- Mayneord F factor
- Ratio used to convert percent depth dose from one source-to-surface distance to another; it multiplies squared inverse-square terms evaluated at maximum depth and at the depth of interest.
- Paddick conformity index
- Target volume inside the prescription isodose, squared, divided by the product of total target volume and prescription isodose volume; it penalizes both undercoverage and dose spill outside the target.
- Gamma index
- Patient-specific QA metric combining dose difference and distance-to-agreement; a point passes if a nearby point satisfies both criteria, and typical intensity-modulated criteria are three percent dose and three millimeters distance.
- Biologically effective dose
- BED equals n times d times one plus d divided by the alpha-beta ratio, where n is the number of fractions and d is dose per fraction; it compares fractionation schemes for a given tissue.
- Equivalent square field size
- A square field with the same collimator scatter as a rectangular field, estimated as four times area divided by perimeter; it lets tabulated square-field output and depth-dose data be applied.
- TERMA
- Total energy released per unit mass by primary photons within a voxel; convolution-superposition algorithms deposit this energy using a dose-spread kernel to obtain absorbed dose.
- Acuros XB
- Deterministic linear Boltzmann transport equation solver that computes dose in heterogeneous media more accurately than pencil-beam methods, particularly for lung, bone, and small radiation fields.
- Temperature-pressure correction
- P_TP equals 760 divided by pressure in millimeters of mercury, multiplied by 273.15 plus temperature in Celsius divided by 295.15; it corrects ion chamber readings to reference conditions.
- Two-voltage method
- Determines the ion chamber recombination correction by measuring collected charge at two bias voltages, usually full and half, and solving the standard quadratic expression from their ratio.
- Winston-Lutz test
- Verifies coincidence of radiation and mechanical isocenters using a small collimated field and a marker at known couch positions across multiple gantry, collimator, and couch angles; tolerance is about one millimeter.
- TG-142 daily output tolerance
- Daily photon and electron output constancy must stay within three percent of baseline, while the monthly calibration tolerance is two percent and the annual tolerance is one percent.
- Picket fence test
- Imaging test in which multileaf collimator leaves form alternate closed and open strips; it detects leaf position errors, calibration drift, and abnormal interleaf leakage.
- Dosimetric leaf gap
- Treatment planning system parameter that compensates for rounded leaf ends and leaf transmission in a multileaf collimator model; it is tuned by comparing calculated with measured doses.
- Energy constancy tolerance
- Per TG-142, photon beam energy constancy must remain within two percent of baseline and electron energy within three percent, determined from a ratio of chamber readings at two depths.
- Air kerma strength
- Brachytherapy source strength equal to air kerma rate in free space multiplied by squared distance; one unit equals one microgray square meter per hour or one centigray square centimeter per hour.
- Dose rate constant lambda
- Absorbed dose rate in water at one centimeter from the source along its transverse axis, divided by air kerma strength; a typical iridium-192 value is about 1.11 centigray per hour per unit.
- Iridium-192 half-life
- Approximately 73.8 days, with a mean photon energy near 380 keV; it is the standard high-dose-rate brachytherapy radionuclide and a source is typically exchanged every three to four months.
- Point A
- Classic gynecologic brachytherapy reference point located two centimeters lateral to the intrauterine tandem and two centimeters superior to the external cervical os along the applicator.
- D2cc
- Minimum dose covering the most irradiated two cubic centimeters of an organ at risk; this volume-based parameter is less sensitive to individual point placement than ICRU 38 reference points.
- Well chamber
- Reentrant ionization chamber used to measure brachytherapy source air kerma strength; it is calibrated against a national standard and corrected for temperature, pressure, and source geometry.
- Stoichiometric CT calibration
- Method that converts computed tomography Hounsfield units into electron density using tissue-equivalent inserts, enabling accurate heterogeneity correction in dose calculation algorithms.
- Four-dimensional CT
- Respiratory-correlated computed tomography that sorts images by breathing phase or amplitude, used to define internal target volumes and motion margins for thoracic and abdominal radiotherapy.
- Cone-beam CT limitations
- Scatter and truncation degrade Hounsfield unit accuracy, so cone-beam computed tomography is used for setup verification rather than primary dose calculation; imaging dose is typically a few milligray per scan.
- RTSTRUCT object
- DICOM radiotherapy object that stores structure contours, names, and references to the planning image series; it transfers target and organ delineations between contouring and planning systems.
- MRI geometric distortion
- Spatial inaccuracy in magnetic resonance images arising from gradient nonlinearity and field inhomogeneity; AAPM Task Group 284 addresses acceptance testing, quality assurance, and distortion correction for simulation.
- Surface-guided radiation therapy
- Uses stereoscopic cameras or structured light to track the patient skin surface in three dimensions for setup and intrafraction motion monitoring, adding no ionizing radiation dose.
- NCRP Report 151
- Provides structural shielding design methods for megavoltage photon and electron radiotherapy facilities, covering workload, use factor, occupancy factor, distance, and tenth-value-layer thickness calculations.
- Use factor
- Fraction of the workload during which the primary beam is directed at a particular barrier; it is one for the floor and commonly one quarter for walls in a four-field arrangement.
- Occupancy factor
- Fraction of time an area beyond a barrier is occupied by the same individual; it is one for offices and work areas and lower for corridors, restrooms, and infrequently used spaces.
- Linac leakage limit
- Leakage radiation from a medical linear accelerator must not exceed 0.2 percent of the useful beam dose rate at any point, with an average no greater than 0.1 percent.
- TG-100 FMEA
- Failure mode and effects analysis scores each process step for severity, occurrence, and detectability; the risk priority number is the product of the three scores and ranks hazards for mitigation.
- TG-263 nomenclature
- AAPM report establishing standardized names for targets, organs at risk, and plan structures so that radiotherapy data can be exchanged, pooled, and analyzed consistently across institutions.