CLIA key terms
CLIA · Get comfortable with the vocabulary.
- CLIA Certificate of Waiver
- Related: Waived Test, Certificate of Compliance, CLIA Certificate of Accreditation
A certificate granted to a laboratory that performs only waived tests. It requires compliance with manufacturer instructions but exempts the lab from routine CLIA inspections and most personnel and proficiency testing requirements.
- Certificate of Compliance
- Related: Certificate of Accreditation, Substantial Compliance, CLIA Certificate of Waiver
A certificate issued to a laboratory once it demonstrates compliance with all applicable CLIA requirements during an initial survey. It authorizes the lab to perform moderate- and high-complexity testing and subjects it to routine inspections.
- CLIA Certificate of Accreditation
- Related: Deemed Status, Certificate of Compliance, CLIA Certificate of Waiver
A certificate granted to a laboratory accredited by a CMS-approved private organization. The lab must meet the accreditor's standards, which CMS deems equivalent to or stricter than CLIA, and is inspected through the accreditor.
- Waived Test
- Related: CLIA Certificate of Waiver, Moderate Complexity Test, High Complexity Test
A test categorized by FDA as simple and having a low risk of erroneous results, such as urine dipsticks or ovulation kits. Laboratories holding a Certificate of Waiver may perform these tests with minimal regulatory oversight.
- Moderate Complexity Test
- Related: High Complexity Test, Waived Test, Proficiency Testing
A test category between waived and high complexity, requiring a laboratory to hold a Certificate of Compliance or Accreditation and to meet CLIA requirements for personnel, quality control, and proficiency testing.
- High Complexity Test
- Related: Moderate Complexity Test, Technical Supervisor, Proficiency Testing
A test classified by FDA as complex, requiring the most stringent CLIA oversight, including qualified laboratory personnel, rigorous quality control, and participation in an approved proficiency testing program where available.
- Test Categorization
- Related: Waived Test, Moderate Complexity Test, High Complexity Test
The FDA process that assigns a test to waived, moderate, or high complexity based on criteria such as knowledge required, training, reagent preparation, and calibration. The category determines the lab's regulatory obligations.
- Laboratory Director
- Related: Technical Supervisor, General Supervisor, Testing Personnel
The individual with overall responsibility for a CLIA laboratory, including quality, personnel qualifications, and compliance. Required credentials vary by test complexity, ranging from a doctoral degree with experience to a high school diploma for waived labs.
- Technical Supervisor
- Related: Laboratory Director, General Supervisor, Proficiency Testing
A person responsible for the technical and scientific oversight of testing, including method selection, quality control review, and proficiency testing evaluation. Required qualifications depend on the complexity of testing performed.
- General Supervisor
- Related: Technical Supervisor, Testing Personnel, Laboratory Director
A role in high-complexity testing responsible for day-to-day supervision of testing personnel and reporting results. Must be accessible and meet CLIA-specified education or experience qualifications.
- Testing Personnel
- Related: General Supervisor, Laboratory Director, Moderate Complexity Test
Staff who perform specimen processing, testing, and result reporting. Their required education and training depend on the complexity of testing, with high-complexity testing demanding specific degrees or documented training.
- Proficiency Testing
- Related: PT Referral, Unsatisfactory PT Performance, Alternative Performance Assessment
External evaluation of a laboratory's performance using unknown samples sent by an approved program. Results are compared with peer laboratories, and poor performance can trigger corrective action or enforcement.
- PT Referral
- Related: Proficiency Testing, PT Attestation, Immediate Jeopardy
The prohibited practice of sending proficiency testing specimens to another laboratory for testing or sharing results between laboratories. Referral can result in certificate revocation and is a serious integrity violation.
- PT Attestation
- Related: PT Referral, Proficiency Testing, Laboratory Director
The signed statement by a laboratory director confirming that proficiency testing samples were tested in the same manner as patient specimens and that no interlaboratory communication or referral occurred.
- Unsatisfactory PT Performance
- Related: Proficiency Testing, Alternative Performance Assessment, Plan of Correction
A proficiency testing score below the acceptable threshold, typically less than 80 percent of samples for a test. Repeated unsatisfactory performance requires corrective action and may lead to sanctions.
- Alternative Performance Assessment
- Related: Proficiency Testing, Unsatisfactory PT Performance, Quality Assessment
A method used to evaluate a laboratory's testing performance when no approved proficiency testing program exists for a test. The laboratory must establish a documented system, such as split-sample comparison or calibration verification.
- Quality Control
- Related: Quality Assessment, Calibration Verification, Control Material
The set of procedures, including control materials and statistical rules, used during testing to detect errors and confirm that results are reliable before patient results are reported.
- Quality Assessment
- Related: Quality Control, Plan of Correction, Proficiency Testing
An ongoing, systematic review of laboratory processes and outcomes, including monitoring of quality indicators and corrective actions, to ensure accuracy and identify opportunities for improvement.
- Control Material
- Related: Quality Control, Calibration Verification, Levey-Jennings Chart
A substance with a known concentration used alongside patient samples to monitor test performance. Results are compared with acceptable ranges to detect shifts, trends, or random errors.
