Mistake 1

On gamma camera uniformity questions, the candidate reports only the integral uniformity value and ignores the differential uniformity value, or applies the wrong formula and mixes the two up when judging whether a daily flood is acceptable.

The two metrics are taught together and both come from the same flood image, so candidates memorise the names but not the definitions. Under time pressure they recall 'uniformity is measured over the useful field of view' and stop there, forgetting that integral uniformity compares maximum and minimum pixel counts across the whole UFOV while differential uniformity compares them within a small moving window.

What to do instead

Before the exam, write out both formulas from memory and rehearse a worked example: integral = ((max - min) / (max + min)) x 100 over the UFOV; differential uses the same expression but over a 5-pixel or manufacturer-specified window. Practise stating which one a daily flood report is judged against and why a camera can pass integral but fail differential.

Mistake 2

When calculating decay or pre-calibration activity, the candidate uses half-life in hours with elapsed time in minutes, or applies the decay formula to the wrong direction (decaying forward when the question asks for activity at calibration).

Most radiopharmacy calculations are simple but unit-sensitive, and candidates rush the arithmetic after correctly identifying the isotope. Tc-99m (6.02 h), F-18 (109.8 min), Ga-68 (67.7 min), and I-131 (8.02 d) are quoted in different units, so mixing them is easy. The forward/backward confusion arises because A = A0 e^(-lambda t) is used for both directions depending on what is known.

What to do instead

Adopt a fixed routine: convert everything to the same unit first, state the half-life you are using on scratch paper, then decide direction by asking 'am I going forward in time or back to calibration?' Do five timed drills covering Tc-99m, F-18, and I-131 in both directions before exam day.

Mistake 3

The candidate answers radiation safety questions using the wrong dose limit for the population in question, for example quoting the 20 mSv/y occupational effective dose limit for a member of the public or for a pregnant radiation worker.

Several limits coexist (occupational 20 mSv/y averaged over 5 years with 50 mSv in any single year; public 1 mSv/y; lens 20 mSv/y; extremity 500 mSv/y; declared pregnant worker 1 mSv to the conceptus over the remainder of the pregnancy). Candidates memorise one number and apply it everywhere, and the declared-pregnancy limit is especially often missed.

What to do instead

Build a one-page table of limits by population and tissue (whole body, lens, extremity/skin, conceptus) and recite it daily. For each practice question, first identify who is exposed, then select the limit. Pay particular attention to the distinction between the 1 mSv public limit and the 1 mSv conceptus limit, which share a number but not a rationale.

Mistake 4

On ALARA and shielding questions, the candidate calculates dose reduction from time and distance but forgets that the inverse-square law applies to a point source and breaks down close to a large-volume source such as a patient or a hot vial.

Textbook inverse-square problems are common in training, so candidates apply 1/r-squared reflexively. Real clinical scenarios involve extended sources, scatter, and the fact that a dose calibrator or generator is not a point, and exam items test whether the candidate recognises the assumption.

What to do instead

When a distance problem appears, first check whether the source is described as small/point-like. If it is a patient or a bulk source, state that inverse-square is an approximation and that measured survey meter readings or published isodose data are preferred. Practice explaining when shielding (lead, tungsten, half-value layer) rather than distance is the practical control.

Mistake 5

The candidate confuses the purpose of the dose calibrator constancy, accuracy, linearity, and geometry tests, and answers a question about which test detects a specific fault with the wrong test name.

All four are performed on the same instrument and are often listed together, so candidates memorise the list without the failure each test is designed to catch. Accuracy (assayed source, within 10 percent), constancy (day-to-day reproducibility with a long-lived source), linearity (activity response across a range, e.g. decay method), and geometry (response vs volume or position in the well) each target a different error mode.

What to do instead

For each test, write one sentence on the fault it detects and the acceptance criterion: accuracy within plus or minus 10 percent of assayed activity; constancy within plus or minus 10 percent of the baseline; linearity within plus or minus 10 percent of expected; geometry consistent across volumes. Then answer practice items by naming the fault first, then the test.

