Ohm's law
The relationship V = IR, where V is voltage, I is current, and R is resistance; used to analyze detector circuits.
Preamplifier function
Converts the small charge pulse from the detector into a voltage pulse, providing initial amplification and impedance matching.
Pulse height analyzer
An electronic circuit that sorts incoming pulses by amplitude, allowing selection of a specific energy window.
RC time constant
The product τ = RC; the time for a capacitor to charge to 63% or discharge to 37% of its initial value.
Paralyzable dead time
A detector model in which events occurring during dead time extend the dead time; observed rate m = n/(1 + nτ).
Nonparalyzable dead time
A detector model in which dead time is fixed after each event; observed rate m = n·e^(−nτ) for random events.
PMT gain
The amplification factor G = δ^n, where δ is the dynode multiplication factor and n is the number of dynodes.
Energy resolution FWHM
The full width at half maximum of a photopeak divided by its centroid energy, expressed as a percentage; improves with more photoelectrons.
NaI(Tl) crystal
A thallium-doped sodium iodide scintillator with high light output and good efficiency for 140 keV gamma rays; resolution ~6–7% at 140 keV.
Well counter
A NaI(Tl) detector with a well-shaped crystal providing high geometric efficiency for counting small radioactive samples.
Dose calibrator
A reentrant ionization chamber used to measure the activity of radionuclides in becquerels or curies; requires daily constancy checks.
Geiger–Müller counter
A gas-filled detector operating in the Geiger region; sensitive to beta and gamma contamination but has low gamma efficiency.
Photopeak
The peak in a gamma-ray spectrum corresponding to full energy absorption of the incident photon; its centroid identifies the radionuclide.
Compton continuum
The broad distribution of energies below the photopeak caused by Compton scattering of gamma rays in the detector.
Anger logic
A positioning method in gamma cameras that uses weighted signals from multiple photomultiplier tubes to determine event location.
Parallel-hole collimator
A collimator with parallel holes that produces an image the same size as the object; sensitivity depends on hole diameter and length.
Pinhole collimator
A collimator with a single small aperture that magnifies the image; used for imaging small organs such as the thyroid.
Energy window
The range of pulse heights accepted by the pulse height analyzer; typically 20% centered on the photopeak.
Uniformity correction map
A stored image of a flood source used to correct nonuniform response of the gamma camera during processing.
Septal penetration
The passage of photons through collimator septa, which degrades image contrast; depends on septal thickness and photon energy.
Pixel size
The field of view divided by the matrix dimension; smaller pixels improve sampling but reduce counts per pixel.
Integral uniformity
A measure of the maximum variation in count density across the useful field of view; typically expressed as a percentage.
Differential uniformity
A measure of local uniformity calculated over adjacent pixels; more sensitive to small-area nonuniformities than integral uniformity.
Bar phantom
A test object with parallel lead bars of varying spacing used to evaluate spatial resolution and linearity of a gamma camera.
Butterworth filter
A frequency-domain smoothing filter characterized by cutoff frequency and order; used to reduce noise in nuclear medicine images.
SPECT reconstruction
The process of creating tomographic slices from projections, typically using filtered backprojection or iterative methods such as OSEM.
PET coincidence detection
Detection of two 511 keV annihilation photons within a timing window (e.g., 4–12 ns) to define a line of response.
Time-of-flight PET
PET technique that measures the arrival time difference of annihilation photons to localize the event along the line of response, improving SNR.
PET randoms
Coincidence events from unrelated photons; rate ≈ 2τ·R1·R2, corrected by delayed-window or model-based methods.
Attenuation correction
Correction for photon attenuation in PET and SPECT, typically using CT or transmission sources, essential for quantitative imaging.
Daily gamma camera QA
Daily tests include uniformity flood and energy peaking to ensure stable camera performance before clinical use.
Dose calibrator constancy
A daily check using a long-lived source to verify that the dose calibrator response is stable within ±10%.