NNC NSREN Specialist flashcards
NNC NSREN Specialist · Build recall with a focused review session.
32 cards. Read the term and the definition here. The deck at the top of this page is optional.
- Photopeak position check
- Daily gamma-camera test confirming that the photopeak for a given radionuclide, such as Tc-99m at 140 keV, falls within the accepted window; drift indicates PMT gain or high-voltage instability.
- Energy resolution definition
- The FWHM of a photopeak divided by its centroid energy, expressed as a percent. Lower values mean better separation of scattered from primary photons; a Tc-99m camera typically gives 9-11%.
- Integral uniformity
- The maximum minus minimum count difference across the entire useful field of view of a flood image, divided by the maximum plus minimum and expressed as a percent.
- Paralyzable dead time
- A detection model in which each event, including one occurring during dead time, extends the dead period. The observed count rate peaks and then decreases as the true rate rises.
- Collimator resolution versus distance
- For a parallel-hole collimator, spatial resolution worsens linearly as the source-to-collimator distance increases; sensitivity stays roughly constant with distance.
- Effective dose formula
- E equals the sum of tissue weighting factors multiplied by the equivalent dose in each tissue. Tissue weights sum to 1.0, and the result is expressed in sieverts.
- Occupational dose limits
- Annual limits: 50 mSv effective dose, 150 mSv lens of eye, 500 mSv skin and extremities, and 5 mSv to the fetus of a declared pregnant worker.
- Half-value layer relation
- HVL equals 0.693 divided by the linear attenuation coefficient. The number of half-value layers needed equals the natural log of the required attenuation factor divided by ln 2.
- Inverse-square law
- Exposure rate from a point source is inversely proportional to the square of the distance. Tripling distance reduces the exposure rate to one ninth.
- Mo-99 breakthrough limit
- The allowable Mo-99 activity in Tc-99m eluate is 0.15 kBq per MBq of Tc-99m (0.15 uCi per mCi) at the time of use.
- Aluminum breakthrough limit
- Alumina column eluate must contain no more than 10 micrograms of aluminum per milliliter, measured by colorimetric assay; excess aluminum affects labeling efficiency.
- Radiochemical purity
- The fraction of total radioactivity present in the desired chemical form, determined by chromatography and required to meet the package-insert specification before patient administration.
- Dose calibrator constancy test
- A daily measurement of a long-lived check source, such as Cs-137, to confirm that the ionization-chamber response has not drifted from its established baseline value.
- MAA particle count limit
- A Tc-99m MAA perfusion dose typically contains 100,000 to 500,000 particles; fewer are used in pulmonary hypertension or right-to-left shunt to limit capillary blockade.
- Bone scan timing
- After Tc-99m MDP injection, imaging begins 2 to 4 hours later to allow skeletal uptake and renal clearance, with hydration and voiding to reduce bladder artifact.
- Tc-99m pertechnetate trapping
- Pertechnetate is trapped by thyroid follicular cells but is not organified, so it cannot evaluate organification defects; iodine-123 is used when organification must be assessed.
- Hepatobiliary scan endpoint
- Visualization of the gallbladder within about 60 minutes after Tc-99m mebrofenin injection makes acute cholecystitis unlikely; nonvisualization suggests cystic duct obstruction.
- FDG uptake preparation
- Patients fast, control blood glucose, avoid strenuous exercise, and rest in a warm quiet room before F-18 FDG injection to reduce muscle and brown adipose uptake.
- Radioiodine release instructions
- Written instructions are required when another person could receive more than 5 mSv from the released patient; they cover distance, contact time, and separate sleeping arrangements.
- Lung shunt fraction
- The fraction of Tc-99m MAA activity reaching the lungs, estimated before Y-90 microsphere therapy; a shunt above 20% contraindicates hepatic arterial treatment.
- Strontium-89 chloride use
- A beta-emitting bone-seeking agent for palliation of painful osteoblastic metastases. It suppresses bone marrow, so blood counts must be checked before administration.
- Pregnancy exclusion before therapy
- A negative pregnancy test is required before radioiodine therapy in patients of childbearing age, because I-131 crosses the placenta and can ablate fetal thyroid tissue.
- Bar phantom QC
- A weekly or monthly test using a bar phantom to check spatial resolution and linearity of a gamma camera; the bars of each size must be visually resolved.
- Dose calibrator linearity test
- Verification of activity measurement across a range of activities, using decay or attenuating sleeves, with results required to fall within about 5% of expected values.
- Survey meter calibration
- Annual calibration traceable to a national standard plus a daily check-source response and battery check; recalibration is required after repair or a dropped instrument.
- PET daily detector check
- A daily scan of detector blocks, such as a blank or normalization acquisition, to confirm that all detectors respond and that timing offsets remain within specification.
- Two-identifier verification
- Before radiopharmaceutical administration, staff confirm the patient's name and date of birth or medical record number against the request and the labeled dose.
- Extravasation response
- Stop the injection, remove the needle, elevate the limb, apply the compress specified for that radiopharmaceutical, and document the event and estimated retained activity.
- Informed consent elements
- Disclosure of the procedure's purpose, risks, benefits, alternatives, and the right to refuse, documented before the study or therapy proceeds.
- Breastfeeding after I-131
- Breastfeeding must be discontinued permanently after radioiodine therapy because iodine is concentrated in breast tissue and secreted in milk, delivering a dose to the infant.
- Absolute risk communication
- Explaining radiation risk with concrete absolute numbers and comparisons rather than vague reassurance, so the patient can weigh benefit against risk.
- Scope of practice
- Technologists perform imaging and therapy procedures under authorized-user supervision and refer diagnostic interpretation and clinical decisions to the physician.