MDCG Pre-Registration Assessment study materials
MDCG Pre-Registration Assessment study resources

MDCG Pre-Registration Assessment cheat sheet

MDCG Pre-Registration Assessment · Keep the key ideas within reach.

Quick reference

EU Medical Device Regulatory Framework

  • MDR 2017/745 applies to medical devices and accessories; IVDR 2017/746 applies to in vitro diagnostic medical devices. Both Regulations are directly applicable in all EU Member States without national transposition.

  • MDR Article 10 and IVDR Article 10 place obligations on manufacturers: risk management, clinical/performance evaluation, technical documentation, conformity assessment, UDI assignment, and post-market surveillance.

  • Transitional provisions: MDR applies from 26 May 2021, IVDR from 26 May 2022. Legacy certificates under MDD 93/42/EEC and AIMDD 90/385/EEC were extended by Regulation (EU) 2023/607 under conditions tied to application lodgement and notified body agreements.

  • MDCG guidance documents are not legislation; they represent the views of the Medical Device Coordination Group and are persuasive but non-binding. Notified bodies and competent authorities use them as assessment benchmarks.

  • MDCG guidance is structured by topic: classification, conformity assessment, clinical evaluation, PMS/vigilance, UDI/EUDAMED, and IVD-specific documents. Document numbers (e.g., MDCG 2021-24) are the citation key.

  • Economic operators are manufacturer, authorised representative, importer, and distributor. Each has distinct verification duties under MDR Articles 11-14 and IVDR Articles 11-14.

  • A manufacturer established outside the EU must appoint an authorised representative before placing a device on the Union market; the mandate must cover regulatory liaison and incident reporting support.

  • The MDR/IVDR do not cover medicinal products, human tissues/cells of non-device origin, or cosmetic products; borderline products follow MDCG 2022-5 (MDR) and MDCG 2022-6 (borderline IVDs).

Quick reference

Device Classification

  • MDR classification uses Annex VIII: 22 rules grouped by invasiveness, duration, and anatomy. Class I, IIa, IIb, III; Class I subdivides into Is (sterile), Im (measuring), and Ir (reusable surgical instruments).

  • Duration definitions: transient <60 minutes; short-term 60 minutes to 30 days; long-term >30 days. Duration and invasiveness together drive the class.

  • Rule 11 (software) classifies software intended to provide information for diagnostic or therapeutic decisions as Class IIa; if the decision may cause death or irreversible deterioration, Class IIb; if death or irreversible deterioration is likely, Class III.

  • Rule 10 (active therapeutic devices) and Rule 9 (active therapeutic devices administering or exchanging energy) are the primary active-device rules; Rule 9 typically yields Class IIa/IIb depending on hazardous energy.

  • IVDR classification uses Annex VIII (IVDR) with Rules 1-7: Class A (lowest), B, C, D (highest). Rule 3 covers transmissible agents; Rule 4 covers blood grouping; Rule 5 covers IVDs for cancer, genetic testing, or monitoring of medicinal products.

  • Class A IVDs that are not sterile and have no measuring function are self-certified (no notified body). Class A sterile devices require notified body involvement for the sterility aspect only.

  • Classification is the manufacturer's responsibility but is verified by the notified body during conformity assessment; misclassification is a common cause of non-conformity findings.

  • MDCG 2021-24 provides classification guidance for MDR; MDCG 2020-16 provides IVDR classification guidance. Both include decision trees and examples.

Quick reference

Conformity Assessment Routes and Notified Body Involvement

  • Class I non-sterile, non-measuring devices: manufacturer self-certifies and issues the EU Declaration of Conformity; no notified body required under MDR Annex II and IV.

  • Class IIa: Annex IX (QMS + technical documentation assessment) or Annex XI Part A (production quality assurance) with Annex IX Section 4 technical documentation sampling.

  • Class IIb: Annex IX (QMS + technical documentation assessment) or Annex XI Part A (production quality assurance) plus Annex IX Section 4; for implantable Class IIb, Annex X (type-examination) may be combined.

  • Class III: Annex IX (QMS + technical documentation assessment) or Annex X (type-examination) plus Annex XI Part B (product quality assurance). Annex IX requires assessment of each technical documentation.

  • IVDR Class A (non-sterile, no measuring function): self-certification. Class A sterile, Class B, C, D require notified body involvement; Class D requires reference laboratory testing under IVDR Article 48(4).

  • Notified body designation is under MDR Article 42 / IVDR Article 38; NANDO database lists designated bodies and their scopes. Scope must cover the device type and annex route used.

