CT key terms
CT · Get comfortable with the vocabulary.
- CTDIvol (volume computed tomography dose index)
- Related: Pitch, DLP (dose-length product), ALARA
A standardized dose descriptor that estimates the average absorbed dose within a scanned volume for one rotation, accounting for pitch. It is displayed on the scanner console and used to compare protocols, but it does not represent patient size or the dose to a specific organ.
- DLP (dose-length product)
- Related: CTDIvol (volume computed tomography dose index), Effective dose, ALARA
The product of CTDIvol and the scanned length in centimeters, expressed in mGy·cm. It reflects total radiation output over a whole examination and is converted to effective dose using a body-region conversion coefficient.
- ALARA
- Related: Automatic exposure control (AEC) in CT, CTDIvol (volume computed tomography dose index), Diagnostic reference level (DRL)
The principle of keeping radiation exposure as low as reasonably achievable while still obtaining diagnostic image quality. In CT it drives protocol optimization, limiting scan length, lowering tube current where possible, and avoiding unnecessary multiphase studies.
- Diagnostic reference level (DRL)
- Related: CTDIvol (volume computed tomography dose index), DLP (dose-length product), ALARA
A benchmark dose value, usually set at the 75th percentile of a survey of facilities, used to flag protocols that deliver unusually high dose for a given examination. Exceeding a DRL prompts review rather than proving unsafe practice.
- Automatic exposure control (AEC) in CT
- Related: Noise (image quality), Pitch, ALARA
A system that modulates tube current during a scan, using a scout image or a localizer to estimate patient attenuation and adjust milliamperage per rotation. It reduces dose and improves noise uniformity across body regions of differing thickness.
- Pitch
- Related: Helical (spiral) scanning, CTDIvol (volume computed tomography dose index), Automatic exposure control (AEC) in CT
For helical scanning, the table feed per gantry rotation divided by the total collimated beam width. A pitch below 1 overlaps data and increases dose; a pitch above 1 reduces dose but may increase noise and artifacts.
- Helical (spiral) scanning
- Related: Pitch, Multiplanar reconstruction (MPR), Step-and-shoot (axial) scanning
A acquisition mode in which the table moves continuously while the tube and detectors rotate, producing a volume of projection data. Images are reconstructed at any position along the volume, enabling overlapping slices and multiplanar reformats without gaps.
- Step-and-shoot (axial) scanning
- Related: Helical (spiral) scanning, Prospective ECG gating, Pitch
A mode in which the table remains stationary during each rotation and advances between rotations. It avoids the continuous tube output of helical scanning and is common in cardiac gating and low-dose screening protocols.
- Hounsfield unit (HU)
- Related: Window width and window level, Linear attenuation coefficient, Region of interest (ROI) measurement
A scaled attenuation value in which water is 0 and air is approximately -1000, with dense bone reaching well above +1000. It allows tissue characterization and window setting independent of the display device.
- Window width and window level
- Related: Hounsfield unit (HU), Windowing artifact (clipping), Region of interest (ROI) measurement
Display settings that map the CT number range to visible gray shades. Level sets the center of the displayed range and width sets its span; narrow widths increase contrast but reduce the range of densities shown.
- Linear attenuation coefficient
- Related: Hounsfield unit (HU), Beam hardening, Photoelectric effect and Compton scatter
A measure of how much a material reduces X-ray intensity per unit path length, depending on photon energy, density, and atomic number. CT reconstruction converts measured attenuation into tissue values relative to water.
- Photoelectric effect and Compton scatter
- Related: Linear attenuation coefficient, Beam hardening, Iodinated contrast media
The two dominant X-ray interaction processes in the diagnostic energy range. Photoelectric absorption rises sharply with atomic number and low energy, giving bone and iodine high contrast; Compton scatter depends mainly on electron density and contributes to image noise.
- Beam hardening
- Related: Linear attenuation coefficient, Metal artifact, Dual-energy CT
The increase in mean photon energy as low-energy photons are preferentially absorbed while passing through tissue. It produces cupping and dark streaks between dense structures, corrected by filtration, calibration, and iterative or dual-energy methods.
