ABSNM NMPI study materials
ABSNM NMPI study resources

ABSNM NMPI key terms

ABSNM NMPI · Get comfortable with the vocabulary.

Photomultiplier tube (PMT)

A vacuum tube that converts light photons from a scintillator into an electrical signal and amplifies it through a series of dynodes. It provides high gain and fast response for radiation detection and gamma camera imaging.

Related: Scintillator, Pulse height analyzer, Quantum mottle
Scintillator

A material that absorbs ionizing radiation and emits visible light photons in proportion to the energy deposited. Common examples include NaI(Tl) for gamma cameras and LSO for PET.

Related: Photomultiplier tube (PMT), Light yield, Energy resolution
Pulse height analyzer (PHA)

An electronic circuit that sorts detector pulses by amplitude into energy bins, allowing selection of a specific energy window. It is used to discriminate between photopeak events and scattered radiation.

Related: Energy window, Photopeak, Scatter
Energy resolution

The ability of a detector to distinguish between radiation of different energies, usually expressed as the full width at half maximum (FWHM) of a photopeak divided by its centroid energy, given as a percentage.

Related: Photopeak, Scintillator, Pulse height analyzer (PHA)
Dead time

The time interval after a detected event during which the detector or processing electronics cannot record another event. It causes count loss at high counting rates and can be paralyzable or nonparalyzable.

Related: Count rate, Paralyzable dead time, Nonparalyzable dead time
Paralyzable dead time

A dead time model in which every event, including those occurring during the dead period, extends the dead time. At very high count rates, the observed count rate can decrease as the true rate increases.

Related: Dead time, Nonparalyzable dead time, Count rate
Nonparalyzable dead time

A dead time model in which events occurring during the dead period do not extend it. The observed count rate approaches a maximum asymptote as the true count rate increases.

Related: Dead time, Paralyzable dead time, Count rate
Photopeak

The peak in a gamma-ray spectrum corresponding to full absorption of the incident photon energy in the detector. It is used for energy window setting and for quantifying radionuclide uptake.

Related: Energy resolution, Energy window, Compton scatter
Compton scatter

Interaction in which a gamma photon transfers part of its energy to an electron and is deflected, producing a scattered photon with lower energy. Scattered photons can degrade image contrast if detected.

Related: Photopeak, Scatter, Energy window
Energy window

A selected range of pulse heights (energies) accepted by the pulse height analyzer. It is typically centered on the photopeak, such as a 20% window around 140 keV for technetium-99m.

Related: Pulse height analyzer (PHA), Photopeak, Scatter
Uniformity

The degree to which a gamma camera produces a constant count rate across the useful field of view when exposed to a uniform flood source. It is expressed as integral and differential uniformity.

Related: Flood field, Integral uniformity, Differential uniformity
Integral uniformity

A measure of the maximum and minimum pixel counts across the entire useful field of view, expressed as a percentage. It reflects the overall variation in camera response.

Related: Uniformity, Differential uniformity, Flood field
Differential uniformity

A measure of local uniformity that compares adjacent pixels or small regions across the field of view. It is more sensitive to small localized nonuniformities than integral uniformity.

Related: Uniformity, Integral uniformity, Flood field
Flood field

A uniform source of radioactivity used to assess gamma camera uniformity and sensitivity. It may be a liquid-filled flood phantom or a sheet source, and images are acquired with and without collimators.

Related: Uniformity, Integral uniformity, Sensitivity
Spatial resolution

The ability of an imaging system to distinguish two closely spaced point sources as separate. It is often expressed as full width at half maximum (FWHM) of a line spread function in millimeters.

Related: Line spread function, Collimator, Intrinsic resolution
Intrinsic resolution

The spatial resolution of a gamma camera without a collimator, determined by the detector and electronics. It is measured using a point or line source and is typically a few millimeters FWHM.

Related: Spatial resolution, Collimator, Line spread function
Collimator

A device made of lead or tungsten with holes or slits that restricts the direction of gamma rays reaching the detector, thereby forming an image. Its geometry determines sensitivity and spatial resolution.

Related: Spatial resolution, Sensitivity, Septal penetration
Septal penetration

The passage of gamma rays through the septa (walls) between collimator holes rather than through the holes. It degrades image contrast and increases with higher-energy radionuclides and thinner septa.

Related: Collimator, Scatter, Spatial resolution
Sensitivity

The number of counts recorded per unit of radioactivity in the source, often expressed as counts per minute per microcurie (cpm/µCi). It depends on collimator geometry and detector efficiency.

Related: Collimator, Count rate, Flood field
Scatter

Detection of gamma rays that have undergone Compton interaction in the patient or surrounding material and lost energy. Scattered events add background counts and reduce image contrast.

