Stress and strain
Stress is force per unit area (Pa); strain is fractional change in length. The slope of the linear stress-strain region is the elastic modulus, a key implant design parameter.
Wolff's law
Bone remodels in response to mechanical load, aligning trabeculae along principal stress directions. Engineers use this to design implants that avoid stress shielding and bone resorption.
Poiseuille flow
Laminar flow rate through a tube is proportional to the fourth power of radius and pressure gradient, and inversely to viscosity and length. Small lumen reductions cause large flow drops.
Foreign body response
Implanted materials trigger protein adsorption, neutrophil and macrophage activity, giant cell formation, and fibrous encapsulation. Surface chemistry and topography influence the response.
Nyquist criterion
Sampling frequency must be at least twice the highest signal frequency to avoid aliasing. Anti-alias filters must attenuate higher-frequency components before the ADC.
ECG bandwidth
Diagnostic ECG requires 0.05–150 Hz bandwidth and at least 500 Hz sampling. Monitoring ECG may use 0.5–40 Hz, which distorts ST segments.
Defibrillation energy
Stored energy in a defibrillator is E = ½CV². Monophasic shocks use 200–360 J; biphasic shocks use lower energy, typically 120–200 J, with comparable efficacy.
Patient leakage current
IEC 60601-1 limits patient leakage current in normal condition to 10 µA for type CF applied parts and 100 µA for type B and BF applied parts.
Acceptance testing
Verification that a newly delivered device meets manufacturer specification and purchase requirements before clinical use. Results are recorded in the equipment history file.
Preventive maintenance interval
PM frequency is set by manufacturer recommendation, equipment risk score, utilisation, and incident history. Typical intervals are 6 or 12 months for clinical devices.
MTBF and MTTR
Mean time between failures is total operating time divided by failure count. Mean time to repair is total repair time divided by repair count; both are HTM performance indicators.
ISO 14971 risk process
Risk management identifies hazards, estimates and evaluates risk, applies controls in a hierarchy, evaluates residual risk, and monitors production and post-production information.
IEC 62304 safety classes
Software safety class A means no injury is possible, B means non-serious injury is possible, and C means death or serious injury is possible. Class drives lifecycle rigour.
HK MDACS
The Medical Device Administrative Control System is Hong Kong's voluntary listing scheme, administered by the Medical Device Division of the Department of Health, covering devices and traders.
Pan-Tompkins algorithm
A QRS detection method using band-pass filtering (5–15 Hz), differentiation, squaring, moving-window integration, and adaptive thresholding to locate heartbeats in ECG.
Hounsfield unit
HU = 1000 × (µ_tissue − µ_water)/µ_water. Water is 0 HU and air is approximately −1000 HU; it standardises CT attenuation across scanners.
CMRR
Common-mode rejection ratio is 20 log₁₀ of the differential gain divided by common-mode gain. High CMRR rejects interference such as mains pickup in biopotential recording.
HL7 FHIR
A RESTful interoperability standard using resources such as Patient, Observation, and DiagnosticReport. It supports modern web APIs and is increasingly used alongside HL7 v2.
PDPO six principles
Hong Kong's Personal Data (Privacy) Ordinance sets six Data Protection Principles covering purpose, accuracy, retention, security, openness, and access/correction rights.
PPV and prevalence
Positive predictive value depends on disease prevalence. A highly specific test can still have low PPV when prevalence is low, which matters for screening decisions.
Tissue engineering triad
Successful tissue engineering combines cells, a scaffold that supports structure and degradation, and biochemical or mechanical signals to guide tissue formation.
Scaffold porosity
Interconnected porosity of roughly 70–90% with pore sizes around 100–500 µm supports cell infiltration, vascularisation, and bone ingrowth in orthopaedic scaffolds.
GMP cell processing
Cell and tissue products require validated aseptic processing, often in ISO 5 cleanrooms, with chain of identity and custody and release testing before clinical use.
Duty of care
Engineers must act within their competence and prioritise public safety. When a task exceeds expertise, they must refer to a qualified specialist and document the decision.
Conflict of interest
Financial or personal interests that could bias professional judgement, such as vendor gifts or shares in a supplier, must be declared and managed under employer policy.
Earned value CPI
Cost performance index is earned value divided by actual cost. A CPI below 1 indicates cost overrun; SPI below 1 indicates schedule delay.
Healthcare facility interfaces
Medical gas, essential electrical supply, HVAC pressure regimes, shielding, and IT segregation must be coordinated with clinical equipment planning during design and commissioning.
Whole-life carbon
Embodied carbon from manufacture, transport, and disposal plus operational carbon from energy and consumables. Single-use devices often dominate the footprint of a procedure.
AI as a medical device
Machine-learning software intended for diagnosis or treatment is regulated as software as a medical device, requiring validation, change control, and post-market performance monitoring.
Galvanic corrosion
Contact between dissimilar metals in an electrolyte creates a galvanic couple that accelerates corrosion. Implant designs avoid mixing stainless steel and titanium components.
Oscillometric NIBP
Cuff pressure oscillations are analysed to estimate mean arterial pressure, from which systolic and diastolic values are derived using an algorithm. Motion reduces accuracy.