Mistake 6

On radiopharmaceutical quality-control items, the candidate applies the wrong acceptance limit to the radiochemical purity test, for example using a 90 percent limit for Tc-99m labelled products or confusing the RCP limit with the radionuclidic purity limit.

Limits differ by product and by monograph, and candidates generalise from one example. They also conflate radiochemical purity (the fraction of activity in the desired chemical form) with radionuclidic purity (the fraction of activity from the desired radionuclide) and with chemical purity.

What to do instead

Make a product-by-product table: for each commonly used agent, list the RCP method (e.g. paper or ITLC, HPLC), the required RCP limit, and the radionuclidic purity requirement. When answering, identify which purity is being asked about before quoting a number, and state the method that would be used to measure it.

Mistake 7

The candidate misidentifies normal biodistribution or a common variant as pathology, or the reverse, on planar and SPECT image-interpretation questions, for example calling physiological bowel or salivary uptake abnormal or missing a classic artefact.

Image questions reward pattern recognition, and candidates who have memorised textbook disease patterns without studying normal variants default to 'abnormal' when they see asymmetric or focal uptake. Common distractors include free pertechnetate in the stomach and thyroid, renal retention, injection-site activity, and patient-motion or attenuation artefacts.

What to do instead

Study a normal-scan atlas alongside a pathology atlas for each procedure (bone, myocardial perfusion, renal, thyroid, V/Q, hepatobiliary) and explicitly list normal variants and artefacts for each. On exam items, first ask 'could this be physiological, artefactual, or free tracer?' before choosing a diagnosis.

Mistake 8

On therapy questions, the candidate manages the patient as if the administered activity were diagnostic, missing the specific pre-therapy, post-therapy, and release requirements such as pregnancy testing, dosimetry, radiation-safety instructions, and the need to consider bone-marrow or renal reserve.

Therapy content is a smaller share of most curricula, so candidates under-prepare and transfer diagnostic habits. They also forget that therapy activities are governed by separate regulatory frameworks (e.g. written directive, authorised user, and patient-release criteria) rather than the diagnostic imaging rules.

What to do instead

Create a separate therapy checklist for each agent used in practice (e.g. I-131 for hyperthyroidism and for thyroid cancer, radioembolisation, bone palliation): pre-treatment workup, activity determination, written directive, post-treatment imaging, patient instructions, and release criteria. Rehearse one full patient pathway end to end.

Mistake 9

The candidate answers patient-care and communication items with generic 'be empathetic' responses instead of naming the specific professional or ethical action, such as informed consent elements, breaking bad news, or recognising and reporting an error.

These items look like common sense, so candidates skim them and pick the most reassuring-sounding option. The exam rewards specific, defensible behaviour: who is told, what is disclosed, what is documented, and what the escalation path is. Generic answers are usually the intended distractors.

What to do instead

For each scenario type (adverse event, incidental finding, patient refusing a study, language barrier, paediatric patient), write the concrete sequence: assess, inform the responsible clinician, document, escalate per policy. Practise choosing the option that names a specific action over the one that only names a feeling.

Mistake 10

The candidate treats quality-assurance schedule questions as memorisation of intervals only, and gives the correct interval but the wrong test, or the right test at the wrong frequency, for example confusing daily flood with weekly or quarterly checks.

QA schedules are dense tables, and candidates learn them as isolated facts rather than as a system in which each test has a frequency tied to its failure mode and risk. The result is that a question about 'what must be done before first patient use each day' pulls up the wrong test.

What to do instead

Rebuild the schedule as a grid: rows are instruments (gamma camera, dose calibrator, survey meter, well counter), columns are frequency (daily, weekly, monthly, quarterly, annual), and each cell names the test and acceptance criterion. Recite the grid in both directions, from instrument to frequency and from frequency to instrument.