- Quality management system hierarchy
- Documents are arranged from quality manual through procedures and work instructions to forms and records, with each level traceable to the level above it.
- BSL-2 requirements
- Moderate-hazard agents require restricted access, biohazard signage, handwashing, a biosafety cabinet for aerosol-generating work, and trained personnel under supervision.
- Safety Data Sheet sections
- Sixteen standardized sections begin with identification and hazard identification and end with transport and regulatory information; Sections 1 through 8 cover first aid and handling.
- Competency assessment timing
- Assessment is required before independent testing, after a method or policy change, and at least annually thereafter, using observation and blind sample testing.
- Percent error formula
- Percent error equals measured minus true value, divided by true value, multiplied by one hundred, expressing the deviation as a percentage.
- Skin puncture order of draw
- After discarding the first drop, collect EDTA tubes first, then other additive tubes, serum tubes, and heparin last to minimize additive and tissue-fluid contamination.
- Citrate tube fill requirement
- Citrate tubes must fill to the indicated line to maintain the nine-to-one blood-to-anticoagulant ratio; short draws falsely prolong clotting times.
- Hemolysis interference effects
- Hemolysis falsely raises potassium, LDH, AST, and phosphate, and may lower glucose or bilirubin depending on the method; significant interference warrants recollection.
- Light-sensitive analyte handling
- Bilirubin, vitamin B12, folate, and porphyrins degrade with light exposure, so specimens are collected in amber or foil-wrapped tubes and transported protected.
- Specimen rejection criteria
- Unlabeled or mislabeled containers, gross hemolysis, insufficient volume, expired devices, and unapproved temperature excursions are grounds for rejection and recollection.
- Validation versus verification
- Validation establishes performance for a new or modified method; verification confirms that a manufacturer's validated method meets claims in the local laboratory.
- Precision versus accuracy
- Precision is agreement among replicate measurements, expressed as SD or CV; accuracy is agreement with the true value, expressed as bias or recovery.
- CV percent formula
- CV percent equals the standard deviation divided by the mean, multiplied by one hundred, and is compared against a predefined acceptance limit.
- Limit of quantification
- The lowest analyte concentration that can be measured with predefined precision and bias goals, typically the lowest point on the validated analytical range.
- Linearity evaluation
- At least five concentrations spanning the reportable range are tested; deviation from the expected line must fall within the laboratory's stated tolerance.
- Westgard 1-3s rule
- A single control value exceeding the mean by three standard deviations is a rejection rule indicating random error, requiring run rejection and investigation.
- Westgard 2-2s rule
- Two consecutive control values exceed the same two-standard-deviation limit, on one or across two materials, indicating systematic error and run rejection.
- Westgard 10x rule
- Ten consecutive control values fall on the same side of the mean, indicating a systematic shift such as a calibration drift or reagent lot change.
- Calibration versus calibration verification
- Calibration establishes the signal-to-concentration relationship using calibrators; calibration verification confirms that relationship using independent materials at multiple levels.
- Sigma metric formula
- Sigma equals the total allowable error minus absolute bias, divided by the CV; higher sigma permits fewer QC rules and lower sigma requires tighter monitoring.
- Beer-Lambert law
- Absorbance equals molar absorptivity times path length times concentration; photometric measurement is valid only within the linear absorbance region.
- Nernst equation application
- Ion-selective electrodes relate measured potential to the logarithm of ion activity, so ionic strength and matrix differences affect reported results.
- Hook effect
- Very high analyte concentrations saturate both capture and detection antibodies in sandwich immunoassays, producing falsely low results that resolve after dilution.
- Competitive immunoassay response
- Signal is inversely proportional to analyte concentration because labeled and unlabeled analyte compete for limited binding sites; used for small molecules.
- Osmolal gap calculation
- Measured osmolality minus calculated osmolality from sodium, glucose, and urea; a gap above roughly ten suggests unmeasured osmotically active substances.
- Reference interval definition
- The central ninety-five percent interval from at least one hundred twenty healthy reference individuals, spanning the 2.5th to 97.5th percentiles.
- Delta check purpose
- Compares a current result to a prior result from the same patient using an absolute or percentage threshold to flag possible specimen or analytic errors.
- Critical value read-back
- The receiving person repeats back patient identifier, test name, result, and their own name, and the entire exchange is documented with date and time.
- Friedewald calculation limits
- LDL cholesterol by Friedewald is invalid when triglycerides exceed about 400 mg/dL or when the patient is not fasting; such results must be suppressed or flagged.
- CLIA test complexity categories
- Tests are classified as waived, moderate, or high complexity, and the laboratory must hold a certificate matching the highest complexity it performs.
- Corrective versus preventive action
- Corrective action eliminates the cause of a detected nonconformity; preventive action eliminates the cause of a potential nonconformity before it occurs.
- Carryover testing procedure
- A high-concentration sample is followed immediately by a low-concentration sample; a positive bias in the low sample indicates carryover requiring wash optimization.