- Noise (image quality)
- Related: Signal-to-noise ratio (SNR), Reconstruction kernel, Automatic exposure control (AEC) in CT
Random variation in pixel values that gives images a grainy appearance, quantified as the standard deviation of CT numbers in a uniform region. Noise falls as tube current, slice thickness, or reconstruction kernel smoothing increases.
- Signal-to-noise ratio (SNR)
- Related: Noise (image quality), Contrast-to-noise ratio (CNR), Slice thickness and collimation
The ratio of the difference in CT number between a structure and its background to the image noise. Higher SNR indicates a structure is more reliably detectable, and it can be improved with more photons or thicker slices.
- Contrast-to-noise ratio (CNR)
- Related: Signal-to-noise ratio (SNR), Noise (image quality), Window width and window level
The difference in CT number between two tissues divided by the background noise. It expresses how visible a lesion is relative to surrounding tissue and is improved by contrast timing, photon count, and appropriate windowing.
- Reconstruction kernel (algorithm)
- Related: Noise (image quality), Iterative reconstruction, Spatial resolution
A mathematical filter applied during image reconstruction that shapes spatial resolution and noise texture. Sharp or bone kernels enhance edges at the cost of noise, while smooth or soft-tissue kernels reduce noise and blur fine detail.
- Iterative reconstruction
- Related: Reconstruction kernel (algorithm), Noise (image quality), Filtered back projection
A reconstruction approach that repeatedly compares modeled projections with measured data to reduce noise while preserving detail. It permits lower tube current than filtered back projection but can alter noise texture and is vendor-specific.
- Filtered back projection
- Related: Iterative reconstruction, Reconstruction kernel (algorithm), Fourier transform reconstruction
The classical CT reconstruction method that back-projects filtered projection data to form an image. It is fast and well understood but amplifies noise, which newer iterative and model-based methods address.
- Spatial resolution
- Related: Reconstruction kernel (algorithm), Slice thickness and collimation, Partial volume averaging
The ability to distinguish two small, closely spaced objects as separate, often expressed as line pairs per centimeter or millimeters. In CT it is limited by detector aperture, focal spot size, slice thickness, and reconstruction kernel.
- Slice thickness and collimation
- Related: Partial volume averaging, Spatial resolution, Z-axis (longitudinal) resolution
The nominal thickness of a reconstructed image and the physical width of the X-ray beam defined by collimators. Thinner slices improve z-axis resolution and reduce partial volume averaging but increase noise and dose per image.
- Partial volume averaging
- Related: Slice thickness and collimation, Spatial resolution, Hounsfield unit (HU)
Blending of attenuation values when a voxel contains more than one tissue type, causing a structure to take an intermediate CT number. It is reduced by thinner slices and is a common cause of missed small lesions.
- Z-axis (longitudinal) resolution
- Related: Slice thickness and collimation, Multiplanar reconstruction (MPR), Partial volume averaging
The ability to resolve structures along the patient's long axis, determined mainly by slice thickness and detector configuration. It is critical for reformats, small-lesion detection, and accurate volume measurement.
- Multiplanar reconstruction (MPR)
- Related: Z-axis (longitudinal) resolution, Helical (spiral) scanning, Maximum intensity projection (MIP)
Creation of coronal, sagittal, or oblique images from a stack of thin axial slices. Quality depends on thin slices and isotropic voxels, and MPR is routine in trauma, vascular, and orthopedic protocols.
- Maximum intensity projection (MIP)
- Related: Multiplanar reconstruction (MPR), Volume rendering, CT angiography (CTA)
A rendering technique that displays the brightest voxel along each line of sight, making vessels and high-density structures stand out. It is widely used for CT angiography and for highlighting lung nodules or bone.
- Volume rendering
- Related: Maximum intensity projection (MIP), Multiplanar reconstruction (MPR)
A three-dimensional display method that assigns color and opacity to voxel values to depict surfaces and internal structures. It aids surgical planning and communication but is a visualization tool, not primary diagnostic data.