- Calibration Verification
- Related: Quality Control, Reportable Range, Verification of Performance Specifications
The process of confirming that a test system's calibration remains accurate across the reportable range. CLIA requires periodic verification, such as every six months for many tests, and after major changes.
- Verification of Performance Specifications
- Related: Calibration Verification, Reportable Range, Method Validation
The process by which a laboratory confirms that a manufacturer's stated accuracy, precision, and reportable range are acceptable for its own use before reporting patient results for a new test system.
- Reportable Range
- Related: Calibration Verification, Verification of Performance Specifications, Critical Value
The span of test values for which a laboratory has verified that results are accurate and reliable. Values outside this range may require dilution or should be reported with appropriate limitations.
- Critical Value
- Related: Result Reporting, Reportable Range, Quality Assessment
A test result that indicates a potentially life-threatening condition and requires immediate notification of the responsible clinician. Laboratories must have written procedures for timely reporting and documentation.
- Result Reporting
- Related: Critical Value, Test Requisition, Reportable Range
The process of communicating test results to the ordering provider, including required elements such as patient identification, test name, result, units, reference range, and any comments or limitations.
- Test Requisition
- Related: Result Reporting, Specimen Rejection, Specimen Collection
The order form or electronic request that authorizes testing. It must include patient identification, the test ordered, the ordering provider, and clinical information needed for accurate testing and interpretation.
- Specimen Collection
- Related: Specimen Rejection, Test Requisition, Chain of Custody
The process of obtaining a patient sample using procedures that ensure integrity, proper labeling, and correct handling. Errors at this stage are a leading cause of inaccurate results.
- Specimen Rejection
- Related: Specimen Collection, Test Requisition, Quality Assessment
The documented refusal to test a sample that fails acceptance criteria, such as inadequate volume, improper labeling, or compromised integrity. Rejection criteria must be written and consistently applied.
- Chain of Custody
- Related: Specimen Collection, Specimen Rejection, Result Reporting
A documented, unbroken record of specimen handling from collection through testing, used for forensic or legal testing. It records each transfer, date, time, and responsible person.
- Standard Precautions
- Related: Exposure Control Plan, Bloodborne Pathogen Exposure, Personal Protective Equipment
Infection control practices applied to all patients regardless of diagnosis, including hand hygiene, gloves, gowns, masks, and safe handling of sharps, to prevent transmission of bloodborne and other pathogens.
- Exposure Control Plan
- Related: Standard Precautions, Bloodborne Pathogen Exposure, Personal Protective Equipment
A written OSHA-required program describing how a laboratory protects workers from bloodborne pathogens, including engineering controls, work practice controls, personal protective equipment, and post-exposure procedures.
- Bloodborne Pathogen Exposure
- Related: Exposure Control Plan, Standard Precautions, Personal Protective Equipment
An incident in which a worker's skin, mucous membranes, or sharps injury contacts potentially infectious blood or body fluids. It requires immediate evaluation, source testing, and post-exposure prophylaxis as indicated.
- Personal Protective Equipment
- Related: Standard Precautions, Exposure Control Plan, Chemical Hygiene Plan
Barriers such as gloves, gowns, masks, and eye protection worn to prevent contact with infectious agents or hazardous materials. Selection depends on the task and anticipated exposure risk.
- Chemical Hygiene Plan
- Related: Personal Protective Equipment, Exposure Control Plan, Standard Precautions
A written OSHA laboratory standard requirement describing safe handling of hazardous chemicals, including storage, labeling, training, and emergency procedures, to protect laboratory staff from chemical hazards.
- Immediate Jeopardy
- Related: Plan of Correction, Substantial Compliance, PT Referral
A situation in which a laboratory's noncompliance has caused or is likely to cause serious injury, harm, or death to patients. It triggers expedited enforcement and possible immediate certificate suspension.
- Plan of Correction
- Related: Immediate Jeopardy, Substantial Compliance, Quality Assessment
A written, time-bound document describing how a laboratory will remedy deficiencies identified during inspection. It identifies responsible persons, corrective actions, and methods for monitoring sustained compliance.
- Substantial Compliance
- Related: Plan of Correction, Immediate Jeopardy, Certificate of Compliance
A finding that a laboratory meets CLIA requirements despite minor deficiencies that do not adversely affect patient care. It allows continued certification, often with a plan of correction for noted issues.
- Deemed Status
- Related: CLIA Certificate of Accreditation, Certificate of Compliance, Substantial Compliance
Recognition that a laboratory accredited by a CMS-approved organization is considered compliant with CLIA. The accreditor conducts inspections and reports results to CMS on the laboratory's behalf.
- Point-of-Care Testing
- Related: Waived Test, Testing Personnel, Quality Control
Testing performed near the patient, often by non-laboratory staff, using portable devices. It requires oversight, documented training, quality control, and a process for reporting results into the patient record.
- Levey-Jennings Chart
- Related: Control Material, Quality Control, Calibration Verification
A control chart that plots quality control results over time against mean and standard deviation limits, helping staff visually detect shifts, trends, and random errors in test performance.