LTP12 · Keep the key ideas within reach.

Quick reference

General Laboratory Practice, Safety, and Quality Management

  • Quality management system hierarchy: quality manual → standard operating procedures → work instructions → forms and records; each level must trace upward to the level above.

  • Biosafety levels: BSL-1 nonpathogenic; BSL-2 agents of moderate hazard with BSL-2 cabinet and restricted access; BSL-3 aerosol-transmitted agents with directional airflow and respiratory protection; BSL-4 dangerous/exotic agents with positive-pressure suit.

  • Chemical hazard communication requires a Safety Data Sheet in 16 sections; Sections 1-8 cover identification, hazards, composition, first aid, firefighting, accidental release, handling/storage, and exposure controls.

  • Sharps and biohazardous waste are segregated at point of generation; chemical waste streams are never combined unless compatibility is documented.

  • Document control rule: only current versions are in use, obsolete documents are removed or clearly marked, and each controlled document carries a unique identifier, version, and effective date.

  • Competency assessment occurs at initial hire, after policy or method change, and at least annually thereafter; direct observation plus blind sample testing is typical.

  • Error rate formula: percent error = (measured − true) / true × 100; relative error uses the same expression and is reported without units.

  • Spill response sequence: alert and evacuate, secure the area, don PPE, contain, disinfect or absorb with appropriate agent, then verify and document.

Quick reference

Specimen Collection, Handling, and Processing

  • Order of draw for skin puncture: discard first drop, then EDTA, other additive tubes, serum tubes, and heparin last to limit tissue-fluid and additive carryover.

  • Tourniquet application should not exceed one minute; prolonged stasis raises potassium, lactate, and total protein and alters protein-bound analytes.

  • Hemolysis falsely increases potassium, LDH, AST, and phosphate, and falsely decreases glucose and bilirubin depending on method; document and recollect when interference is significant.

  • Additive inversion counts after collection: EDTA and heparin tubes 8-10 inversions, serum separator 5, citrate 3-4; citrate tubes must fill to the indicated line for a 9:1 blood-to-anticoagulant ratio.

  • Transport stability rules: ammonia and blood gases on ice and analyzed promptly; light-sensitive analytes (bilirubin, vitamin B12) protected from light; cryoglobulin and cold agglutinin specimens kept at 37 °C until separation.

  • Centrifugation for serum/plasma is typically 10-15 minutes at 1000-1200 × g at controlled temperature; incomplete clotting causes fibrin strands and instrument probe obstruction.

  • 24-hour urine collection requires a documented start and end time, proper preservative, and total volume; results are reported as concentration × volume per time period.

  • Rejection criteria include unlabeled or mislabeled containers, gross hemolysis, insufficient volume, expired collection devices, and unapproved transport temperature deviations.

Quick reference

Method Selection, Validation, and Verification

  • Validation establishes performance for a new in-house method; verification confirms that a manufacturer's validated method performs as claimed in the local laboratory.

  • Precision components: repeatability (within-run), intermediate precision (within-laboratory, across days/operators), and reproducibility (across laboratories); report as SD and CV percent.

  • CV percent = (SD / mean) × 100; compare observed CV against the laboratory's defined acceptance limit derived from biologic variation or manufacturer claims.

  • Accuracy is assessed by recovery studies, comparison with a reference method (Deming or Passing-Bablok regression), or bias against a certified reference material.

  • Linearity is evaluated across at least five concentration levels spanning the reportable range; deviation from linearity must remain within the laboratory's stated tolerance.

  • Limit of detection is the lowest amount detectable with stated confidence; limit of quantification is the lowest amount meeting predefined precision and bias goals.

  • Analytical measurement range is the interval producing results within acceptable accuracy and precision without dilution; clinical reportable range may extend it by validated dilution.

  • Interference testing evaluates endogenous substances (hemoglobin, bilirubin, lipids) and common drugs at defined concentrations; results are documented as percent bias at each level.

Quick reference

Quality Control, Calibration, and Proficiency Testing

  • Westgard multirule: 1-2s is a warning; 1-3s, 2-2s, R-4s, and 4-1s are rejection rules; 10x indicates systematic shift and is a rejection rule in most schemes.

  • Levey-Jennings chart plots control values over time against mean and ±1s, ±2s, ±3s limits; shifts are abrupt mean changes while trends are six or more consecutive values in one direction.

  • Calibration establishes the relationship between signal and concentration using calibrators; calibration verification confirms that relationship with independent materials at multiple levels.

  • Proficiency testing requires results be handled as patient samples, not shared between analysts, and not referred to another laboratory; unacceptable results trigger investigation and corrective action.

  • Control material selection: matrix should resemble patient samples, analyte levels should span medical decision points, and lot changes require establishment of new means and SDs.