  • A notified body assessment includes QMS audit, technical documentation review, and where applicable, unannounced audits and product testing. The certificate states the annex route and device scope.

  • The manufacturer must lodge an application with a notified body before the certificate expiry under Regulation (EU) 2023/607; written agreement must be signed by 26 September 2024 for the extension to apply.

Quick reference

Technical Documentation

  • MDR Annex II and IVDR Annex II define technical documentation content: device description, labelling, design and manufacturing information, GSPR checklist, benefit-risk analysis, product verification and validation, and clinical/performance evaluation.

  • The GSPR checklist (MDR Annex I / IVDR Annex I) must address every applicable requirement and justify any non-applicability with a rationale.

  • Technical documentation must be kept for at least 10 years after the last device is placed on the market (MDR Article 10(8)); IVDs also 10 years, or 10 years after the last IVD manufactured.

  • Post-market surveillance plan, PSUR (Class IIa/IIb/III and IVD B/C/D), and PMCF plan are part of the technical documentation under MDR Annex III and IVDR Annex III.

  • The technical documentation must demonstrate conformity with the GSPRs, including risk management per ISO 14971 and, where applicable, IEC 62304, IEC 62366, and IEC 60601 series.

  • The Declaration of Conformity (MDR Annex IV / IVDR Annex IV) must be kept up to date and reference the applicable regulation, class, and notified body certificate number.

  • Technical documentation must be provided to the notified body and, on request, to competent authorities; it must be available in an official EU language as required by the Member State.

  • For devices incorporating software, the documentation must include software description, architecture, verification and validation evidence, and cybersecurity documentation per MDCG 2019-16.

Quick reference

Annexes and Pre-Registration Submission Requirements

  • MDR Annex I: General Safety and Performance Requirements (GSPRs), 23 chapters. IVDR Annex I: GSPRs for IVDs, 20 chapters.

  • MDR Annex II: technical documentation; Annex III: technical documentation on post-market surveillance; Annex IV: EU Declaration of Conformity; Annex V: CE marking; Annex VI: UDI; Annex VII: notified bodies; Annex VIII: classification; Annex IX-XI: conformity assessment; Annex XII-XV: certificates and miscellaneous.

  • IVDR Annex II: technical documentation; Annex III: PMS technical documentation; Annex IV: EU Declaration of Conformity; Annex V: CE marking; Annex VI: UDI; Annex VII: notified bodies; Annex VIII: classification; Annex IX-XI: conformity assessment.

  • Pre-registration submission typically requires: device description, classification rationale, GSPR checklist, risk management file, clinical evaluation report, PMS plan, and UDI-DI assignment.

  • The notified body application form must specify the annex route, device scope, and manufacturing sites; incomplete applications are rejected at the completeness check.

  • The EU Declaration of Conformity must be signed and dated; a draft DoC is not acceptable for certificate issuance.

  • For Class III and implantable devices, the notified body must review the clinical evaluation assessment report (CEAR) and may consult the expert panel under MDR Article 54.

  • IVDR Class D devices require a reference laboratory opinion under Article 48(4) before the notified body issues the certificate.

Quick reference

Clinical Evaluation, Clinical Investigation and Performance Evaluation

  • MDR Article 61 requires clinical evaluation for all devices; clinical investigation is required where clinical evidence is insufficient, except for certain well-established technologies.

  • Clinical evaluation follows MDR Annex XIV Part A: identify clinical evidence, appraise it, and generate new evidence if needed. The clinical evaluation report (CER) must be updated throughout the lifecycle.

  • Clinical investigation is governed by MDR Articles 62-80 and Annex XV; it requires an ethics committee opinion and competent authority authorisation in each Member State.

  • IVDR Article 56 requires performance evaluation; IVDR Annex XIII Part A covers scientific validity, analytical performance, and clinical performance.

  • IVDR performance studies are governed by Articles 57-77 and Annex XIII Part B; they require ethics committee and competent authority approval.

  • PMCF (MDR Annex XIV Part B) and PMPF (IVDR Annex XIII Part B) are required for Class IIa and above (and IVD B and above) unless justified as not applicable.

  • The clinical evaluation must be based on a systematic literature search with documented methodology, inclusion/exclusion criteria, and appraisal per MDCG 2020-1 and MDCG 2020-6.

  • Equivalence claims require demonstration of technical, biological, and clinical equivalence per MDR Annex XIV Section 3; the equivalence data must be contractually accessible.

Quick reference

Post-Market Surveillance, Vigilance and PMCF

  • MDR Article 83-86 and IVDR Article 78-81 require a PMS system proportionate to the device risk class; the PMS plan is part of the technical documentation.

