CSPDT flashcards
CSPDT · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Enzymatic detergent
- Cleaning solution containing enzymes that break down protein, fat, and blood; it is diluted, warmed, and timed according to the manufacturer's instructions and replaced when soiled.
- Point-of-use treatment
- Keeping contaminated instruments moist with an approved gel, foam, or water immediately after use so blood and saline do not dry and harden before cleaning.
- Negative-pressure decontamination room
- Decontamination area held at negative air pressure relative to surrounding rooms so airborne microorganisms stay contained; typically at least ten air exchanges per hour are required.
- Cavitation
- Formation and violent collapse of tiny bubbles in an ultrasonic cleaner that pull soil from instrument surfaces; items must be completely submerged and the solution degassed first.
- Thermal disinfection
- Hot-water rinse in a washer-disassembler at a validated time and temperature that reduces microbial bioburden on instruments; it is not a sterilization process.
- Lumen flushing
- Flushing the internal channel of a hollow instrument with detergent solution and water using a correctly sized brush and connector, following the manufacturer's written instructions.
- Instrument inspection
- Checking every item for cleanliness, proper function, alignment, sharpness, corrosion, and intact insulation before packaging; defective instruments are removed and repaired or replaced.
- Hinged instrument position
- Box locks and ratchets are opened to the first notch or fully open during preparation so steam or chemical sterilant can contact all instrument surfaces.
- Internal chemical indicator
- Indicator placed inside every package or tray to confirm exposure to the sterilization process; it does not prove sterility and does not replace biological monitoring.
- Instrument set weight limit
- Instrument sets generally should not exceed 25 pounds and textile packs should not exceed 12 pounds so sterilant penetrates and packages dry properly.
- Event-related sterility
- Shelf life based on storage and handling events rather than a fixed expiration date; packages remain sterile until torn, wet, compressed, or otherwise compromised.
- Package labeling elements
- Each package is labeled with the sterilizer identifier, cycle or load number, sterilization date, preparer's initials, and the facility's expiration or event-related date.
- Geobacillus stearothermophilus spores
- Highly heat-resistant spores used as the biological indicator for steam, ethylene oxide, and hydrogen peroxide gas plasma sterilization; a positive result indicates process failure.
- Bacillus atrophaeus spores
- Spore-forming organism used as the biological indicator for dry heat and ethylene oxide sterilization; it is more resistant than vegetative bacteria to those processes.
- Bowie-Dick test
- Daily warm-up test in an empty prevacuum steam sterilizer that detects air removal and steam penetration problems; a failed test means the load cannot be released.
- Implant load quarantine
- Implants are held until the load's biological indicator is negative; a rapid-read biological indicator may permit earlier release only when facility policy allows.
- Immediate-use steam sterilization
- Sterilizing unwrapped items for urgent use with no storage; the item is transported aseptically in a covered container and used immediately, formerly called flash sterilization.
- Class 5 integrating indicator
- Chemical indicator that responds to all critical sterilization parameters and is designed to correlate with a biological indicator at a specified location in the sterilizer.
- Positive biological indicator
- Result showing surviving spores after a cycle; all items processed since the last negative indicator from that sterilizer are recalled and the sterilizer is investigated.
- Dry heat sterilization parameters
- Hot-air sterilization commonly run at 160 C for two hours or 170 C for one hour, used for items that cannot tolerate moisture such as powders and oils.
- Sterility assurance level
- Probability that an item remains nonsterile after sterilization; the accepted level for medical devices is one in one million, written as 10 to the minus sixth.
- Sterile storage humidity range
- Sterile storage areas are maintained near 20-24 C (68-75 F) with 30-60% relative humidity to protect package integrity and keep items sterile.
- Storage clearance distances
- Sterile supplies are stored at least 8-10 inches above the floor, 18-20 inches below the ceiling, and 2 inches from exterior walls for cleaning and air circulation.
- First in, first out
- Inventory practice of issuing the oldest sterile stock first; each package is checked for integrity and correct labeling before it leaves the department.
- Closed transport cart
- Sterile items are moved in closed or covered carts to protect packages from dust, moisture, and contamination during distribution within the facility.
- Compromised sterile package
- A package that is torn, wet, soiled, punctured, or opened is considered contaminated and must be returned for inspection and reprocessing, never issued.
- Spaulding classification
- System that groups devices as critical, semicritical, or noncritical based on contact with sterile tissue, mucous membranes, or intact skin, which determines the required processing level.
- Critical device
- Item that enters sterile tissue or the vascular system; it must be sterilized before use because any microbial contamination carries a high risk of infection.
- Semicritical device
- Item that contacts mucous membranes or nonintact skin but not sterile body cavities; it requires at least high-level disinfection, and sterilization when feasible.
- OSHA Bloodborne Pathogens Standard
- Federal rule 29 CFR 1910.1030 requiring an exposure control plan, hepatitis B vaccination offer, personal protective equipment, training, and sharps injury recordkeeping.
- AAMI ST79
- Comprehensive national guideline for steam sterilization and sterility assurance in health care facilities, covering decontamination, packaging, loading, monitoring, and sterile storage.
- Hazard Communication Standard
- OSHA rule requiring a chemical inventory, accessible safety data sheets, labeled containers, and staff training so workers know the hazards of chemicals they handle.