- MDR application date
- The Medical Device Regulation (EU) 2017/745 has applied since 26 May 2021, replacing the MDD 93/42/EEC and AIMDD 90/385/EEC.
- IVDR application date
- The In Vitro Diagnostic Regulation (EU) 2017/746 has applied since 26 May 2022, replacing IVDD 98/79/EC.
- MDCG guidance legal status
- MDCG guidance is non-binding but is used by notified bodies and competent authorities as the assessment benchmark; deviations must be justified in the technical documentation.
- Authorised representative mandate
- A non-EU manufacturer must appoint an EU authorised representative whose written mandate covers regulatory liaison, incident reporting support, and access to technical documentation.
- MDR Rule 11 software
- Software providing information for diagnostic or therapeutic decisions is Class IIa; if death or irreversible deterioration is possible, Class IIb; if likely, Class III.
- IVDR Rule 3 transmissible agents
- IVDs intended to detect, diagnose, or monitor transmissible agents are Class D when they detect a life-threatening disease with high or suspected high risk of propagation.
- Class I sterile subclass
- Class I devices that are sterile are subclass Is and require notified body involvement for the sterility aspect only; the manufacturer remains responsible for the rest.
- Duration categories
- Transient is less than 60 minutes; short-term is 60 minutes to 30 days; long-term is more than 30 days. Duration combines with invasiveness to set the class.
- Class IIa conformity route
- Class IIa devices use Annex IX (QMS plus technical documentation assessment) or Annex XI Part A (production quality assurance) with Annex IX Section 4 sampling.
- Class III conformity route
- Class III devices use Annex IX (QMS plus technical documentation assessment of each device) or Annex X type-examination combined with Annex XI Part B.
- IVDR Class D reference laboratory
- IVDR Class D devices require a reference laboratory opinion under Article 48(4) before the notified body issues the certificate.
- Notified body scope check
- The notified body's NANDO designation scope must cover the device type and the conformity assessment annex route used; otherwise the certificate is invalid.
- GSPR checklist requirement
- The MDR Annex I and IVDR Annex I GSPR checklist must address every applicable requirement and justify each non-applicability with a documented rationale.
- Technical documentation retention
- Technical documentation must be kept for at least 10 years after the last device is placed on the market, or 10 years after the last IVD is manufactured.
- Declaration of Conformity content
- The EU Declaration of Conformity must identify the regulation, device class, manufacturer, notified body certificate number, and be signed and dated.
- Software documentation standard
- MDCG 2019-16 requires software description, architecture, verification and validation evidence, and cybersecurity documentation in the technical documentation.
- MDR Annex VIII classification
- MDR Annex VIII contains 22 classification rules grouped by invasiveness, duration, and anatomy; Class I subdivides into Is, Im, and Ir.
- IVDR Annex VIII classification
- IVDR Annex VIII contains Rules 1-7 classifying IVDs into Class A, B, C, and D, with Rule 3 covering transmissible agents and Rule 4 blood grouping.
- Class III expert panel
- Under MDR Article 54, the notified body must review the clinical evaluation assessment report for Class III and implantable devices and may consult the expert panel.
- Clinical evaluation report update
- The CER must be updated throughout the device lifecycle using PMS and PMCF data, per MDR Annex XIV Part A and MDCG 2020-1.
- Equivalence three pillars
- Equivalence requires technical, biological, and clinical equivalence per MDR Annex XIV Section 3, with contractual access to the equivalence data.
- IVDR performance evaluation
- IVDR Article 56 requires performance evaluation covering scientific validity, analytical performance, and clinical performance per Annex XIII Part A.
- Clinical investigation authorisation
- Clinical investigations require an ethics committee opinion and competent authority authorisation in each Member State under MDR Articles 62-80 and Annex XV.
- PSUR frequency
- PSUR is required for Class IIa every 2 years, Class IIb and III annually, IVD B and C every 2 years, and IVD D annually.
- Serious incident reporting deadline
- Serious incidents must be reported within 15 days; serious public health threats within 2 days; device deficiencies that could lead to serious incidents within 10 days.
- Trend reporting obligation
- Under MDR Article 88, statistically significant increases in non-serious incidents or expected undesirable side-effects must be reported as a trend.
- ISO 13485 status
- ISO 13485:2016 is the harmonised QMS standard for medical devices; certification is not mandatory but is the practical route to MDR Annex IX QMS conformity.
- ISO 14971 risk management
- ISO 14971:2019 requires a risk management system covering hazard identification, risk estimation, evaluation, control, and residual risk evaluation throughout the lifecycle.
- Critical supplier audit
- Critical suppliers such as contract manufacturers and sterilisation providers must be audited on a risk-based frequency documented in the QMS.
- Basic UDI-DI role
- The Basic UDI-DI is the primary key for device grouping in EUDAMED and must be consistent across the technical documentation, DoC, and EUDAMED registration.
- EUDAMED six modules
- EUDAMED comprises actor registration, UDI/device registration, notified bodies and certificates, clinical investigations, vigilance, and market surveillance modules.
- MDCG 2021-24 scope
- MDCG 2021-24 provides classification guidance for medical devices under MDR Annex VIII, including decision trees and examples for each rule.