CSPDT study materials
CSPDT study resources

CSPDT cheat sheet

CSPDT · Keep the key ideas within reach.

Quick reference

Decontamination

  • PPE for decontamination: fluid-resistant gown, gloves, mask with eye protection or face shield, shoe covers, and hair cover; remove all PPE before leaving the area.

  • Cleaning is the first and most critical step - residual soil protects microorganisms and can inactivate disinfectants and sterilants.

  • Point-of-use treatment: keep contaminated items moist with approved gel, foam, or water; never let blood or saline dry on the device.

  • Manual cleaning: brush instruments while submerged below the water surface to prevent aerosol generation; use a brush sized to the lumen.

  • Ultrasonic cleaner: degas the solution, fully submerge items, and open hinges and box locks; cavitation removes soil but does not sterilize.

  • Decontamination area is held at negative air pressure relative to adjacent spaces, with a minimum of 10 air exchanges per hour.

  • Final rinse water in washer-disassemblers should be treated (RO/DI) to reduce mineral and pyrogen residue on instruments.

  • Washer-disassembler verification each load: check spray arms for blockage, detergent level, and the recorded time-temperature printout.

  • Disassemble multi-part instruments per the manufacturer's instructions; open box locks and remove all detachable parts before cleaning.

Quick reference

Preparation and Packaging

  • Every instrument is inspected for cleanliness, function, alignment, sharpness, corrosion, and cracked or frayed insulation before packaging.

  • Hinged instruments are opened to the first ratchet notch or fully open so sterilant contacts all surfaces.

  • Instrument lubricants must be water-based, steam-permeable, and approved by the device manufacturer; oil-based lubricants are not used.

  • An internal chemical indicator goes inside every package and an external indicator (tape or label) on every package; neither proves sterility.

  • Weight limits: instrument sets about 25 lb, textile packs about 12 lb, and textile density about 7.2 lb/ft3 to allow penetration and drying.

  • Peel pouch loading: the item should occupy no more than about three-quarters of the pouch and must not touch or stress the seals.

  • Label each package with sterilizer identifier, cycle or load number, date of sterilization, preparer's initials, and expiration or event-related date.

  • Nested basin sets: separate basins with absorbent material and position them so condensate can drain; avoid tightly nested configurations.

  • Use only FDA-cleared packaging validated for that specific sterilization method; inspect wraps for tears and punctures after the cycle.

Quick reference

Sterilization

  • Steam cycles (AAMI-referenced): 121-123 C (250-252 F) gravity for 30 min; 132-135 C (270-275 F) prevacuum for 3-4 min; the device IFU always governs.

  • Biological indicator organisms: Geobacillus stearothermophilus for steam, ethylene oxide, and hydrogen peroxide plasma; Bacillus atrophaeus for dry heat and EO.

  • Bowie-Dick test: prevacuum steam only, run daily in an empty warmed chamber; it detects air removal and steam penetration failure, not sterility.

  • Chemical indicator classes: 1 process indicator, 2 Bowie-Dick, 3 single parameter, 4 multi-parameter, 5 integrating, 6 emulator.

  • Implant loads are quarantined until the biological indicator is negative; a rapid-read BI may permit earlier release only if facility policy allows.

  • Positive biological indicator: recall all items processed in that sterilizer back to the last negative BI and do not release new loads until the cause is resolved.

  • Dry heat parameters: 160 C (320 F) for 2 hours or 170 C (340 F) for 1 hour, used for powders, oils, and items that cannot tolerate moisture.

  • EO parameters are typically 55-60 C, 40-80% relative humidity, 450-1200 mg/L concentration, 1-6 hours exposure, followed by required aeration.

  • Load configuration: do not overload; stand paper-plastic pouches on edge, place rigid containers flat, and tilt basins so condensate drains.

Quick reference

Sterile Storage and Distribution

  • Storage environment: approximately 20-24 C (68-75 F) and 30-60% relative humidity, with temperature and humidity documented.

  • Clearances: at least 8-10 inches above the floor, 18-20 inches below the ceiling, and 2 inches from exterior walls; nothing stored under sinks or water pipes.

  • Event-related sterility: shelf life depends on storage and handling events rather than elapsed time alone.

  • Store paper-plastic pouches on edge rather than flat or folded; store rigid containers per the container manufacturer's instructions.

  • Transport sterile items in closed or covered carts; never carry them against the body or set them on the floor.

  • Issuance check before distribution: package integrity, correct label, correct item, and valid event-related status.

  • Rotate stock first in, first out (FIFO); segregate recalled or compromised packages and return them for reprocessing.

  • Traffic control: separate clean from contaminated flow and minimize personnel traffic in the sterile storage area.

Quick reference

Professional and Regulatory Standards

  • Spaulding classification: critical devices require sterilization, semicritical devices require at least high-level disinfection, noncritical devices require low-level disinfection.

  • OSHA Bloodborne Pathogens Standard (29 CFR 1910.1030): exposure control plan, hepatitis B vaccination offered within 10 working days of assignment, PPE, training, and sharps injury log.

  • OSHA Hazard Communication (29 CFR 1910.1200): chemical inventory, accessible safety data sheets, labeled containers, and staff training.

  • AAMI ST79 covers steam sterilization and sterility assurance; ST91 covers flexible endoscope processing; ST58 covers chemical sterilization and high-level disinfection.

  • FDA clears sterilizers, biological indicators, chemical indicators, and sterilization packaging as medical devices before marketing.

  • The CDC Guideline for Disinfection and Sterilization in Healthcare Facilities sets device processing levels; Joint Commission and CMS survey compliance against them.

  • Device, sterilizer, and packaging manufacturer instructions for use must all be compatible; where they conflict, the most restrictive instruction governs.

  • Records must be legible and traceable to the load and patient recall; retention periods follow facility policy and regulatory requirements.

  • CSPDT certification requires passing the CBSPD examination and maintaining continuing education as required for recertification.