  • PSUR (Periodic Safety Update Report) is required for Class IIa, IIb, III, and IVD B, C, D; frequency: Class IIa every 2 years, Class IIb/III annually, IVD B/C every 2 years, IVD D annually.

  • Serious incident reporting under MDR Article 87: 15 days for serious incidents, 2 days for serious public health threats, and 10 days for device deficiencies that could lead to serious incidents.

  • Field Safety Corrective Action (FSCA) must be reported and a Field Safety Notice (FSN) issued; the FSN must be reviewed by the competent authority where required.

  • Trend reporting (MDR Article 88) applies to statistically significant increases in non-serious incidents or expected undesirable side-effects; the manufacturer must report the trend.

  • Vigilance reporting uses the Manufacturer's Incident Report (MIR) form and EUDAMED; until EUDAMED is fully functional, national reporting channels apply per MDCG 2021-1.

  • PMCF must be part of the PMS plan; PMCF results feed back into the clinical evaluation and risk management file.

  • The PMS system must be audited by the notified body; the PSUR and PMS plan are reviewed during surveillance audits.

Quick reference

Quality Management System (ISO 13485, ISO 14971) and Supplier Controls

  • ISO 13485:2016 is the harmonised QMS standard for medical devices; certification is not mandatory but is the practical route to demonstrating MDR Annex IX QMS conformity.

  • ISO 14971:2019 is the risk management standard; MDR Annex I GSPR 1-3 require a risk management system throughout the device lifecycle.

  • Supplier controls require: supplier evaluation and selection, documented purchasing information, verification of purchased product, and a supplier monitoring programme.

  • Critical suppliers (e.g., contract manufacturers, sterilisation providers) must be audited; the audit frequency is risk-based and documented in the QMS.

  • Outsourced processes remain the manufacturer's responsibility; a written quality agreement is required with critical suppliers.

  • The QMS must cover design controls, document control, record control, CAPA, non-conforming product, and complaint handling.

  • ISO 13485 certification is verified by the notified body during the QMS audit; the certificate scope must cover the device types and manufacturing sites.

  • Risk management must be integrated with the QMS; the risk management file must be maintained and updated with PMS data.

Quick reference

Unique Device Identification (UDI), EUDAMED Registration and Data Governance

  • MDR Article 27 and IVDR Article 24 require UDI assignment; the UDI consists of a UDI-DI (device identifier) and UDI-PI (production identifier).

  • UDI must be placed on the label and, for reusable devices, on the device itself; Class III implantable devices require direct marking.

  • UDI-DI must be registered in EUDAMED before placing the device on the market; the Basic UDI-DI is the primary key for device grouping.

  • EUDAMED has six modules: actor registration, UDI/device registration, notified bodies and certificates, clinical investigations and performance studies, vigilance, and market surveillance.

  • EUDAMED actor registration is mandatory for manufacturers, authorised representatives, importers, and distributors before they can register devices.

  • Data governance: the manufacturer is responsible for the accuracy of EUDAMED data; changes must be updated within the required timelines.

  • UDI issuing entities are designated by the European Commission; GS1, HIBCC, ICCBBA, and IFA are the designated entities.

  • The Basic UDI-DI is used for device grouping and must be consistent across the technical documentation, DoC, and EUDAMED.

Quick reference

MDCG Guidance Documents and Their Application

  • MDCG 2021-24: classification of medical devices under MDR; includes decision trees and examples for each rule.

  • MDCG 2020-1: clinical evaluation of medical devices; MDCG 2020-6: clinical evidence for legacy devices; MDCG 2020-5: equivalence; MDCG 2020-7: PMCF plan; MDCG 2020-8: PMCF evaluation report.

  • MDCG 2019-16: cybersecurity for medical devices; MDCG 2019-11: qualification and classification of software; MDCG 2022-14: transition to MDR.

  • MDCG 2021-1: vigilance reporting; MDCG 2022-5: borderline products; MDCG 2022-6: borderline IVDs; MDCG 2021-5: clinical investigation of devices.

  • MDCG 2022-21: performance evaluation of IVDs; MDCG 2022-9: performance study applications; MDCG 2023-3: IVDR classification questions.

  • MDCG guidance is not legally binding but is used by notified bodies and competent authorities as the assessment benchmark; deviation must be justified.

  • Guidance documents are published on the European Commission's MDCG page; the document number and revision date must be cited in the technical documentation.

  • MDCG position papers (e.g., MDCG 2022-14 on transition) clarify interpretation but do not create new legal obligations.