Quick reference

Nuclear Medicine Instrumentation and Radiation Detection

  • Photomultiplier tube (PMT) gain rises steeply with applied high voltage; a small HV drift shifts energy-peak position, so daily photopeak checks detect gain instability.

  • Anger-camera position logic: each PMT signal is weighted by its position and summed; the ratio of weighted sums to total energy gives X and Y coordinates, while total energy gives the Z (pulse-height) signal.

  • Energy resolution is reported as FWHM divided by photopeak energy, expressed as a percent; for Tc-99m at 140 keV, a gamma camera typically achieves about 9-11% FWHM.

  • Sensitivity is counts per unit activity (cps/MBq or cpm/uCi) measured with a source in air; it is distinct from spatial resolution, which is measured as FWHM of a line-spread function in mm.

  • Uniformity is assessed with a flood field: integral uniformity is the maximum-minimum count difference over the whole field; differential uniformity uses a 5-pixel-wide moving window along each row and column.

  • Collimator choice sets both resolution and sensitivity: a low-energy high-resolution (LEHR) parallel-hole collimator gives finer resolution but lower sensitivity than a low-energy general-purpose (LEGP) collimator.

  • A parallel-hole collimator's resolution degrades with distance from the collimator face; a pinhole collimator magnifies small organs but its field of view shrinks as the aperture-to-organ distance decreases.

  • Dead-time losses follow a paralyzable (extending) or nonparalyzable model; the observed rate approaches a maximum and then falls with a paralyzable system, so decay-corrected count rates must be checked for linearity.

Quick reference

Radiation Safety, Dosimetry, and Regulatory Compliance

  • Dose equivalent: H = D x w_R (radiation weighting factor); 1 Sv = 100 rem. For gamma and beta radiation w_R = 1, so 1 Gy = 1 Sv for these radiations.

  • Effective dose E = sum over tissues of w_T x H_T; tissue weighting factors sum to 1.0, and the ICRP 103 values are used in current practice.

  • Inverse-square law: exposure rate falls as 1/r^2, so doubling distance from a point source cuts the exposure rate to one quarter.

  • Half-value layer (HVL) = 0.693/mu; shielding thickness for a required attenuation factor follows n = ln(attenuation)/ln(2) half-value layers.

  • Annual occupational effective dose limit is 50 mSv (5 rem); lens of eye 150 mSv (15 rem); skin and extremities 500 mSv (50 rem); declared pregnant worker 5 mSv (0.5 rem) to the fetus over the declared gestation.

  • Public dose limit is 1 mSv/y (100 mrem/y); a minor (under 18) occupational limit is 5 mSv/y (0.5 rem/y).

  • ALARA is implemented with time, distance, and shielding; a lead apron of 0.5 mm Pb attenuates Tc-99m 140 keV photons by roughly 90%.

  • A dose calibrator must be checked for constancy, accuracy (with a certified source such as Cs-137 or Co-57), linearity, and geometry; linearity is verified with a decay method or a set of calibrated attenuators.

  • Records of occupational dose are maintained on an NRC Form 5 or equivalent; a licensee must report a dose exceeding 250 mSv (25 rem) to the individual and the agency within the required timeframe.

Quick reference

Radiopharmacy and Radiopharmaceutical Quality Control

  • Tc-99m is eluted from a Mo-99/Tc-99m generator; the Mo-99 breakthrough limit in eluate is 0.15 kBq Mo-99 per MBq Tc-99m (0.15 uCi Mo-99 per mCi Tc-99m).

  • Alumina column breakthrough (aluminum ion) limit is 10 ug/mL of eluate; measured with a colorimetric assay or a commercial test kit.

  • Radionuclidic purity of eluate: Mo-99 breakthrough is tested at first elution and at the end of the generator's useful life; the generator's Mo-99 activity must not exceed 0.15 kBq per MBq of Tc-99m at the time of use.

  • Radiochemical purity is the fraction of total activity in the desired chemical form; it is measured by thin-layer chromatography, paper chromatography, or HPLC, and each radiopharmaceutical has a specified minimum (often 90-95%).

  • For Tc-99m MDP, radiochemical purity is checked with a chromatographic system that separates free pertechnetate and hydrolyzed-reduced technetium from the labeled complex; the labeled fraction must meet the package-insert limit.

  • A dose calibrator uses a well ionization chamber; constancy is checked daily, accuracy with a certified source, and geometry with a syringe versus vial of the same activity.

  • Radiopharmaceuticals must be tested for sterility (membrane filtration or direct inoculation) and bacterial endotoxin (LAL test); the 0.22 um filter is used for terminal filtration of Tc-99m products.

  • The expiration time of a Tc-99m kit is set by the manufacturer's instructions; beyond it, radiochemical purity may fall below the specification even though Tc-99m activity remains.

Quick reference

Clinical Nuclear Medicine Imaging Procedures

  • Myocardial perfusion: Tc-99m sestamibi or tetrofosmin, about 740-1110 MBq (20-30 mCi) per injection; redistribution is minimal, so rest and stress images are acquired separately.

  • Bone scan: Tc-99m MDP, about 740-925 MBq (20-25 mCi) adult dose; imaging begins 2-4 h after injection, and hydration plus voiding before imaging reduces bladder and pelvic artifact.

