PET Scanner Sensitivity Loss After Detector Block Replacement

ABSNM NMPI · Put your knowledge to work in context.

All case studies

PET Scanner Sensitivity Loss After Detector Block Replacement

A PET/CT scanner had one detector block replaced. A daily quality control scan using a cylindrical phantom shows a sensitivity of 5.2 cps/kBq, compared with a baseline of 6.8 cps/kBq (a 24% drop). The technologist also notices that the uniformity image shows a cold region corresponding to the replaced block. The scanner is scheduled for oncology patients in 1 hour. The department's acceptance criterion is that sensitivity must be within 10% of baseline.

1. Determine whether the scanner meets the acceptance criterion and explain the clinical significance of the cold region.

Read the worked answer

Baseline = 6.8 cps/kBq; current = 5.2 cps/kBq. Percent change = (6.8 − 5.2)/6.8 × 100 = 23.5%, which exceeds the 10% acceptance criterion. The scanner fails.

The cold region corresponds to the replaced block, indicating that the block is either not fully coupled to its electronics, has a gain/energy calibration issue, or is partially inoperative. A cold detector region causes:

  • Underestimation of activity in that region (quantification bias).
  • Ring or streak artifacts in reconstructed images.
  • Potential missed or mischaracterized lesions.

The scanner should not be used for clinical PET until the block is calibrated and sensitivity returns within 10% of baseline. The 1-hour schedule should be rescheduled or diverted.

2. List the calibration and QC steps that should be performed after a detector block replacement before clinical use.

Read the worked answer

After block replacement:

  1. Verify block identification and electronic connections.
  2. Perform detector gain/energy calibration (photopeak position for the 511 keV photons).
  3. Run a normalization scan to correct for detector efficiency variations.
  4. Perform a well-counter cross-calibration if quantification is used.
  5. Run the daily QC (sensitivity, uniformity, and image quality) and confirm results are within 10% of baseline.
  6. Review the uniformity image for residual cold/hot regions.
  7. Document all results and obtain physicist approval before clinical use.