Bioburden

The amount of organic soil, including blood, tissue, and microorganisms, remaining on an item before it is cleaned or sterilized. High bioburden can shield microbes and cause disinfection or sterilization to fail.

Related: Enzymatic Detergent, Sterility Assurance Level
Spaulding Classification

A system that sorts medical devices into critical, semicritical, and noncritical groups based on how they contact the body, which then sets the minimum level of reprocessing each device needs.

Related: High-Level Disinfection, Sterility Assurance Level
High-Level Disinfection

A process that destroys all microorganisms except large numbers of bacterial spores. It is used for semicritical devices, such as endoscopes, that touch mucous membranes but do not enter sterile body areas.

Related: Spaulding Classification, Bioburden
Personal Protective Equipment

Barriers such as gloves, gowns, masks, eye protection, and shoe covers worn to protect the technician from contact with blood, body fluids, and contaminated items during reprocessing.

Related: Standard Precautions, Decontamination Area
Decontamination Area

The designated, physically separated room where soiled items are received, cleaned, and disinfected. It is kept at negative air pressure relative to clean areas so contaminants stay contained.

Related: Personal Protective Equipment, Bioburden
Enzymatic Detergent

A cleaning product containing enzymes that break down proteins, fats, and carbohydrates on instruments. It is used at the manufacturer's recommended dilution, temperature, and contact time before disinfection or sterilization.

Related: Bioburden, Ultrasonic Cleaner
Ultrasonic Cleaner

A machine that uses high-frequency sound waves to form bubbles that implode and lift soil from instrument surfaces, including hinges and serrations. It performs cleaning only, not disinfection or sterilization.

Related: Enzymatic Detergent, Inspection and Assembly
Point-of-Use Treatment

Immediate bedside care of a soiled instrument, such as wiping it or applying a moistening gel, so soil does not dry and harden before the item reaches the decontamination area.

Related: Bioburden, Decontamination Area
Prion

An abnormally folded infectious protein that causes diseases such as Creutzfeldt-Jakob disease. Prions resist routine disinfection and sterilization, so special manufacturer and public health protocols are required.

Related: Bioburden, Spaulding Classification
Inspection and Assembly

The step in which cleaned instruments are checked for cleanliness, damage, and function, then arranged into a set according to its count sheet before packaging.

Related: Count Sheet, Instrument Tray
Count Sheet

A list of every item and its quantity that belongs in a particular instrument set. It is used during assembly and again at the point of use to confirm the set is complete.

Related: Instrument Tray, Inspection and Assembly
Instrument Tray

A basket or container holding a specific group of instruments needed for a procedure, configured according to a count sheet and processed as one packaged unit.

Related: Count Sheet, Rigid Sterilization Container
Hinged Instrument Open Position

Instruments with box locks, hinges, or ratchets are packaged unlocked and open so steam or sterilant can reach every surface. Closed or locked instruments can trap air and block sterilization.

Related: Instrument Tray, Steam Sterilization
Sterilization Wrap

A medical-grade material, such as nonwoven polypropylene or woven cotton, used to wrap items so sterilant can enter and leave while microorganisms are kept out after processing.

Related: Peel Pouch, Rigid Sterilization Container
Peel Pouch

A sealed paper-and-film package used for small or single instruments. The clear film lets staff see the item, and the package peels open cleanly for aseptic presentation.

Related: Heat Sealer, Sterilization Wrap
Rigid Sterilization Container

A reusable metal or plastic box with a lid and a filter or valve system that holds instruments during sterilization and protects them afterward. It must be validated for the sterilizer type used.

Related: Instrument Tray, Sterile Storage Area
Heat Sealer

A machine that closes peel pouches using controlled temperature, pressure, and dwell time to form a continuous seal. Seal quality is checked visually and by periodic strength testing.

Related: Peel Pouch, Quality Control
Chemical Indicator

A strip, dot, or tape that changes color when exposed to one or more sterilization parameters such as temperature, steam, or a chemical agent. It shows exposure but does not prove sterility.

Related: Biological Indicator, Bowie-Dick Test
Biological Indicator

A test strip or vial containing bacterial spores with known high resistance to a specific sterilization process. It is the most reliable routine evidence that the process actually killed microorganisms.

Related: Geobacillus stearothermophilus, Sterility Assurance Level
Geobacillus stearothermophilus

A spore-forming bacterium used in biological indicators for steam and hydrogen peroxide sterilization because its spores are highly resistant to those processes.

