HKIE Professional Assessment - Biomedical Discipline flashcards
HKIE Professional Assessment - Biomedical Discipline · Build recall with a focused review session.
31 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Stress and strain
- Stress is force per unit area (Pa); strain is fractional change in length. The slope of the linear stress-strain region is the elastic modulus, a key implant design parameter.
- Wolff's law
- Bone remodels in response to mechanical load, aligning trabeculae along principal stress directions. Engineers use this to design implants that avoid stress shielding and bone resorption.
- Poiseuille flow
- Laminar flow rate through a tube is proportional to the fourth power of radius and pressure gradient, and inversely to viscosity and length. Small lumen reductions cause large flow drops.
- Foreign body response
- Implanted materials trigger protein adsorption, neutrophil and macrophage activity, giant cell formation, and fibrous encapsulation. Surface chemistry and topography influence the response.
- Nyquist criterion
- Sampling frequency must be at least twice the highest signal frequency to avoid aliasing. Anti-alias filters must attenuate higher-frequency components before the ADC.
- ECG bandwidth
- Diagnostic ECG requires 0.05–150 Hz bandwidth and at least 500 Hz sampling. Monitoring ECG may use 0.5–40 Hz, which distorts ST segments.
- Defibrillation energy
- Stored energy in a defibrillator is E = ½CV². Monophasic shocks use 200–360 J; biphasic shocks use lower energy, typically 120–200 J, with comparable efficacy.
- Patient leakage current
- IEC 60601-1 limits patient leakage current in normal condition to 10 µA for type CF applied parts and 100 µA for type B and BF applied parts.
- Acceptance testing
- Verification that a newly delivered device meets manufacturer specification and purchase requirements before clinical use. Results are recorded in the equipment history file.
- Preventive maintenance interval
- PM frequency is set by manufacturer recommendation, equipment risk score, utilisation, and incident history. Typical intervals are 6 or 12 months for clinical devices.
- MTBF and MTTR
- Mean time between failures is total operating time divided by failure count. Mean time to repair is total repair time divided by repair count; both are HTM performance indicators.
- ISO 14971 risk process
- Risk management identifies hazards, estimates and evaluates risk, applies controls in a hierarchy, evaluates residual risk, and monitors production and post-production information.
- IEC 62304 safety classes
- Software safety class A means no injury is possible, B means non-serious injury is possible, and C means death or serious injury is possible. Class drives lifecycle rigour.
- HK MDACS
- The Medical Device Administrative Control System is Hong Kong's voluntary listing scheme, administered by the Medical Device Division of the Department of Health, covering devices and traders.
- Pan-Tompkins algorithm
- A QRS detection method using band-pass filtering (5–15 Hz), differentiation, squaring, moving-window integration, and adaptive thresholding to locate heartbeats in ECG.
- Hounsfield unit
- HU = 1000 × (µ_tissue − µ_water)/µ_water. Water is 0 HU and air is approximately −1000 HU; it standardises CT attenuation across scanners.
- CMRR
- Common-mode rejection ratio is 20 log₁₀ of the differential gain divided by common-mode gain. High CMRR rejects interference such as mains pickup in biopotential recording.
- HL7 FHIR
- A RESTful interoperability standard using resources such as Patient, Observation, and DiagnosticReport. It supports modern web APIs and is increasingly used alongside HL7 v2.
- PDPO six principles
- Hong Kong's Personal Data (Privacy) Ordinance sets six Data Protection Principles covering purpose, accuracy, retention, security, openness, and access/correction rights.
- PPV and prevalence
- Positive predictive value depends on disease prevalence. A highly specific test can still have low PPV when prevalence is low, which matters for screening decisions.
- Tissue engineering triad
- Successful tissue engineering combines cells, a scaffold that supports structure and degradation, and biochemical or mechanical signals to guide tissue formation.
- Scaffold porosity
- Interconnected porosity of roughly 70–90% with pore sizes around 100–500 µm supports cell infiltration, vascularisation, and bone ingrowth in orthopaedic scaffolds.
- GMP cell processing
- Cell and tissue products require validated aseptic processing, often in ISO 5 cleanrooms, with chain of identity and custody and release testing before clinical use.
- Duty of care
- Engineers must act within their competence and prioritise public safety. When a task exceeds expertise, they must refer to a qualified specialist and document the decision.
- Conflict of interest
- Financial or personal interests that could bias professional judgement, such as vendor gifts or shares in a supplier, must be declared and managed under employer policy.
- Earned value CPI
- Cost performance index is earned value divided by actual cost. A CPI below 1 indicates cost overrun; SPI below 1 indicates schedule delay.
- Healthcare facility interfaces
- Medical gas, essential electrical supply, HVAC pressure regimes, shielding, and IT segregation must be coordinated with clinical equipment planning during design and commissioning.
- Whole-life carbon
- Embodied carbon from manufacture, transport, and disposal plus operational carbon from energy and consumables. Single-use devices often dominate the footprint of a procedure.
- AI as a medical device
- Machine-learning software intended for diagnosis or treatment is regulated as software as a medical device, requiring validation, change control, and post-market performance monitoring.
- Galvanic corrosion
- Contact between dissimilar metals in an electrolyte creates a galvanic couple that accelerates corrosion. Implant designs avoid mixing stainless steel and titanium components.
- Oscillometric NIBP
- Cuff pressure oscillations are analysed to estimate mean arterial pressure, from which systolic and diastolic values are derived using an algorithm. Motion reduces accuracy.