SPI flashcards
SPI · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Soft tissue propagation speed
- Ultrasound travels through average soft tissue at approximately 1540 meters per second, or 1.54 millimeters per microsecond. This assumed constant underlies all depth and distance calculations.
- Wavelength equation
- Wavelength equals propagation speed divided by frequency. In soft tissue, wavelength in millimeters equals 1.54 divided by frequency in megahertz, so higher frequency gives a shorter wavelength.
- Acoustic impedance
- Impedance is tissue density multiplied by propagation speed, expressed in rayleighs. A large impedance mismatch at an interface causes strong reflection and little transmission.
- Axial resolution equation
- Axial resolution equals spatial pulse length divided by two, or wavelength times cycles per pulse divided by two. Shorter pulses and higher frequencies improve axial resolution.
- Piezoelectric effect
- Certain crystals such as lead zirconate titanate convert applied voltage into mechanical vibration and convert returning vibration back into voltage, letting one element both transmit and receive.
- Curie temperature
- The temperature above which a piezoelectric crystal loses its polarization and stops producing ultrasound. Heating a transducer above this point permanently reduces its efficiency.
- Matching layer thickness
- A layer one-quarter wavelength thick placed at the transducer face. It provides intermediate impedance between crystal and skin, reducing reflection and increasing transmitted intensity.
- Backing (damping) material
- Material behind the crystal that absorbs rearward sound and stops crystal ringing. It shortens the pulse, broadens bandwidth, and improves axial resolution at the cost of sensitivity.
- Frame rate equation
- Frame rate equals pulse repetition frequency divided by the number of pulses per frame. Deeper imaging, wider sectors, and higher line density all lower the frame rate.
- Harmonic imaging
- The transducer transmits at the fundamental frequency and receives echoes at twice that frequency. Harmonics arise within tissue and improve lateral resolution while reducing side-lobe clutter.
- Doppler shift equation
- The Doppler shift equals twice the transmitted frequency times flow velocity times the cosine of the angle, divided by propagation speed. The shift is maximal at zero degrees.
- Nyquist limit
- The highest Doppler shift that can be displayed without aliasing, equal to half the pulse repetition frequency. Shifts above this value wrap to the opposite side of the display.
- Doppler aliasing
- A wraparound error on spectral or color Doppler occurring when the Doppler shift exceeds the Nyquist limit, so peak velocities appear on the opposite channel or as a color reversal.
- Continuous wave Doppler
- Two separate crystals transmit and receive continuously, so there is no pulse and no aliasing. It detects the highest velocities but has no depth discrimination along the beam.
- Resistive index
- Peak systolic velocity minus end-diastolic velocity, divided by peak systolic velocity. Values near zero indicate low-resistance flow; values approaching one indicate high-resistance flow.
- Bernoulli equation
- The pressure gradient across a stenosis is approximately four times the square of the peak velocity in meters per second, allowing severity to be estimated from a Doppler measurement.
- Continuity equation
- Volume flow equals cross-sectional area multiplied by velocity. Because flow is constant along a vessel, velocity increases wherever the cross-sectional area narrows.
- Power Doppler
- Displays the amplitude of the Doppler signal rather than its frequency shift. It is angle independent, does not alias, and detects slow flow, but shows no direction or velocity.
- Wall filter
- A high-pass filter that removes strong, low-frequency Doppler signals originating from moving vessel walls and cardiac motion. Excessive filtering can eliminate genuine slow-flow signals.
- Tissue-mimicking phantom
- A test object with known attenuation, propagation speed, and target positions, used to measure resolution, distance accuracy, and sensitivity. Its cyst-like and solid targets simulate tissue contrast.
- Dead zone
- The near-field distance from the transducer face to the first resolvable echo. It reflects ring-down and pulse duration, and a long dead zone hides shallow structures.
- Distance accuracy tolerance
- Measured distances on a phantom should agree with true values within about two percent or two millimeters, whichever is greater, confirming correct depth calibration of the system.
- Uniformity test
- Evaluates whether echo brightness is consistent across the whole image at a fixed depth. Bright or dark vertical bands reveal time-gain compensation errors or failed transducer elements.
- Anechoic cyst contrast test
- Measures the ability to detect and display a fluid-filled target against background tissue. Loss of contrast resolution appears as fill-in of the cyst interior.
- Transducer element dropout
- Failure of individual transducer elements produces vertical bands of missing or reduced echoes in the image, most easily detected with a uniformity phantom.
- Thermal index
- An estimate of temperature rise from absorbed ultrasound, calculated as output power divided by the power needed to raise tissue one degree Celsius. Subscripts denote soft tissue, bone, or cranial.
- Mechanical index
- Peak rarefactional pressure in megapascals divided by the square root of frequency in megahertz. Higher values increase the likelihood of cavitation, especially when gas bodies are present.
- Inertial cavitation
- Bubble collapse during the compression phase of an ultrasound wave, producing high local temperatures and shock waves. The threshold falls sharply when contrast microbubbles are present.
- Reverberation artifact
- Equally spaced false echoes produced when sound bounces repeatedly between two strong reflectors, such as the transducer face and a needle or a layer of fat.
- Acoustic shadowing
- Reduced echoes deep to a highly attenuating or strongly reflecting structure such as bone, calculus, or gas, appearing as a dark band below the object.
- Mirror image artifact
- A duplicate of a structure appearing beyond a strong specular reflector, such as the diaphragm, because the reflector acts as a mirror and the extra path length is not recognized.
- ALARA principle
- As low as reasonably achievable: ultrasound exposure is kept to the minimum that yields diagnostic information, using the lowest acceptable output and the shortest scanning time.