MDCG Pre-Registration Assessment study materials
MDCG Pre-Registration Assessment study resources

MDCG Pre-Registration Assessment study plan

MDCG Pre-Registration Assessment · Give each week a clear focus.

8 weeks · About 6 hours per week

Week 1

EU Medical Device Regulatory Framework (MDR 2017/745, IVDR 2017/746, MDCG Guidance Structure)

  • Map the structure of MDR 2017/745 article by article, writing a one-line purpose statement for each chapter and annex from memory after a closed-book pass.

  • Distinguish the roles and legal responsibilities of economic operators (manufacturer, authorised representative, importer, distributor) and state who carries which obligation under Articles 10-14.

  • Trace how MDCG guidance documents are produced, endorsed and published, and explain their legal status relative to the Regulation text.

  • Compare MDR and IVDR scope, definitions and transitional provisions in a two-column table covering at least ten divergence points.

Your checkpoint

A one-page annotated map of MDR chapters and annexes plus a 10-row MDR-versus-IVDR divergence table, both produced without reference material.

Week 2

EU Medical Device Regulatory Framework (MDR 2017/745, IVDR 2017/746, MDCG Guidance Structure) & MDCG Guidance Documents, Position Papers and Their Application in Assessment

  • Locate and summarise the MDCG guidance most relevant to pre-registration assessment, noting for each the document number, issue date and the assessment question it answers.

  • Explain how MDCG position papers and Q&A documents are used by notified bodies and competent authorities when the Regulation text is ambiguous.

  • Reproduce the classification rules of MDR Annex VIII for the classes most often disputed, and justify a classification decision for three worked device examples.

  • Identify where MDCG guidance has been superseded or amended and describe how to verify you are reading the current version.

Your checkpoint

A guidance register of at least 12 MDCG documents with number, date, status (current/superseded) and a one-sentence application note each.

Week 3

Device Classification, Conformity Assessment Routes and Notified Body Involvement

  • Apply the MDR Annex VIII classification rules to at least 15 practice devices spanning Classes I, IIa, IIb and III, and record the decisive rule for each.

  • Match each device class to its conformity assessment route under Articles 52 and Annexes IX, X and XI, including when a notified body is mandatory.

  • Explain the Class I carve-outs (sterile, measuring function, reusable surgical instruments) and why they pull a device into notified body involvement.

  • Describe the notified body's role in the pre-registration phase: application, scoping, quotation, and the boundary between assessment and certification.

Your checkpoint

A classification and route matrix covering 15 devices with class, decisive Annex VIII rule, chosen conformity assessment annex and whether a notified body is required.

Week 4

Technical Documentation, Annexes and Pre-Registration Submission Requirements

  • Reconstruct the Annex II technical documentation structure and state what evidence satisfies each section for a representative Class IIb device.

  • Explain the role of the GSPR checklist (Annex I) and how each GSPR is evidenced, including the standards presumption of conformity under Article 8.

  • Draft the pre-registration submission package checklist a notified body expects, covering application form, declaration, technical documentation and classification rationale.

  • Identify the common causes of pre-registration deficiency letters and describe how to pre-empt each in the submission.

Your checkpoint

A complete mock pre-registration submission checklist for one fictional Class IIb device, with each Annex II section mapped to a specific evidence item.

Week 5

Clinical Evaluation, Clinical Investigation and Performance Evaluation Requirements

  • Outline the clinical evaluation process under Article 61 and Annex XIV, including the equivalence route and when it is defensible.

  • Distinguish clinical evaluation, clinical investigation and PMCF, and state the documentation each produces.

  • Explain the performance evaluation requirements under IVDR Article 56 and Annex XIII, including scientific validity, analytical performance and clinical performance.

  • Summarise the content and purpose of the clinical evaluation assessment report (CEAR) and how notified bodies use it in the pre-registration dialogue.

Your checkpoint

A one-page clinical evidence plan for the fictional device, stating the chosen evidence route, justification, and the documents that will be submitted.

Week 6

Post-Market Surveillance, Vigilance and Post-Market Clinical Follow-Up

  • Build the PMS system per Article 83 and Annex III, listing the inputs, outputs and review cadence for each PMS element.

  • Explain the vigilance workflow under Articles 87-92, including serious incident reporting timelines, trend reporting and field safety corrective actions.

  • Design a PMCF plan under Annex XIV Part B that specifies methods, endpoints, sample size rationale and linkage to the clinical evaluation.

  • Describe how PSUR and periodic safety update requirements differ by device class and what triggers an update.

Your checkpoint

A PMS and PMCF plan outline for the fictional device containing at least six PMS data sources, reporting timelines and three PMCF methods with endpoints.

Week 7

Quality Management System, Risk Management (ISO 13485, ISO 14971) and Supplier Controls & Unique Device Identification (UDI), EUDAMED Registration and Data Governance

  • Map ISO 13485:2016 clauses to MDR Annex IX quality system requirements and flag the clauses that need MDR-specific extension.

  • Apply ISO 14971 risk management to the fictional device, producing a hazard analysis with risk acceptability decisions and residual risk justification.

  • Explain supplier controls: selection, evaluation, monitoring and the written agreement requirements for critical suppliers and subcontractors.

  • Describe UDI assignment under Article 27, the Basic UDI-DI concept, and how device data is entered and governed in EUDAMED.

Your checkpoint

A combined artefact: a 10-row ISO 13485-to-MDR clause mapping table plus a hazard analysis of at least eight hazards with risk control measures and a draft UDI-DI assignment.

Week 8

Technical Documentation, Annexes and Pre-Registration Submission Requirements & MDCG Guidance Documents, Position Papers and Their Application in Assessment

  • Integrate all prior deliverables into one coherent pre-registration dossier for the fictional device and check internal consistency across classification, clinical, PMS and UDI sections.

  • Re-read the guidance register from week 2 and state, for each document, exactly where it shaped a decision in the dossier.

  • Complete a timed 60-question mock assessment weighted by domain percentage and score at least 75 percent.

  • Review every mock question answered incorrectly and write a corrected one-sentence rule for each.

Your checkpoint

An integrated mock pre-registration dossier plus a scored mock assessment with a written error log covering every missed question.

Milestones

Week 2: Can name at least 12 MDCG documents with correct numbers and dates and explain the legal weight of guidance versus the Regulation text, without notes.

Week 4: Can classify any given device under Annex VIII and name the required conformity assessment route and notified body involvement in under five minutes per device.

Week 6: Can explain the full clinical and PMS evidence chain from clinical evaluation through PMCF to vigilance reporting in a single uninterrupted five-minute verbal walkthrough.

Week 8: Scores at least 75 percent on the weighted 60-question mock and has a written correction for every missed item.