstructions.
- Contrast phase timing
- Non-contrast, arterial, venous, and delayed phases answer different clinical questions. Timing depends on circulation and injection rate; incorrect timing can miss vascular or tumor enhancement.
- Iterative reconstruction
- Iterative reconstruction reduces image noise and allows lower dose compared with filtered back projection. It can alter image texture, so protocols and kernels must be validated for each anatomy.
- Beam hardening artifact
- Dense bone or metal absorbs low-energy photons, causing streaks and dark bands. Centering the patient, using appropriate kV, and applying metal artifact reduction can reduce the effect.
- Motion artifact
- Patient movement or breathing during acquisition causes blurring and streaks. Clear breath-hold instructions, immobilization, and faster scan times reduce motion artifacts in CT.
- Isocenter positioning
- Centering the patient at gantry isocenter improves dose distribution and image quality. Off-center positioning can increase dose and produce shading or streak artifacts.
- X-ray production
- Electrons accelerated by high voltage strike a tungsten target, producing bremsstrahlung and characteristic radiation. kVp controls beam energy and penetration; mAs controls photon quantity.
- Exponential attenuation
- Intensity after attenuation follows I = I0 e^(-mu x). The linear attenuation coefficient depends on tissue density, atomic number, and photon energy; higher energy reduces photoelectric absorption.
- Photoelectric effect
- The photoelectric effect dominates at low photon energies and in high-atomic-number tissues. It increases patient dose and image contrast; higher kVp reduces its contribution.
- Compton scatter
- Compton scatter occurs when photons interact with outer-shell electrons, changing direction and contributing to image noise. Higher energy increases Compton relative to photoelectric interactions.
- Detector types
- Scintillation detectors convert x-rays to light and then to electrical signal. Solid-state and gas detectors differ in efficiency, stability, and afterglow, affecting image quality and calibration.
- Spatial resolution
- Spatial resolution is the ability to distinguish small high-contrast objects, measured in line pairs per centimeter. It depends on detector size, focal spot, slice thickness, and reconstruction.
- Contrast resolution
- Contrast resolution is the ability to distinguish tissues with small attenuation differences. It improves with higher dose, lower noise, and appropriate window settings and reconstruction kernels.
- Reconstruction kernels
- Soft-tissue kernels reduce noise but blur edges; bone kernels sharpen edges but increase noise. Kernel choice should match the anatomy and clinical question being evaluated.
- Slice thickness tradeoff
- Thinner slices improve z-axis resolution but increase noise and data volume. Overlapping reconstruction intervals improve multiplanar reformats without additional patient dose.
- Partial volume artifact
- Partial volume occurs when a voxel contains tissues of different attenuation, averaging their values. Thin slices and appropriate reconstruction reduce partial volume effects in small structures.
- Ring artifact
- Ring artifacts appear as concentric circles from detector calibration errors or a faulty channel. Phantom quality control scans can detect and localize the defective detector element.
- Quality control phantom
- A uniform phantom checks CT number accuracy, noise, and uniformity. Daily or periodic scans detect calibration drift and artifacts before they affect clinical images.
- Diagnostic reference levels
- Diagnostic reference levels are local or national dose benchmarks for common studies. Protocols exceeding them should be reviewed and optimized to keep dose as low as reasonably achievable.
- Pregnancy and CT
- CT is justified when medically necessary during pregnancy; diagnostic fetal risk is far below deterministic thresholds. Document justification, limit scan length, and use dose-reduction techniques.
- Patient breath-hold instruction
- Clear breath-hold instructions and confirmation of understanding reduce motion artifact. Explain warmth or flushing from contrast and the need to remain still before the scout image.