ABR TMP study plan
ABR TMP · Give each week a clear focus.
8 weeks · About 6 hours per week
Treatment planning and dosimetry
Reconstruct the photon dose-calculation chain end to end: CT number to electron density conversion, beam commissioning data, and the path from percent depth dose and tissue-air ratio to absorbed dose.
Solve by hand at least six monitor-unit and dose problems covering SSD and SAD setups, inverse-square corrections, tray and wedge factors, and field-size output factors.
Derive and interpret plan-evaluation metrics (D95, D2, V20, conformity index, homogeneity index) and state the clinical failure mode each one is designed to catch.
Build a one-page equation sheet from memory, then verify every entry against your references and mark each correction in a different color.
Your checkpoint
A verified one-page dosimetry equation sheet plus six worked MU/dose problems, each graded against an independent recalculation with errors annotated.
Treatment planning and dosimetry & Treatment delivery, machine calibration, and quality assurance
Compare 3DCRT, IMRT, VMAT, and SBRT plans for the same hypothetical target and justify the chosen technique using dose-gradient, integral-dose, and treatment-time arguments.
Evaluate a written DVH summary for a target and two organs at risk, and write the specific replan or accept recommendation with the tradeoff quantified.
Explain how heterogeneity correction, grid size, and calculation algorithm choice shift calculated dose, and predict the direction of the shift in a lung-like geometry.
Draft the acceptance criteria you would apply to a patient-specific IMRT QA measurement and state which failures trigger a plan review versus a measurement repeat.
Your checkpoint
A two-page technique-selection and DVH-evaluation memo with at least four explicit numeric decision thresholds you would defend in an oral exam.
Treatment delivery, machine calibration, and quality assurance
Reconstruct the reference-dosimetry calibration chain from ion chamber reading to stated machine output, including temperature-pressure and electrometer corrections, and compute output for one changed condition.
Reproduce from memory the daily, monthly, and annual QA tests for a linear accelerator with their tolerances, then flag every item you could not recall and close those gaps.
Diagnose three written fault scenarios (output drift, MLC position error, flatness change) by naming the confirming measurement and the corrective action sequence.
Sequence an end-to-end IGRT workflow from simulation to treatment, stating the action taken at each matching threshold and who authorizes it.
Your checkpoint
A QA tolerance table with at least 30 rows (test, frequency, tolerance, action on failure) that is complete enough to run a monthly QA session from.
Treatment delivery, machine calibration, and quality assurance & Brachytherapy
Calculate source strength conversions and decay corrections for iridium-192, iodine-125, and palladium-103 high- and low-dose-rate sources, working from stated half-lives.
Apply the TG-43 formalism by hand to compute dose at a point from a single source with anisotropy and radial dose function values supplied in the problem, and state each assumption.
Compare HDR, PDR, and LDR approaches on fractionation, radiobiology, shielding, and workflow, and select the appropriate modality for two contrasting case vignettes.
Outline the response to a source dislodgement or stuck-source event, naming the survey, documentation, and notification steps in order.
Your checkpoint
A brachytherapy calculation worksheet with three fully worked source-strength/decay problems and one TG-43 point-dose calculation, plus a scored 20-item recall quiz on nuclide data.
Brachytherapy
Interpret implant quality metrics (D90, V100, V150, dose to organs at risk) for a prostate and a gynecologic implant and state the corrective action for each under-covered case.
Trace the brachytherapy workflow from applicator insertion through imaging, planning, treatment, and source removal, listing the verification checkpoint at each transition.
Explain how applicator reconstruction uncertainty and imaging modality choice propagate into reported dose for gynecologic implants.
Audit a written brachytherapy plan against a checklist you construct, and document every discrepancy found.
Your checkpoint
A completed two-case brachytherapy plan audit using your own checklist, with discrepancies ranked by clinical severity and a remediation note for each.
Brachytherapy & Imaging for radiation therapy and simulation
Compare CT simulation, kV and MV cone-beam CT, ultrasound, and MRI on contrast, spatial accuracy, artifacts, and dose, and match each to the clinical question it answers best.
Quantify image-quality tradeoffs by explaining how mAs, kVp, slice thickness, and reconstruction kernel change noise, contrast, and geometric fidelity.
Evaluate rigid versus deformable registration output for a written case and state when deformable registration must not be used for dose accumulation.
Integrate image guidance into a brachytherapy and an SBRT scenario, specifying the registration method, tolerance, and escalation path for each.
Your checkpoint
An imaging-modality comparison matrix with at least 12 attributes filled for five modalities, plus a one-paragraph justification of the imaging choice for each of two clinical scenarios.
Imaging for radiation therapy and simulation & Radiation protection, safety, informatics, and professionalism
Compute primary and secondary barrier transmission for one treatment room using stated workload, use factor, occupancy factor, and distance, and compare the result to the required limit.
Apply ALARA and dose-limitation principles to a written scenario involving a pregnant worker or a member of the public, and state the recommended action with the governing limit cited.
Trace a DICOM image set and RT plan through the informatics chain from scanner to treatment record and verify, identifying where a mismatch could cause a wrong-site or wrong-dose error.
Explain the reporting obligations and root-cause process following a written near-miss, and distinguish an error from a violation.
Your checkpoint
A shielding calculation with all inputs and the final comparison to the limit, plus a DICOM data-flow diagram annotated with at least five failure points and their controls.
Radiation protection, safety, informatics, and professionalism & Treatment planning and dosimetry
Synthesize professionalism content - informed consent, peer review, scope of practice, and conflict-of-interest disclosure - into answers for four written vignettes.
Complete a timed full-length mock examination under test conditions and score it by domain to identify residual weak areas.
Re-derive every formula on your equation sheet from memory and reconcile any drift against your corrected version.
Build a prioritized remediation list for the final days, ordered by expected score gain per hour of study.
Your checkpoint
A scored full-length timed mock exam with a per-domain score breakdown and a ranked remediation list of no more than ten items.
Milestones
Week 2: Equation sheet verified error-free and all six MU/dose problems independently reproduced within 2% of the reference answer.
Week 4: QA tolerance table at least 85% complete from memory on a blind re-test, and all brachytherapy calculation problems solved without reference lookup.
Week 6: Mid-plan self-assessment across all five domains scores at least 70%, with no single domain below 60%.
Week 8: Timed mock exam scores at least 80% overall with every domain at or above 70%, and the remediation list is closed out or explicitly deferred.