  • V/Q lung scan: Tc-99m DTPA aerosol or Technegas for ventilation and Tc-99m MAA (about 100-200 ug MAA, 100,000-500,000 particles) for perfusion; MAA particle count is limited in pulmonary hypertension.

  • Renal scan: Tc-99m MAG3 or DTPA for dynamic imaging; MAG3 is preferred when renal function is reduced, and furosemide challenge is used for suspected obstruction.

  • Thyroid uptake and scan: I-123 (about 7.5-14.8 MBq, 200-400 uCi) or Tc-99m pertechnetate (about 74-185 MBq); Tc-99m pertechnetate is trapped but not organified, so it cannot assess organification.

  • Hepatobiliary scan: Tc-99m mebrofenin or disofenin; gallbladder visualization within 60 min excludes acute cholecystitis, while a morphine-augmented study can shorten the wait.

  • Sentinel lymph node: Tc-99m sulfur colloid or filtered Tc-99m sulfur colloid injected around the tumor; imaging and gamma-probe counting localize the first draining node.

  • PET with F-18 FDG: about 185-370 MBq (5-10 mCi) adult dose; blood glucose should be controlled and the patient rested in a quiet, warm room before uptake to reduce muscle and brown-fat uptake.

Quick reference

Therapeutic Nuclear Medicine and Patient Management

  • I-131 therapy for hyperthyroidism uses fixed activities (often 185-555 MBq, 5-15 mCi) or calculated activity from thyroid uptake and mass; ablation after thyroidectomy typically uses 1.1-3.7 GBq (30-100 mCi).

  • I-131 is concentrated by thyroid tissue; blocking agents (iodine, thyroid hormone, amiodarone) must be stopped before therapy, and pregnancy must be excluded with a negative hCG.

  • Radioiodine patient release criteria: the measured dose rate at 1 m is used with the NRC release rule; written instructions are required when the total effective dose equivalent to another individual could exceed 5 mSv (0.5 rem).

  • I-131 bodily fluids are radioactive for days; patients are instructed to flush twice, use separate towels, and avoid sharing utensils, and to sleep in a separate bed for the recommended period.

  • Y-90 microspheres for hepatic tumors are delivered by interventional radiology; a pre-treatment Tc-99m MAA scan estimates lung shunt fraction, and a shunt fraction above 20% contraindicates therapy.

  • Sr-89 chloride and Sm-153 lexidronam palliate painful bone metastases; both suppress marrow, so platelet and white-cell counts must be checked before each dose.

  • Lu-177 DOTATATE for somatostatin-receptor-positive tumors is given with amino-acid infusion to protect the kidneys; the dose is fractionated over multiple cycles.

  • After therapeutic administration, patient room surveys and dose-rate measurements document exposure; the therapy room is surveyed and decontaminated before reuse.

Quick reference

Quality Assurance, Quality Control, and Equipment Calibration

  • Daily gamma-camera QC: photopeak position and energy resolution for each radionuclide used, plus a flood-field uniformity image, are performed before patient imaging.

  • Weekly or monthly QC: spatial resolution and linearity with a bar phantom, and sensitivity with a calibrated source, are performed on a defined schedule.

  • Dose calibrator daily constancy test uses a long-lived check source (Cs-137 or Co-57); accuracy is tested annually with a certified source traceable to a national standard.

  • Dose calibrator linearity is tested at installation and after repair by the decay method or with attenuating sleeves; the result must be within +/-5% of expected activity.

  • Gamma-camera uniformity: integral uniformity should typically be within about 5% and differential uniformity within about 5% for a useful field of view, per the manufacturer's specification.

  • PET scanner QC includes daily detector checks (e.g., a blank or normalization scan), periodic well-counter calibration, and image quality phantoms; time-of-flight systems require timing-offset calibration.

  • A survey meter is calibrated annually and after repair; it is checked daily with a check source for response and battery status, and the calibration must be traceable.

  • Equipment records must document the date, procedure, results, and corrective action; a QC result outside the action level triggers repeat testing and, if persistent, service before clinical use.

Quick reference

Patient Care, Communication, and Professional Ethics

  • Informed consent requires disclosure of purpose, risks, benefits, alternatives, and the right to refuse; for research or novel therapy, written consent is obtained.

  • Pregnancy screening: ask about pregnancy and last menstrual period for patients of childbearing age before any ionizing radiation procedure; document the response in the record.

  • Breastfeeding: after Tc-99m MAA or pertechnetate, interruption periods are specified by the radiopharmaceutical; for I-131 therapy, breastfeeding must be stopped permanently.

  • Patient identification uses two identifiers (name and date of birth or medical record number) checked against the request and the radiopharmaceutical label before administration.

  • Extravasation management: stop the injection, remove the needle, elevate the limb, apply a cold or warm compress per the radiopharmaceutical, and document the event and estimated activity.

  • Radiation risk communication should use absolute risk and comparison terms rather than vague reassurance; the ALARA principle and benefit-risk balance are explained in plain language.

  • Patient confidentiality is protected under HIPAA; images and reports are shared only with authorized persons for treatment, payment, or operations, and de-identification is required for other uses.

  • Professional scope of practice: technologists perform procedures under physician supervision and refer clinical questions to the authorized user or radiologist rather than interpreting results independently.