ABSNM NMPI study materials
ABSNM NMPI study resources

ABSNM NMPI study plan

ABSNM NMPI · Give each week a clear focus.

8 weeks · About 6 hours per week

Week 1

Gamma camera components and operation

  • Diagram the gamma camera signal path from collimator to photomultiplier tube to pulse-height analyzer, labeling each component's function

  • Explain how collimator hole geometry (septal thickness, hole diameter, length) affects sensitivity and spatial resolution

  • Compare NaI(Tl) scintillation crystal properties relevant to camera design, including light output, decay time, and hygroscopic handling

  • Describe how photomultiplier tube gain and position-weighted Anger logic localize an event

Your checkpoint

A hand-drawn labeled gamma camera block diagram plus a one-page written explanation of collimator trade-offs, self-graded against a checklist

Week 2

Gamma camera components and operation & Radiation detection and counting systems

  • Trace how pulse-height analysis and energy window settings discriminate photopeak events from Compton scatter

  • Calculate energy resolution from a given photopeak spectrum width and explain its clinical impact on scatter rejection

  • Contrast detector configurations (continuous crystal versus pixelated) for uniformity and count-rate behavior

  • Explain uniformity correction maps and when they are applied during acquisition

Your checkpoint

A worked problem set of 10 energy-window and energy-resolution calculations with written rationales for each answer

Week 3

Radiation detection and counting systems

  • Describe the operating principle and appropriate use of gas-filled, scintillation, and semiconductor detectors

  • Explain dead time, paralyzable versus nonparalyzable behavior, and how each distorts measured count rates

  • Apply the decay equation to compute activity at a later time and to perform dose calibrator constancy checks

  • Distinguish counting statistics concepts including Poisson behavior, standard deviation, and the effect of acquisition time on precision

Your checkpoint

A one-page reference sheet of detector types, dead-time formulas, and decay/statistics equations, validated by solving 8 practice problems

Week 4

Basic instrumentation principles and electronics & Radiation detection and counting systems

  • Explain the function of each stage in a nuclear counting chain: preamplifier, amplifier, discriminator, and scaler

  • Describe how pulse shaping and amplification affect energy resolution and count-rate performance

  • Interpret basic electronic noise sources and grounding/shielding practices relevant to instrumentation

  • Identify common causes of spectral artifacts from electronics rather than from the radionuclide

Your checkpoint

A timed 25-question self-made quiz on electronics and detection with a written error analysis of any missed items

Week 5

Image acquisition, processing, and quality control

  • Differentiate static, dynamic, gated, and whole-body acquisition modes and select appropriate parameters for each

  • Explain matrix size, zoom, pixel size, and their relationship to spatial resolution and count density

  • Describe uniformity, spatial resolution, and energy-peak QC tests, including acceptance criteria and frequency

  • Outline the purpose of flood-field, bar-phantom, and line-source images in camera QC

Your checkpoint

A completed QC log template listing each test, its frequency, acceptance criteria, and corrective action, with a short written justification for each criterion

Week 6

Emission tomography (SPECT and PET) instrumentation

  • Explain SPECT projection acquisition, reconstruction, and the role of filtered backprojection versus iterative reconstruction

  • Describe attenuation, scatter, and resolution-recovery corrections and when each is applied

  • Explain PET detector physics including annihilation photon detection, coincidence timing window, and time-of-flight principles

  • Compare SPECT and PET instrumentation capabilities for spatial resolution, sensitivity, and quantitation

Your checkpoint

A two-page comparison write-up of SPECT versus PET instrumentation with a labeled reconstruction workflow for each modality

Week 7

Instrumentation quality assurance and regulatory compliance

  • Map each major QA test to its regulatory or accreditation driver and required documentation

  • Explain the roles of the radiation safety officer and medical physicist in equipment oversight

  • Describe acceptance testing versus routine QC and the actions triggered by out-of-tolerance results

  • Identify recordkeeping requirements for instrument QC and survey results

Your checkpoint

A one-page compliance matrix linking QA tests to frequency, responsible party, and documentation requirement

Week 8

Basic instrumentation principles and electronics & Radiation detection and counting systems & Gamma camera components and operation & Image acquisition, processing, and quality control & Emission tomography (SPECT and PET) instrumentation & Instrumentation quality assurance and regulatory compliance

  • Integrate concepts across all six domains by working mixed-topic practice questions under timed conditions

  • Identify and remediate the three weakest domains based on practice-question performance

  • Re-derive key equations (decay, energy resolution, dead time, counting statistics) from memory

  • Summarize the highest-yield instrumentation facts into a single condensed review sheet

Your checkpoint

A 60-question timed mixed-domain practice exam with a scored error log and a completed one-page condensed review sheet

Milestones

Week 2: Explain and sketch the full gamma camera signal path and correctly solve at least 8 of 10 energy-window and resolution problems without notes

Week 4: Score at least 80% on the self-made electronics and detection quiz and correctly explain dead-time distortion in both paralyzable and nonparalyzable systems

Week 6: Produce the SPECT-versus-PET comparison write-up with accurate reconstruction workflows and no factual errors flagged for review

Week 8: Score at least 80% on the 60-question mixed-domain practice exam and demonstrate remediation of previously weak domains in the error log