Related: Compton scatter, Energy window, Scatter correction
Scatter correction

A processing method that estimates and subtracts scattered counts from an image to improve contrast and quantitation. Common approaches include energy-window-based methods and deconvolution.

Related: Scatter, Energy window, Attenuation correction
Attenuation correction

A method that compensates for gamma-ray absorption and scattering within the patient, improving image uniformity and quantitative accuracy. It is commonly performed in SPECT/CT and PET/CT using transmission data.

Related: Scatter correction, Transmission scan, Standardized uptake value (SUV)
Transmission scan

An imaging procedure in which an external radioactive source or X-ray tube is used to measure attenuation through the patient. The data are used to create an attenuation map for correction.

Related: Attenuation correction, CT-based attenuation correction, Standardized uptake value (SUV)
Standardized uptake value (SUV)

A semiquantitative measure of radiotracer uptake in a region of interest, normalized to injected dose and patient body size. It is commonly used in PET oncology to assess lesion metabolism.

Related: Attenuation correction, Region of interest (ROI), Time-of-flight (TOF) PET
Region of interest (ROI)

A defined area on an image used for quantitative analysis, such as measuring counts or SUV. The ROI should be placed consistently to ensure reproducible results.

Related: Standardized uptake value (SUV), Count rate, Uniformity
Line spread function

The count rate profile obtained by imaging a thin line source, used to measure spatial resolution. Its full width at half maximum (FWHM) indicates the system's blurring.

Related: Spatial resolution, Intrinsic resolution, Modulation transfer function (MTF)
Modulation transfer function (MTF)

A measure of how well an imaging system preserves contrast at different spatial frequencies. It is derived from the line spread function and describes spatial resolution performance.

Related: Line spread function, Spatial resolution, Intrinsic resolution
Quantum mottle

A grainy appearance in images caused by statistical fluctuations in the number of detected photons. It becomes more noticeable with lower counts and can be reduced by increasing acquisition time or dose.

Related: Count rate, Noise, Sensitivity
Count rate

The number of detected events per unit time, usually expressed in counts per second or counts per minute. It is affected by source activity, detector efficiency, and dead time.

Related: Dead time, Sensitivity, Quantum mottle
Time-of-flight (TOF) PET

A PET technique that measures the difference in arrival times of annihilation photons at opposing detectors to localize the annihilation event along the line of response. It improves image signal-to-noise ratio.

Related: Positron emission tomography (PET), Line of response (LOR), Attenuation correction
Positron emission tomography (PET)

An imaging modality that detects pairs of 511 keV annihilation photons emitted from positron-emitting radionuclides. Coincidence detection localizes the annihilation event along a line of response.

Related: Time-of-flight (TOF) PET, Line of response (LOR), Coincidence detection
Line of response (LOR)

The straight line connecting two detectors that record a coincidence event in PET. The annihilation event is assumed to have occurred somewhere along this line.

Related: Coincidence detection, Positron emission tomography (PET), Time-of-flight (TOF) PET
Coincidence detection

The process of identifying two nearly simultaneous detector events in PET as originating from a single annihilation. A timing window is used to accept true coincidences and reject random events.

Related: Positron emission tomography (PET), Line of response (LOR), Random coincidence
Random coincidence

A coincidence event in which two unrelated photons from different annihilations are detected within the timing window. It adds background noise and is reduced by shorter timing windows and lower count rates.

Related: Coincidence detection, Scatter, Dead time
Single-photon emission computed tomography (SPECT)

A tomographic imaging technique that uses a rotating gamma camera to acquire multiple planar projections and reconstructs cross-sectional images. It typically uses single-photon-emitting radionuclides such as technetium-99m.

Related: Gamma camera, Attenuation correction, Scatter correction
Gamma camera

An imaging device that detects gamma rays using a large scintillation crystal, photomultiplier tubes, and a collimator to produce planar images of radionuclide distribution. It is the core instrument for conventional nuclear medicine.

Related: Collimator, Photomultiplier tube (PMT), Scintillator
Quality control (QC)

A set of scheduled tests and measurements performed to ensure that imaging equipment operates within specified performance limits. QC includes uniformity, resolution, sensitivity, and energy peaking checks.

Related: Uniformity, Sensitivity, Spatial resolution
Acceptance testing

A series of measurements performed after installation of equipment to verify that it meets manufacturer specifications and regulatory requirements before clinical use. It establishes baseline performance values for future QC.

Related: Quality control (QC), Uniformity, Sensitivity
Regulatory compliance

Adherence to laws, regulations, and standards governing the use of radioactive materials and medical imaging equipment. It includes licensing, radiation safety, and documentation requirements.

Related: Quality control (QC), Acceptance testing, Radiation safety