  • Random error raises imprecision and triggers 1-3s or R-4s; systematic error shifts the mean and triggers 2-2s, 4-1s, or 10x.

  • Sigma metric = (TEa − |bias|) / CV; higher sigma indicates fewer QC rules are needed, and lower sigma requires tighter rules or more frequent QC.

  • Documented QC review must precede patient result release; out-of-control runs require corrective action, documentation, and repeat testing before reporting.

Quick reference

Analytical Measurement Techniques and Instrumentation

  • Beer-Lambert law: absorbance = molar absorptivity × path length × concentration; photometric methods are valid only in the linear absorbance region, generally below about 2.0 AU.

  • Spectrophotometer components in order: light source → monochromator → cuvette → photodetector → signal processor; stray light limits the upper measurable absorbance.

  • Ion-selective electrodes measure activity, not concentration; the Nernst equation relates potential to the logarithm of ion activity, and ionic strength affects results.

  • Potentiometric pH and blood gas analyzers use separate electrodes for pH, pCO2 (Severinghaus), and pO2 (Clark); each requires periodic two-point calibration.

  • Immunoassay formats include competitive (small analytes, signal inversely proportional to concentration) and sandwich (large analytes, signal directly proportional); the hook effect causes falsely low results at very high concentration.

  • Chromatography separates by differential partitioning between mobile and stationary phases; retention time identifies and peak area quantifies.

  • Mass spectrometry quantifies by mass-to-charge ratio; internal standards correct for ionization variability and matrix effects.

  • Osmolality by freezing point depression measures total particle concentration; the osmolal gap is measured osmolality minus calculated osmolality, normally under about 10 mOsm/kg.

Quick reference

Data Evaluation, Result Reporting, and Reference Intervals

  • Reference intervals are established from at least 120 healthy reference individuals per partition; the central 95 percent interval spans the 2.5th to 97.5th percentiles.

  • Partitioning by age and sex is justified when the difference between subgroup means or intervals is clinically and statistically meaningful.

  • Delta check compares a current result to a prior result from the same patient using an absolute or percentage threshold; failure triggers specimen review before reporting.

  • Critical values must be reported promptly with documented read-back of patient identifier, test, result, and the name of the person receiving it.

  • Calculated results (LDL cholesterol by Friedewald, anion gap, osmolality, eGFR) must be suppressed or flagged when input values fall outside the formula's validated range.

  • Results below the limit of detection are reported as less than that value, not as zero; results above the analytical range require validated dilution and a dilution factor notation.

  • Autoverification rules must be validated before use and must include range checks, delta checks, interference flags, and QC status before release.

  • Rounding and significant figures follow the measuring instrument's resolution; unit conversions must be verified, for example glucose mg/dL × 0.0555 = mmol/L.

Quick reference

Regulatory Compliance, Ethics, and Professional Conduct

  • CLIA categorizes tests as waived, moderate, or high complexity; laboratories hold a certificate appropriate to the testing performed and are subject to unannounced surveys.

  • Patient confidentiality rules prohibit disclosure of identifiable results without authorization except where required by law, public health reporting, or for treatment.

  • Research involving human specimens requires institutional review board approval and informed consent unless the protocol qualifies for exemption.

  • Method performance documentation must be retained per the laboratory's record retention policy, commonly at least two years for quality records and longer for some regulated records.

  • Conflict of interest must be disclosed when an employee has a financial interest in a vendor, test, or referral relationship.

  • Competency and proficiency testing records are personnel records; falsification of results or QC is a reportable integrity violation.

  • Specimen and result ownership: the laboratory owns records, the patient owns the information; release follows authorization and applicable privacy regulation.

  • Error disclosure: clinically significant errors are communicated to the responsible clinician promptly and documented, regardless of whether harm occurred.

Quick reference

Troubleshooting, Error Investigation, and Corrective Action

  • Root cause analysis tools: five whys, fishbone (cause-and-effect) diagram, and fault tree; the goal is to distinguish proximate from underlying causes.

  • Corrective action addresses the cause of a detected nonconformity; preventive action addresses a potential nonconformity before it occurs.

  • Investigation sequence for an aberrant result: verify specimen identity and integrity, review QC and calibration, inspect reagent and instrument status, then repeat the test.

  • Random error is suggested by scattered control values on both sides of the mean; systematic error is suggested by consecutive values on one side or a step change.

  • Hook effect is suspected when a result is unexpectedly low in a patient with clinical findings of high analyte; repeat after dilution to confirm.

  • Carryover is tested by running a high-concentration sample followed by a low-concentration sample; a positive bias in the low sample indicates carryover.

  • Corrective action effectiveness must be verified and documented, and the action closed only after monitoring shows the problem has not recurred.

  • Turnaround time failures are analyzed by timestamp audit at each phase: collection, receipt, analysis, and reporting, to locate the delay.