- Region of interest (ROI) measurement
- Related: Hounsfield unit (HU), Noise (image quality), Window width and window level
A drawn area in which mean CT number and standard deviation are calculated. It is used to characterize lesions, verify contrast enhancement, and measure image noise in quality assurance.
- Iodinated contrast media
- Related: Contrast-induced acute kidney injury (CI-AKI), Extravasation, CT angiography (CTA)
Intravascular or oral agents containing iodine that increase X-ray attenuation, used to opacify vessels and enhance tissue. Risks include contrast-induced kidney injury and allergic-like reactions, and patients require screening before administration.
- Contrast-induced acute kidney injury (CI-AKI)
- Related: Iodinated contrast media, Extravasation, Informed consent
An acute rise in serum creatinine after iodinated contrast administration, typically within days. Risk is higher with pre-existing renal impairment, diabetes, dehydration, and high contrast volume, so renal function is checked and volume minimized.
- Extravasation
- Related: Iodinated contrast media, Power injector, Patient positioning and immobilization
Leakage of contrast from a vein into surrounding soft tissue during injection, causing pain, swelling, and rarely compartment syndrome. Management includes stopping the injection, elevating the limb, cold compresses, and monitoring.
- Informed consent
- Related: Iodinated contrast media, Pregnancy screening in CT
The process of explaining a procedure's purpose, risks, benefits, and alternatives so the patient can decide voluntarily. For CT it covers contrast risks, radiation, and alternatives, and is documented before the examination.
- Pregnancy screening in CT
- Related: ALARA, Informed consent, Effective dose
The practice of asking patients of childbearing age about possible pregnancy before scanning and weighing whether the examination is justified. If needed, the scan is modified to limit dose to the pelvis and fetus while preserving diagnostic value.
- Effective dose
- Related: DLP (dose-length product), Pregnancy screening in CT, ALARA
A whole-body equivalent dose that reflects the relative radiation risk of a partial-body exposure, expressed in millisieverts. It allows comparison across modalities but is a population estimate, not a patient-specific risk.
- Metal artifact
- Related: Beam hardening, Motion artifact, Dual-energy CT
Bright and dark streaks radiating from dense implanted material such as hardware, dental fillings, or clips. It degrades adjacent tissue visibility and is reduced by thinner slices, higher tube output, and dedicated metal-reduction algorithms.
- Motion artifact
- Related: Patient positioning and immobilization, Prospective ECG gating, Metal artifact
Blurring, ghosting, or misregistration caused by patient movement, respiration, or cardiac motion during acquisition. It is limited by breath-hold coaching, immobilization, faster gantry rotation, and ECG gating.
- Prospective ECG gating
- Related: Retrospective ECG gating, Step-and-shoot (axial) scanning, Motion artifact
A cardiac CT technique in which the tube is triggered during a selected phase of the cardiac cycle, usually diastole. It lowers dose compared with retrospective gating but requires a stable, slow heart rate and regular rhythm.
- Retrospective ECG gating
- Related: Prospective ECG gating, Helical (spiral) scanning, Motion artifact
A cardiac CT technique in which projection data are acquired continuously and images are reconstructed at chosen cardiac phases. It allows functional assessment and works with irregular rhythms but delivers a higher radiation dose.
- CT angiography (CTA)
- Related: Bolus tracking (test bolus), Iodinated contrast media, Maximum intensity projection (MIP)
A contrast-enhanced study timed so that arteries are opacified during acquisition, often followed by thin-slice reformats. It requires accurate bolus timing or bolus tracking and careful attention to renal function and radiation dose.
- Bolus tracking (test bolus)
- Related: CT angiography (CTA), Power injector, Iodinated contrast media
A method for timing contrast acquisition by monitoring a vessel's CT number after injection and triggering when a threshold is reached, or by using a small test injection. It helps ensure peak arterial enhancement during the scan.