Related: Biological Indicator, Steam Sterilization
Steam Sterilization

A process that uses saturated steam under pressure at a set temperature and time, commonly 132 degrees Celsius for four minutes, to kill microorganisms by denaturing their proteins.

Related: Prevacuum Sterilizer, Bowie-Dick Test
Prevacuum Sterilizer

A steam sterilizer that first pulls a vacuum to remove air from the chamber before admitting steam. Air removal is more complete than in gravity units, so cycles run faster.

Related: Steam Sterilization, Bowie-Dick Test
Bowie-Dick Test

A test pack run in an empty chamber of a prevacuum steam sterilizer to detect air leaks and poor air removal. It is typically performed at the start of each operating day.

Related: Prevacuum Sterilizer, Chemical Indicator
Ethylene Oxide Sterilization

A low-temperature process using a toxic, flammable gas to sterilize heat- or moisture-sensitive items. Loads require aeration afterward so toxic residues are removed before use.

Related: Biological Indicator, Instructions for Use
Sterility Assurance Level

The probability that a single item is nonsterile after a validated sterilization process. For medical devices it is usually set at one in a million, written as 10 to the minus sixth power.

Related: Validation, Biological Indicator
Immediate-Use Steam Sterilization

Also called flash sterilization: processing an unpackaged item for immediate use when a full cycle cannot be done. It is reserved for urgent situations and needs documented justification.

Related: Steam Sterilization, Sterile Storage Area
Event-Related Sterility

The principle that a properly packaged sterile item stays sterile indefinitely unless the package is damaged, wet, or compromised, or storage conditions fail. It replaces fixed expiration dates.

Related: Compromised Package, Sterile Storage Area
Sterile Storage Area

A clean, restricted room where processed items are held before use. It is separated from soiled areas and has controlled temperature, humidity, and traffic to protect package integrity.

Related: Traffic Control, Event-Related Sterility
Compromised Package

A sterile package that is torn, punctured, wet, stained, or has a broken seal. Sterility can no longer be assumed, so the item must be reprocessed before use.

Related: Event-Related Sterility, Sterilization Wrap
First In, First Out

A stock rotation method in which the oldest processed items are used first. It helps ensure supplies are used before materials degrade or storage conditions affect them.

Related: Sterile Storage Area, Event-Related Sterility
Case Cart

A cart assembled with the supplies and instrument sets needed for one scheduled procedure. It is prepared in advance and delivered so operating room staff do not search for items.

Related: Instrument Tray, Count Sheet
Traffic Control

Restricting who enters sterile storage and how people move through it, such as limiting access and requiring clean attire, to reduce airborne contamination and package damage.

Related: Sterile Storage Area, Personal Protective Equipment
Instructions for Use

The manufacturer's written directions for cleaning, packaging, sterilizing, and storing a device or sterilizer. Following them is required because the device's validation depends on those exact steps.

Related: Validation, AAMI ST79
Validation

Documented evidence that a process consistently produces the intended result, such as a sterilization cycle proven to reach the required sterility assurance level under defined conditions.

Related: Sterility Assurance Level, Instructions for Use
AAMI ST79

A widely used consensus standard for steam sterilization and sterility assurance in healthcare facilities. It covers cleaning, packaging, loading, monitoring, storage, and quality control practices.

Related: Instructions for Use, Quality Control
Standard Precautions

Infection control practices applied to every patient's blood and body fluids, including gloves, gowns, masks, eye protection, and hand hygiene, because any patient may carry a transmissible disease.

Related: Personal Protective Equipment, Bloodborne Pathogens Standard
Bloodborne Pathogens Standard

An OSHA regulation requiring employers to protect workers from blood and other potentially infectious materials through an exposure control plan, protective equipment, training, and hepatitis B vaccination.

Related: Standard Precautions, Personal Protective Equipment
Quality Control

Planned checks and records, such as indicator results, temperature logs, and seal tests, that show reprocessing steps are working as intended and let problems be found early.

Related: Biological Indicator, Documentation
Documentation

Written records of each reprocessing step, including load contents, cycle parameters, and monitoring results. Good records provide traceability and are essential when investigating a sterilization failure.

Related: Load Recall, Quality Control
Load Recall

The process of retrieving and reprocessing or discarding items from a load after a failed biological indicator or other sterilization failure, and notifying anyone who may have received them.

Related: Biological Indicator, Documentation