Quick reference

Hearing Conservation Program and OSHA Requirements

  • Action level: 85 dBA 8-hour TWA; a hearing conservation program is required for employees at or above this exposure.

  • PEL: 90 dBA 8-hour TWA; feasible engineering or administrative controls are required before hearing protectors are used to meet the PEL.

  • Core HCP components: noise monitoring, audiometric testing, hearing protection devices, annual training, and recordkeeping.

  • Baseline audiogram: within 6 months of first exposure at or above 85 dBA; mobile-van testing may extend this to 12 months with 14 hours without workplace noise before testing.

  • Annual audiogram: required for employees exposed at or above 85 dBA; compare with baseline to identify standard threshold shifts.

  • Hearing protectors: offered at 85 dBA TWA, mandatory at 90 dBA TWA, and required after a confirmed STS until professional review or return to baseline.

  • STS notification: employee must be informed in writing within 21 days of the determination.

  • Records: noise exposure measurements retained 2 years; audiometric records retained for employment plus 30 years.

Quick reference

Noise Measurement, Exposure Assessment, and Dosimetry

  • Sound level meter: measures sound pressure level at a location; use for area surveys and source identification.

  • Dosimeter: worn by worker; integrates dose over shift and computes TWA for personal exposure.

  • OSHA exchange rate: 5 dB; a 5 dB increase doubles dose and halves allowable duration.

  • Dose reference: 90 dBA for 8 hours = 100% dose; 85 dBA for 8 hours = 50% dose.

  • TWA formula: TWA = 16.61 log10(D/100) + 90, where D is percent dose for an 8-hour shift.

  • Allowable time: 90 dBA = 8 h; 95 dBA = 4 h; 100 dBA = 2 h; 105 dBA = 1 h under the 5 dB exchange rate.

  • Measure with A-weighting and slow response for steady noise; use integrating meter/dosimeter for fluctuating exposure.

  • Document calibration before/after, microphone location, sampling time, tasks, sources, and representative conditions.

Quick reference

Anatomy and Physiology of the Ear and Noise Effects

  • Outer ear: pinna and ear canal collect and direct sound to the tympanic membrane.

  • Middle ear: tympanic membrane and ossicles (malleus, incus, stapes) transfer sound to the cochlea.

  • Inner ear: cochlea contains hair cells that convert mechanical vibration into auditory nerve signals.

  • Conductive loss: outer/middle ear problem; sensorineural loss: cochlea or auditory nerve problem.

  • Noise-induced hearing loss is usually sensorineural, bilateral, and irreversible; initial pattern is a high-frequency notch.

  • Typical notch: 3000-6000 Hz, often greatest at 4000 Hz; speech frequencies may be affected with continued exposure.

  • Temporary threshold shift: reversible elevation after noise; repeated TTS can progress to permanent threshold shift.

  • Hair cell damage reduces amplification, frequency selectivity, and clarity; tinnitus often accompanies noise injury.

Quick reference

Audiometric Testing Procedures and Test Environment

  • Required test frequencies: 500, 1000, 2000, 3000, 4000, and 6000 Hz air conduction in each ear.

  • Test environment: booth/room must meet OSHA Appendix D octave-band background noise limits.

  • Baseline timing: within 6 months of first exposure at or above 85 dBA; precede baseline with 14 hours without workplace noise.

  • Threshold method: ascending 5 dB presentations; threshold is lowest level heard in at least 50% of responses.

  • Annual testing: compare current audiogram with baseline; retest or refer when STS or validity concerns arise.

  • Record: date, examiner, calibration status, ear-specific thresholds, and test conditions.

  • Instruction: employee should respond whenever tone is heard, even faint; avoid visual cues.

  • Validity: inconsistent responses, background noise, earphone fit, or tinnitus can affect thresholds; note and retest.

Quick reference

Audiometer and Equipment Calibration and Biological Checks

  • Daily biological check: test a person with stable hearing before daily use; compare with reference thresholds.

  • Daily listening check: listen for distortion, unwanted noise, intermittency, and proper earphone output.

  • Acoustic calibration: at least annually to ANSI S3.6; verify frequencies, output levels, attenuator linearity, and distortion.

  • Exhaustive calibration: perform at interval required by OSHA/manufacturer; document all results and corrective actions.

  • Booth noise check: verify background noise meets Appendix D limits; recheck after relocation or construction.

  • Earphone inspection: check cushions, cords, headband tension, and labels; damaged parts must be replaced.

  • Biological check limits: compare listener thresholds with reference; investigate unexpected differences at any frequency.

  • Calibration records: retain dates, methods, equipment IDs, results, and repair history.

Quick reference

Audiogram Interpretation, Baseline Revision, and STS

  • STS definition: average shift of 10 dB or more at 2000, 3000, and 4000 Hz in either ear.

  • Compare annual audiogram to baseline; if STS, retest within 30 days to confirm.

  • Confirmed STS: revise baseline, notify employee, and apply hearing protector requirement.

  • Recordability: STS is recordable on OSHA 300 if average hearing level at 2,3,4 kHz is 25 dB or more above audiometric zero.

  • Age correction: allowed for STS determination; document method and apply consistently.

  • Not confirmed: original baseline remains; investigate cause and retest as needed.

  • Interpretation: high-frequency notch suggests noise; asymmetric or conductive loss warrants professional referral.

  • Notification: provide STS results in writing within 21 days.

Quick reference

Hearing Protection Devices, Fitting, Attenuation, and Fit Testing

  • NRR: single-number laboratory attenuation rating; actual protection is usually lower.

  • OSHA A-weighted estimate: protected exposure = TWA - (NRR - 7); C-weighted estimate = TWA - NRR.

  • NIOSH derating: earmuffs 25%, formable earplugs 50%, other earplugs 70%.

  • Selection: match protector to exposure, comfort, communication needs, and concurrent hazards.

  • Foam plug fit: roll, pull pinna back, insert deeply, hold for expansion; poor fit reduces attenuation.

  • Earmuff fit: seal must fit around ear; hair, glasses, and caps can break the seal.

  • Fit testing: yields a personal attenuation rating (PAR); use to verify fit and train workers.

  • Care: clean reusable protectors, replace worn parts, store properly, and reissue when fit changes.

Quick reference

Employee Education, Motivation, Recordkeeping, and Program Evaluation

  • Annual training content: noise effects, audiometric testing, HPD selection/fit/use/care, and program purpose.

  • Training must be understandable to employees and documented with date, content, and attendees.

  • Motivation: use personal audiograms, fit testing, demonstrations, and visible management commitment.

  • Noise exposure records: retain 2 years; audiometric records: employment plus 30 years.

  • Employee access: provide monitoring results and audiometric records on request.

  • Program evaluation: track STS rates, dose exceedances, HPD use, training completion, and calibration compliance.

  • Review program when processes, equipment, or exposure levels change, and at least annually.

  • Recordkeeping: maintain calibration logs, training records, fit-test results, and noise measurement data.