SPI · Keep the key ideas within reach.

Quick reference

Physical Principles and Instrumentation

  • Propagation speed in average soft tissue is 1540 m/s (1.54 mm/microsecond); wavelength in millimeters = 1.54 divided by frequency in MHz.

  • Attenuation averages about 0.5 dB per centimeter per megahertz, so total attenuation (dB) = 0.5 x frequency (MHz) x path length (cm).

  • Acoustic impedance = tissue density x propagation speed; intensity reflection coefficient = [(Z2 - Z1) / (Z2 + Z1)] squared.

  • Axial resolution = spatial pulse length / 2 = (wavelength x number of cycles per pulse) / 2; higher frequency and heavier damping shorten the pulse.

  • Lateral resolution is beam width, narrowest at the focal zone; it improves with higher frequency, larger aperture, and a lower f-number.

  • Elevational (slice-thickness) resolution is usually the poorest dimension and is fixed by the lens or by a 1.5-D array's elevation aperture.

  • Backing (damping) material shortens ring-down, lowers Q, and broadens bandwidth; the quarter-wave matching layer improves transmission from crystal into tissue.

  • Frame rate = pulse repetition frequency / number of pulses per frame (lines per frame); greater depth, wider sector, and higher line density each lower frame rate.

  • Receiver order is amplification, time-gain compensation, compression, demodulation, and rejection; scan conversion and postprocessing occur in the display stage.

Quick reference

Doppler and Hemodynamics

  • Doppler shift = (2 x transmitted frequency x velocity x cosine of angle) / propagation speed; the shift is maximal at 0 degrees and zero at 90 degrees.

  • Doppler angle should be kept at 60 degrees or less, because cosine changes steeply near 90 degrees and small angle errors create large velocity errors.

  • Nyquist limit = pulse repetition frequency / 2, the highest displayable shift; a shift above it aliases (wraps to the opposite channel or reverses in color).

  • Aliasing is reduced by raising PRF or scale, lowering transmitted frequency, shifting the baseline, or switching to continuous wave Doppler.

  • Continuous wave Doppler has no aliasing and no depth discrimination; pulsed Doppler provides a depth-selective sample gate but is subject to aliasing and range ambiguity.

  • Resistive index = (peak systolic - end-diastolic velocity) / peak systolic velocity; pulsatility index = (peak systolic - end-diastolic) / time-averaged mean velocity.

  • Pressure gradient across a narrowing is approximately 4 x velocity squared (Bernoulli); by continuity (A1V1 = A2V2), velocity rises where cross-sectional area falls.

  • Wall filters remove high-amplitude, low-frequency clutter from vessel wall and cardiac motion, but over-filtering erases genuine slow-flow signals.

  • Power Doppler encodes Doppler amplitude: it is angle independent, does not alias, and detects slow flow, but shows no direction or velocity; Reynolds number above about 2000 indicates turbulence.

Quick reference

Quality Assurance

  • Tissue-mimicking phantoms provide known attenuation, propagation speed, target depths, and cyst-like or solid lesions for reproducible measurements.

  • Axial resolution is the minimum separable reflector spacing along the beam; lateral resolution is measured across the beam at a stated depth.

  • Depth calibration should agree with true distance within about 2% or 2 mm, whichever is greater.

  • Dead zone is the distance from the transducer face to the first resolvable echo; a long dead zone obscures superficial structures.

  • Uniformity testing checks echo brightness across the image at fixed depth; vertical bands indicate failed elements and horizontal gradients indicate TGC errors.

  • Sensitivity (maximum penetration) is the deepest visible reflector at a given output, and anechoic cyst contrast tests contrast resolution.

  • Baseline QA is performed at installation, then on a scheduled interval such as semiannually, and always after transducer repair, drop, or replacement.

  • Transducer inspection looks for lens cracks, delamination, pitting, cable damage, and element failure, since these degrade images before electronics fail.

  • The output display standard requires real-time display of thermal and mechanical indices, supporting ALARA-based power reduction.

Quick reference

Safety and Artifacts

  • Thermal index estimates temperature rise as output power divided by the power needed to raise tissue 1 degree C; TIS is soft tissue, TIB bone at focus, TIC cranial.

  • Mechanical index = peak rarefactional pressure (MPa) / square root of frequency (MHz); values above roughly 0.5 raise cavitation concern when gas bodies exist.

  • FDA limits for general imaging: derated ISPTA at or below 720 mW/cm2 and MI at or below 1.9; ophthalmic limits are far lower (ISPTA 50 mW/cm2, MI 0.23, TI 1.0).

  • Stable cavitation is bubble oscillation; inertial cavitation is bubble collapse producing shock waves and local heating, with a lower threshold in the presence of contrast agent.

  • Reverberation produces equally spaced false echoes between two strong reflectors; comet tail and ring-down arise from fluid or gas traps.

  • Shadowing is reduced echogenicity deep to high-attenuation structures such as bone or calculi; enhancement is increased echogenicity deep to low-attenuation fluid.

  • Mirror image artifact duplicates a structure beyond a strong specular reflector (for example the diaphragm) because the extra path is not accounted for.

  • Speed-error artifact occurs when a tissue segment conducts faster or slower than the assumed 1540 m/s, displacing structures and refracting the beam.

  • Side lobes and grating lobes return echoes from off-axis reflectors that are displayed as if located along the main beam; color flash and twinkle artifacts arise from motion and rough calcific surfaces.