HKIE Professional Assessment - Biomedical Discipline study plan
HKIE Professional Assessment - Biomedical Discipline · Give each week a clear focus.
8 weeks · About 7 hours per week
Biomedical engineering fundamentals (biomechanics, biomaterials, physiology for engineers)
Derive and apply the stress-strain relationships for cortical bone and a typical titanium alloy, then compute the axial stiffness of a simplified femoral stem under a 2 kN load
Explain the physiology of the cardiac cycle, glomerular filtration and pulmonary gas exchange in engineering terms, linking each to a measurable quantity such as pressure, flow or clearance
Classify biomaterials by degradation behaviour and contrast the host response to a permanent metal implant versus a resorbable polymer
Solve three worked biomechanics problems on joint reaction forces and beam bending of long bones without referring to notes
Your checkpoint
A one-page formula sheet plus three fully worked numerical problems with units checked, self-marked against textbook solutions
Medical devices and instrumentation (design, sensors, signal acquisition, safety)
Design a signal chain for an ECG front end, selecting electrode type, amplifier gain, filter cut-offs and sampling rate with stated justification
Calculate the leakage current and patient auxiliary current limits applicable to a Type CF applied part and identify which IEC 60601-1 clauses govern them
Compare optical, piezoelectric and capacitive sensing principles for SpO2, blood pressure and respiration respectively, stating one failure mode for each
Draw a block diagram of a closed-loop infusion pump safety architecture including independent over-infusion protection
Your checkpoint
A two-page instrument design note for one device, including block diagram, component selection table and a safety limits summary
Clinical engineering and healthcare technology management (maintenance, calibration, lifecycle) & Medical device standards, regulation and quality systems (ISO 13485, ISO 14971, IEC 60601, HK MDACS/MDD)
Build a preventive maintenance schedule for a ventilator and an electrosurgical unit, specifying intervals, tests and acceptance criteria
Perform a full ISO 14971 risk assessment on one device, producing a hazard list, severity and probability scoring, and at least three risk control measures traced to verification evidence
Map the clauses of ISO 13485 to the lifecycle stages of a device from design input to post-market surveillance
Explain how the Hong Kong MDACS listing process differs from the mainland NMPA and EU MDR routes, and state which documents each requires
Calibrate a simulated pressure transducer against a reference, record readings and compute measurement uncertainty
Your checkpoint
A risk management file excerpt for one device containing a hazard analysis table, risk control traceability and a calibration record with uncertainty budget
Medical device standards, regulation and quality systems (ISO 13485, ISO 14971, IEC 60601, HK MDACS/MDD) & Professional practice, ethics, safety and risk management in healthcare
Interpret the IEC 60601-1 third edition collateral and particular standards structure and identify which particular standard applies to three named device types
Analyse two published medical device incident reports, identifying the root cause, the violated control and the corrective action taken
Apply the four principles of biomedical ethics to a case where a clinician requests off-label use of a device, and justify a recommendation
Draft a health technology assessment recommendation for a new imaging modality using clinical, technical and economic criteria
Explain the professional obligations of an engineer signing off a medical device acceptance test in Hong Kong, including record retention and duty to report
Your checkpoint
A written ethics and incident analysis of 800 words covering two cases with a defensible recommendation for each
Biomedical signal and image processing, health informatics and data
Implement a digital filter in code to remove baseline wander and 50 Hz interference from a sampled ECG record, and report the signal-to-noise improvement
Compute and interpret the frequency spectrum of a biomedical signal, distinguishing physiological content from artefact
Explain the reconstruction pipeline of CT and MRI in engineering terms, including projection, back-projection, k-space and image quality metrics
Assess a health informatics scenario for data privacy obligations under the Personal Data (Privacy) Ordinance and propose an anonymisation approach
Evaluate the performance of a diagnostic classifier using sensitivity, specificity, ROC curve and positive predictive value at a stated prevalence
Your checkpoint
A reproducible analysis notebook or script with one filtered signal plot, one spectrum, one ROC curve and a short interpretation of each
Tissue engineering, regenerative medicine and advanced therapeutics engineering & Sustainability, emerging technologies and continuous professional development
Compare scaffold material options for a cartilage repair application on mechanical, degradation and biocompatibility criteria
Explain the regulatory pathway and evidence expectations for a combined cell-scaffold product, and identify where device and biological requirements diverge
Critically appraise one emerging technology such as surgical robotics or AI-based diagnosis, stating its maturity, evidence base and adoption barriers
Produce a personal continuing professional development plan aligned to the competencies the assessment tests, with specific activities and evidence
Your checkpoint
A two-page technology and CPD brief covering one regenerative medicine product and one emerging technology, with a personal 12-month development plan
Engineering management, project delivery and healthcare facility interfaces
Produce a project schedule for installing a new imaging suite, including dependencies, critical path, procurement lead times and commissioning milestones
Identify the facility interfaces for that installation, covering electrical supply, cooling, shielding, structural loading and infection control
Draft a stakeholder communication plan distinguishing clinical users, estates, infection control and finance
Quantify the cost of downtime for a critical device and justify a redundancy or service contract decision
Your checkpoint
A one-page project plan with Gantt summary, interface register and risk log for a simulated equipment installation
Biomedical engineering fundamentals (biomechanics, biomaterials, physiology for engineers) & Medical devices and instrumentation (design, sensors, signal acquisition, safety) & Clinical engineering and healthcare technology management (maintenance, calibration, lifecycle) & Medical device standards, regulation and quality systems (ISO 13485, ISO 14971, IEC 60601, HK MDACS/MDD) & Biomedical signal and image processing, health informatics and data & Tissue engineering, regenerative medicine and advanced therapeutics engineering & Professional practice, ethics, safety and risk management in healthcare & Engineering management, project delivery and healthcare facility interfaces & Sustainability, emerging technologies and continuous professional development
Sit a timed mock examination under exam conditions and mark it strictly against model answers
Re-derive the ten formulas most frequently needed across the domains from memory
Identify the three weakest domains from the mock result and close each gap with targeted practice
Rehearse concise verbal explanations of risk management, device lifecycle and professional ethics suitable for a viva
Your checkpoint
A marked mock examination with a scored breakdown by domain and a written remediation list for the three weakest areas
Milestones
Week 2: Can explain a full device signal chain and its safety limits without notes, and has completed the design note deliverable
Week 4: Can produce a hazard analysis and cite the applicable IEC 60601-1 standard for any named device; scoring at least 60 percent on a self-set 20-question standards and ethics quiz
Week 6: Can filter a signal, interpret an image reconstruction pipeline and appraise an emerging technology; all three week 5 and 6 deliverables complete
Week 8: Mock examination completed under time pressure with a domain-by-domain score; every domain at or above 50 percent and the three weakest areas documented with a remediation action